INTRODUCTION: Wearable limb-socket displacement sensors may help patients and prosthetists identify a deteriorating socket fit and justify the need for repair or replacement. METHODS: A novel sensor using an inductive sensing modality was developed to detect limb-to-socket distances. Key detection elements were a coil antenna placed in the socket wall and a magnetic composite sheath worn over the outside of the prosthesis user's elastomeric liner. The sheath was a nylon or cotton prosthetic stocking coated with a polyurethane composite. The polyurethane composite contained embedded iron particles (75 wt%). RESULTS: Brushing γ-glycidoxypropyltriethoxysilane onto the sheath fabric, coating it first with unfilled polyurethane and then iron-filled polyurethane, enhanced bonding between the sheath and the composite and overcame mechanical degradation problems. A γ-glycidoxypropyltriethoxysilane-rich fumed silica layer applied to the outside of the sheath reduced friction and improved durability. Field testing demonstrated less than a 3% signal degradation from four weeks of field use. CONCLUSIONS: The developed wearable displacement sensor meets durability and performance needs, and is ready for large-scale clinical testing.
INTRODUCTION: Wearable limb-socket displacement sensors may help patients and prosthetists identify a deteriorating socket fit and justify the need for repair or replacement. METHODS: A novel sensor using an inductive sensing modality was developed to detect limb-to-socket distances. Key detection elements were a coil antenna placed in the socket wall and a magnetic composite sheath worn over the outside of the prosthesis user's elastomeric liner. The sheath was a nylon or cotton prosthetic stocking coated with a polyurethane composite. The polyurethane composite contained embedded iron particles (75 wt%). RESULTS: Brushing γ-glycidoxypropyltriethoxysilane onto the sheath fabric, coating it first with unfilled polyurethane and then iron-filled polyurethane, enhanced bonding between the sheath and the composite and overcame mechanical degradation problems. A γ-glycidoxypropyltriethoxysilane-rich fumed silica layer applied to the outside of the sheath reduced friction and improved durability. Field testing demonstrated less than a 3% signal degradation from four weeks of field use. CONCLUSIONS: The developed wearable displacement sensor meets durability and performance needs, and is ready for large-scale clinical testing.
Wearable sensors to measure displacements between residual limbs and prosthetic
sockets of lower-limb prosthesis users may have important clinical use towards
monitoring socket fit. Elevated limb–socket displacements reflect pistoning
(excessive vertical translation of the limb in the socket) or excessive sagittal
plane rotation, both of which over time contribute to skin irritation and gait
instability.[1-3] Many prosthesis
users cannot sense excessive motion until their residual limb becomes sore, soft
tissues break down, or they stumble and fall.[4] Typically, the person's prosthesis use is restricted or discontinued until
damaged tissues heal, worsening the disability. Potentially, displacement
measurements can be used to convey to the prosthesis user in the field, via a mobile
phone for example, the need for accommodation to maintain fit (e.g. sock addition,
size change of an adjustable socket).[5] Over the longer term, collected data may help practitioners justify the need
for socket shape adjustment or replacement. Wearable displacement sensors also have
potential use in automatically controlled adjustable socket systems.Commercial displacement sensors have been used by researchers to investigate
limb–socket motion during ambulation. Sanders et al. positioned a photoelectric
sensor of dimension 6-mm diameter and 12-mm length in a hole in the distal socket to
measure limb–socket vertical position and pistoning.[6] Results from a single participant with diabetes wearing a
patellar-tendon-bearing socket with a single sock but no liner demonstrated that a
5-min rest period with the prosthesis doffed caused the limb to shift proximally
4.8 mm upon re-donning, and to piston an additional 3.5 mm during walking. Gerschutz
et al. used an 18-mm diameter, 65-mm long inductive sensor to measure displacements
of the inferior surface of trans-femoral amputee participants' elastomeric liners
during ambulation using elevated vacuum or suction suspension.[7] Pistoning during ambulation for five participants ranged from
0.05( ± 0.04) mm for 20 in Hg elevated vacuum to 2.65( ± 1.21) mm for suction. These
results were comparable to Board et al.'s radiological findings using stationary
techniques to simulate swing phase conditions.[8] Eleven participants averaged 5( ± 2) mm pistoning between liner and socket
for suction suspension and 1( ± 1) mm for vacuum suspension.The commercial sensors used in Sanders et al.'s and Gerschutz et al.'s investigations
provided meaningful data but were bulky and required permanent modification (e.g.
drilled holes) to the participants' sockets. Further, sensor energy demands did not
allow for long-term monitoring. A smaller inductive sensor that used a conductive
fabric target affixed to participant elastomeric liners was much smaller, and under
battery power was wearable for longer term use outside the laboratory.[9,10] It provided useful clinical
data during two-day field tests. Displacement measurements between the anterior
distal liner and socket during ambulation ranged from approximately 5 mm to 10 mm.[9] Within two weeks, however, the sensor suffered degradation problems from
debris build-up between fibers within the fabric target that resulted in an unusable signal.[10]The purpose of this research was to extend from prior investigations to design and
evaluate a novel wearable inductive sensor system for long-term measurement of
limb–socket displacements. A key aspect of the technology was a composite sheath
(stocking) that served as a magnetic target for inductive sensing. The design and
results from bench testing and field testing of the new wearable system on people
with trans-tibial limb loss are presented.
