| Literature DB >> 29422512 |
Katharina Hellbach1, Andrea Baehr2, Fabio De Marco3, Konstantin Willer3, Lukas B Gromann3, Julia Herzen3, Michaela Dmochewitz2, Sigrid Auweter4, Alexander A Fingerle5, Peter B Noël3,5, Ernst J Rummeny5, Andre Yaroshenko6, Hanns-Ingo Maack6, Thomas Pralow6, Hendrik van der Heijden6, Nataly Wieberneit6, Roland Proksa7, Thomas Koehler7,8, Karsten Rindt6, Tobias J Schroeter9, Juergen Mohr9, Fabian Bamberg4,10, Birgit Ertl-Wagner4,10, Franz Pfeiffer3,5,8, Maximilian F Reiser4,10.
Abstract
The aim of this study was to assess the diagnostic value of x-ray dark-field radiography to detect pneumothoraces in a pig model. Eight pigs were imaged with an experimental grating-based large-animal dark-field scanner before and after induction of a unilateral pneumothorax. Image contrast-to-noise ratios between lung tissue and the air-filled pleural cavity were quantified for transmission and dark-field radiograms. The projected area in the object plane of the inflated lung was measured in dark-field images to quantify the collapse of lung parenchyma due to a pneumothorax. Means and standard deviations for lung sizes and signal intensities from dark-field and transmission images were tested for statistical significance using Student's two-tailed t-test for paired samples. The contrast-to-noise ratio between the air-filled pleural space of lateral pneumothoraces and lung tissue was significantly higher in the dark-field (3.65 ± 0.9) than in the transmission images (1.13 ± 1.1; p = 0.002). In case of dorsally located pneumothoraces, a significant decrease (-20.5%; p > 0.0001) in the projected area of inflated lung parenchyma was found after a pneumothorax was induced. Therefore, the detection of pneumothoraces in x-ray dark-field radiography was facilitated compared to transmission imaging in a large animal model.Entities:
Mesh:
Year: 2018 PMID: 29422512 PMCID: PMC5805747 DOI: 10.1038/s41598-018-20985-y
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Pneumothoraces of different sizes depicted in x-ray dark-field and transmission images.Chest radiographs of the same pig (in vivo) just before (A,B) and after application of 250 ml air (C,D) and 500 ml air (E,F) in the left pleural space, comparing transmission images (lower row) with the corresponding dark-field images (upper row). A pneumothorax is indicated in dark-field images as a dark area adjacent to the rib cage on the left side (arrows). After application of 250 ml of air, a small pneumothorax was only visible in the dark-field image, indicated by a discrete widening of the left pleural space. The white scale bar is approximately 5 cm.
Figure 2Unilateral pneumothorax depicted in x-ray dark-field and transmission images.Chest radiographs of a pig (in vivo) after a pneumothorax (arrows) was induced on the right side. The transmission image (B) can be compared directly to the corresponding dark-field image (A). The white scale bar is approximately 5 cm.
Figure 3Contrast-to-noise ratio (CNR) between the air-filled pleural cavity and the adjacent lung tissue for transmission (white bar) and dark-field images (grey bar).CNR was significantly higher for dark-field compared to transmission imaging (p < 0.01).
Figure 4Depiction of a dorsal pneumothorax of a living pig in dark-field and transmission imaging.Scattering lung parenchyma can be visualized directly in dark-field images, even if located behind the diaphragm, making quantification of the area of the inflated lung possible. Before the induction of a pneumothorax (A,B) affected side (orange) 72 cm2, unaffected side (blue) 55 cm2. After the induction of a pneumothorax (right): affected side (orange) 61 cm2, unaffected side (blue) 56 cm2. Indirect signs, such as an increased hyper transparency in the basal parts of the affected lung and consecutively sharper edges of the pleura, indicate the presence of a pneumothorax in the corresponding transmission image (arrows). The white scale bar is approximately 5 cm.