| Literature DB >> 31213645 |
Andreas P Sauter1, Jana Andrejewski2, Fabio De Marco2, Konstantin Willer2, Lukas B Gromann2, Wolfgang Noichl2, Fabian Kriner3, Florian Fischer3, Christian Braun3, Thomas Koehler4, Felix Meurer5, Alexander A Fingerle5, Daniela Pfeiffer5, Ernst Rummeny5, Julia Herzen2, Franz Pfeiffer5,2.
Abstract
Grating-based X-ray dark-field imaging is a novel imaging modality which has been refined during the last decade. It exploits the wave-like behaviour of X-radiation and can nowadays be implemented with existing X-ray tubes used in clinical applications. The method is based on the detection of small-angle X-ray scattering, which occurs e.g. at air-tissue-interfaces in the lung or bone-fat interfaces in spongy bone. In contrast to attenuation-based chest X-ray imaging, the optimal tube voltage for dark-field imaging of the thorax has not yet been examined. In this work, dark-field scans with tube voltages ranging from 60 to 120 kVp were performed on a deceased human body. We analyzed the resulting images with respect to subjective and objective image quality, and found that the optimum tube voltage for dark-field thorax imaging at the used setup is at rather low energies of around 60 to 70 kVp. Furthermore, we found that at these tube voltages, the transmission radiographs still exhibit sufficient image quality to correlate dark-field information. Therefore, this study may serve as an important guideline for the development of clinical dark-field chest X-ray imaging devices for future routine use.Entities:
Year: 2019 PMID: 31213645 PMCID: PMC6582156 DOI: 10.1038/s41598-019-45256-2
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1X-ray dark-field (a) and transmission (b) radiograph of a human thorax at 60kVp. The dark-field image was low-pass filtered by convolution with a 2D Gaussian kernel (σ = 3.2 px). In the transmission image, an infiltrate is visible in the lower left lung, which translates to a lower dark-field signal in this zone. Of all examined tube voltages (see also Fig. 2), the highest dark-field signal strength can be found at 60 kVp.
Figure 2X-ray dark-field radiographs of a human thorax at the examined tube voltages. All images presented here and in Fig. 1a are shown with the same windowing and the same low-pass filter as in Fig. 1a was applied. The dark-field signal declines significantly toward higher tube voltages. Also, the image quality decreases with higher tube voltages due an increase of imaging artifacts.
Figure 3Signal strength score (median and range) of dark-field (top row) and transmission (bottom row) images for the examined tube voltages. Signal strength score in dark-field images increases from the upper left towards the lower left lung, as the projected thickness of the lung increases. In the right lung, this effect is counteracted by an infiltrate in the lower right lung. In every examined zone, signal strength score decreases for increasing tube voltages. In transmission, signal strength score decreases from the upper lungs to the lower lungs. In every examined zone, the score remains nearly constant for all tube voltages.
Figure 4Image quality score (median and range) of dark-field (right) and transmission (left) images for the examined tube voltages. Image quality score decreases for higher tube voltages in dark-field images. In transmission images, the score remains constant.
Inter-reader agreement (both reading sessions) and agreement for each reader between “grouped” and “individual” reading sessions. Agreement is given in Spearman’s rho. Values of 0.81–1.0 are considered as almost perfect agreement, 0.61–0.80 as substantial agreement, and 0.41–0.60 as moderate agreement. N/A denotes undefined values for Spearman’s rho due to zero variance of one reader’s transmission image quality score.
| Signal strength | Image quality | |||
|---|---|---|---|---|
| Dark-field | Transmission | Dark-field | Transmission | |
|
| ||||
| Readers 1 and 2 | 0.755 | 0.623 | 0.498 | N/A |
| Readers 1 and 3 | 0.754 | 0.674 | 0.310 | 0.861 |
| Readers 2 and 3 | 0.658 | 0.918 | 0.677 | N/A |
|
| ||||
| Reader 1 | 0.814 | 0.999 | 0.473 | 1.000 |
| Reader 2 | 0.952 | 0.980 | 0.882 | N/A |
| Reader 3 | 0.945 | 1.000 | 0.823 | 0.730 |
Figure 5Dark-field contrast as a function of tube voltage in six ROIs (a–f), with respect to an ROI in the vicinity of the lung (g). Each ROI is a 30 × 30 pixel square, placed at highlighted locations in the central overview image. Graphs correspond to ROIs with the same letter. ROIs were placed such that ribs were avoided. Contrast decreases towards higher tube voltages in all lung zones. The horizontal dashed lines in the overview image approximate the extent of the six lung zones assumed in the reader study.
Image acquisition parameters. At every tube voltage one image was acquired. All seven scans were acquired in a time span of 40 min. During each scan, every part of the field of view received 25 × -ray pulses of 20 each. Visibility is given as mean ± standard deviation across the field of view. ESD: entrance surface dose measured at source distance of 152 cm (16 cm above patient table).
| Tube voltage [kV] | 60 | 70 | 80 | 90 | 100 | 110 | 120 |
|---|---|---|---|---|---|---|---|
| Visibility [%] | 36 ± 4 | 29 ± 3 | 22 ± 3 | 19 ± 2 | 18 ± 2 | 18 ± 2 | 18 ± 2 |
| Tube current [mA] | 600 | 360 | 450 | 400 | 360 | 327 | 300 |
| ESD [mGy] | 0.9 | 0.8 | 1.3 | 1.5 | 1.5 | 2.0 | 2.2 |