Performance evaluation of Sysmex XN hematology analyzer in umbilical cord blood: a comparison study with Sysmex XE-2100.

Hanah Kim, Mina Hur, Sang-Gyeu Choi, Hee-Won Moon, Yeo-Min Yun, Han Sung Hwang, Han Sung Kwon, In Sook Sohn.   

Abstract

BACKGROUND: The Sysmex XN (XN) modular system (Sysmex, Kobe, Japan) is a new automated hematology analyzer equipped with different principles from its previous version, Sysmex XE-2100. We compared the performances of Sysmex XN and XE-2100 in umbilical cord blood (CB) specimens.
METHODS: In 160 CB specimens, complete blood count (CBC) parameters and white blood cells (WBC) differentials were compared between the two analyzers. Their flagging performances for blasts, abnormal/atypical lymphocytes, immature granulocytes and/or left-shift (IG), and nucleated red blood cells (NRBC) counts were compared with manual counts. For the blast flagging, Q values by Sysmex XN were further compared with manual slide review.
RESULTS: Sysmex XN and XE-2100 showed high or very high correlations for most CBC parameters but variable correlations for WBC differentials. Compared with XE-2100, XN showed significantly different flagging performances for blasts, abnormal/atypical lymphocytes, and IG. The flagging efficiency for blasts was significantly better on Sysmex XN than on XE-2100 (85.0% vs. 38.8%): Sysmex XN showed a remarkably increased specificity of blast flag, compromising its sensitivity of blast flag. Among the 24 specimens with blasts (range, 0.5%-1.5%), only one (4.2%) showed a positive Q value.
CONCLUSIONS: This study highlighted the remarkable differences of flagging performances between Sysmex XN and XE-2100 in CB specimens. The Sysmex XN modular system seems to be a suitable and practical option for the CB specimens used for hematopoietic stem cell transplantation as well as for the specimens from neonates.

Mesh:

Year:  2014        PMID: 24978899     DOI: 10.1515/cclm-2014-0392

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Comparing automated vs manual leukocyte differential counts for quantifying the 'left shift' in the blood of neonates.

Authors:  B C MacQueen; R D Christensen; B A Yoder; E Henry; V L Baer; S T Bennett; H M Yaish
Journal:  J Perinatol       Date:  2016-06-09       Impact factor: 2.521

2.  Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.

Authors:  Jing Wang; Shengmei Zhao; Zhenguo Su; Xiangdong Liu
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352

3.  The Novel Use of Umbilical Cord Blood to Obtain Complete Blood Counts for Critical Neonatal Assessment.

Authors:  Alexandra P Hansen; Gayle D Haischer-Rollo; Jonathan B Shapiro; James K Aden; Jude M Abadie; Thornton S Mu
Journal:  Cureus       Date:  2022-08-14
  3 in total