Evaluation of Mindray BC-3600 hematology analyzer in a university hospital.

G Shu1, H Lu, H Du, J Shi, G Wu.   

Abstract

INTRODUCTION: The BC-3600 Auto Hematology Analyzer (hereinafter call BC-3600) is a quantitative, automated hematology analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories.
METHODS: The analyzer was evaluated and compared with the Mindray BC-3200 3-part differential (BC-3200) and Sysmex XE-2100 5-part differential (XE-2100) Hematology Analyzer in the hematology laboratory of a university hospital. The BC-3600 was evaluated according to guidelines published by Clinical and Laboratory Standards Institute (CLSI), the International Committee for Standardization in Hematology (ICSH), and Department of Food and Drug Administration (FDA).
RESULTS: There were no background, minimal carryover (<0.5%), and excellent linearity for white blood cell (WBC), hemoglobin (Hb) level, red blood cell (RBC), and platelet (PLT) counts (r > 0.999). Precision was good at all levels for the routine cell blood count (CBC) parameters: CV% being ≤2.0, except for platelet count (PLT) at the low level with CV% of ≤5.0% and WBC at the low level with CV% of <3.0%. Correlation between the BC-3600 and BC-3200, XE-2100 were excellent (r > 0.99) for all major CBC parameters.
CONCLUSION: It is concluded that the overall performance of the BC-3600 is excellent and compares well with that of BC-3200 and XE-2100.
© 2012 Blackwell Publishing Ltd.

Mesh:

Year:  2012        PMID: 22925195     DOI: 10.1111/j.1751-553X.2012.01464.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

2.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

3.  Point-of-Care Testing for Anaemia in Children Using Portable Haematocrit Meter: A Pilot Study from Southwest Nigeria and Implications for Developing Countries.

Authors:  Oladele Olatunya; Olatunde Ogundare; Abiola Olaleye; Oyinkansola Agaja; Evelyn Omoniyi; Babajide Adeyefa; Kehinde Oluwadiya; Oyeku Oyelami
Journal:  Ethiop J Health Sci       Date:  2016-05

4.  Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.

Authors:  Jing Wang; Shengmei Zhao; Zhenguo Su; Xiangdong Liu
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.