Evaluation of Mindray BC-5000 hematology analyzer: a new miniature 5-part WBC differential instrument.

D Xiang1, J Yue1, Y Lan1, C Sha1, S Ren1, Y Li1, M Li1, C Wang1.   

Abstract

INTRODUCTION: The Mindray BC-5000 automated hematology analyzer is a miniature, automated hematology analyzer, and 5-part leukocyte differential counter for in vitro diagnostic use in clinical laboratories.
METHODS: Precision, linearity, carryover, and method comparison were carried out. The analyzer was evaluated and compared with the Sysmex XE-2100 hematology analyzer and manual microscopic in the hematology laboratory of a tertiary hospital in Chinese.
RESULTS: There were minimal carryover (<0.25%), and excellent linearity for white blood cell, and platelet counts (r > 0.99). Within-run precision was good at all levels for the routine cell blood count parameters (CV < 3.5%). Between-run precision was acceptable at all levels for the analysis parameters (CV < 5%) except for eosinophil and basophil (CV% >10%). BC-5000 displayed very good correlation (r > 0.94) with the XE-2100 for cell blood count and cell differential parameters except for basophil (r = 0.72). The comparison of 258 leukocyte differential count results analyzed in parallel with manual microscopy, BC-5000 analyzer showed perfect specificity (93%) and negative predictive value (90%) on abnormal blood samples.
CONCLUSION: It is concluded that the overall performance of the BC-5000 is acceptable. The miniature analyzer is suitable for use in small- to medium-size laboratories.
© 2015 John Wiley & Sons Ltd.

Keywords:  Automated hematology analyzer; Mindray BC-5000; leukocyte differential flag; performance evaluation

Mesh:

Year:  2015        PMID: 25944368     DOI: 10.1111/ijlh.12370

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.

Authors:  Jing Wang; Shengmei Zhao; Zhenguo Su; Xiangdong Liu
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352

2.  Effect of haemolysis on an enzymatic measurement of ethanol.

Authors:  Abdulkadir Çat; Kamil Taha Uçar; Alper Gümüş
Journal:  Biochem Med (Zagreb)       Date:  2020-12-15       Impact factor: 2.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.