Sensor development
Sensor design
The sensor included an inductive sensing chip (LDC1614, Texas Instruments,
Dallas, Texas), a custom-designed flexible coil antenna (diameter 32.0 mm,
thickness 0.15 mm) with a capacitor (220 pF), and a composite sheath (thin sock)
made with magnetic particles embedded within it (Figure 1(a) and (b)). The composite
sheath was worn on the outside of the participant's elastomeric liner, the same
location as traditional socks or sheaths are worn in clinical practice. When
powered, the inductive sensing chip and antenna operated as an
inductor-capacitor (LC) tank oscillator. The magnetic permeability of the sheath
rich with magnetic filler material reinforced the inductor and so lowered the
oscillation frequency when the sheath was near the antenna. This effect was
distance dependent, and so by measuring changes in the LC tank oscillator
frequency a measurement reflecting the distance of the sheath from the antenna,
i.e. the liner-to-socket distance, was collected.
Figure 1.
Sensor system. (a) Block diagram showing arrangement of sensor
components. (b) Antenna and instrumented socket. Left: A schematic
of the antenna. The black rectangle is the tank capacitor. Right:
The regions of the composite sheath embedded with magnetic particles
are shown as red stripes. The gold areas on the socket represent the
antennas embedded within the socket. The blue layer is the
elastomeric prosthetic liner.
Sensor system. (a) Block diagram showing arrangement of sensor
components. (b) Antenna and instrumented socket. Left: A schematic
of the antenna. The black rectangle is the tank capacitor. Right:
The regions of the composite sheath embedded with magnetic particles
are shown as red stripes. The gold areas on the socket represent the
antennas embedded within the socket. The blue layer is the
elastomeric prosthetic liner.
Composite design
Three silicones (Shore 00–10 platinum cure; Shore 30A platinum cure; and a Shore
30A tin cure (Smooth On, Macungie, Pennsylvania)) and two polyurethanes (Shore
10A; Shore 30A (Smooth On)) were considered for the composite matrix of the
sheath. Flexible polymers were chosen since addition of the magnetic powder was
expected to stiffen the material. The composite sheath needed to not be so stiff
as to affect normal limb–socket interactions during prosthesis use. For the
magnetic filler material, the following materials with strong magnetic
permeability were considered: iron (Sigma Aldrich, St Louis, Missouri),
iron–nickel alloy (Alfa Aesar, Haverhill, Massachusetts), and
manganese–zinc–ferrite (Powder Tech International, Valparaiso, Indiana). Two
particle sizes were tested: 6–10 μm diameter and ∼40 μm diameter (300-mesh
material). Pucks of diameter 50 mm and thickness 3 mm were fabricated at filler
concentrations between 60 wt% and 85 wt%. Signal strength was tested using a
testing jig that held the target above the antenna (Figure 2). Distances between the antenna
and target were measured with a digital height gauge (Mitutoyo 570, Aurora,
Illinois). In this setup, the sensor output tank frequency values relative to an
external clock, thus the signal from the sensor was the ratio of the tank
frequency to the external clock frequency in units of Hz/Hz. Changes in signal
amplitude were shifts in this ratio. The sample was adhered to the benchtop
using double-sided tape. The antenna was lowered into contact with the target
and data collection was initiated. Signal strength was measured every 0.05 mm up
to 1.00 mm and then every 0.25 mm thereafter up to a point where no signal
change was observed.
Figure 2.
Test setup with digital scale and test sample. This device was used
to evaluate sensor sensitivity for different composite designs.
Test setup with digital scale and test sample. This device was used
to evaluate sensor sensitivity for different composite designs.The iron–nickel alloy and manganese–zinc–ferrite-filled pucks produced signal
amplitudes 61% and 68%, respectively, of that of the iron-filled pucks. Because
of the weaker signals, iron–nickel alloy and manganese–zinc–ferrite-filled pucks
were not tested further, and only iron was used in subsequent testing. Further,
some people may develop contact sensitization to nickel.[11] Two instances of contact sensitization to iron have been reported in the
literature but those were to iron salts, specifically ferric chloride
(FeCl3) and ferric sulfate
(Fe2(SO4)3) rather than metallic iron as
used here.[12] The concentration of iron affected signal sensitivity, with a maximum
amplitude achieved at 83 wt% iron. Concentrations greater than 83 wt% prevented
all of the tested polymer from reaching full cure. Concentrations above 75 wt%
but below 83 wt% were good. The response curves were non-linear (Figure 3). A third-order
polynomial fit the data within 1.7% full-scale error. The 75 wt% concentration
proved the best combination of strong signal intensity and sufficient
elasticity. At this stage in development, elasticity was qualitatively assessed,
using our expertise gained from testing and using liner materials in prosthetics
research for many years. Excessively stiff material may cause the prosthesis
user discomfort during ambulation, and by restricting radial expansion may limit
fluid movement within the residual limb.
Figure 3.
Sensitivity testing results from evaluation of puck samples.
Sensitivity increased with iron concentration. A polynomial curve
fit for the 75 wt% is shown.
Sensitivity testing results from evaluation of puck samples.
Sensitivity increased with iron concentration. A polynomial curve
fit for the 75 wt% is shown.Addition of iron created a fluid with excessive viscosity for both polymer types
that proved difficult to spread evenly onto a sheath. Thus, a heating profile
was incorporated into the fabrication procedure. Preheating the polymer in an
oven at 50℃ for 3 min reduced viscosity of the polyurethanes to an acceptable
level but the silicones still remained too viscous to spread evenly, and pot
life of the silicones was decreased below an acceptable level (Table 1). As a result,
the silicones were removed from further consideration.
Table 1.
Pot life sensitivity of matrix candidates to heating profiles.
Pot life (minutes)
Polyurethane (shore 10A and 30A)
Silicone (platinum cure 10A, 30A; tin cure 30A)
Unheated (control)
30.0
30.0
Preheated polymer
20.0
10.0
Heated layup
12.5
8.0
Preheated and heated layup
7.5
5.0
Pot life sensitivity of matrix candidates to heating profiles.Testing of samples using the preheated polymer was conducted on 10.0 cm square
iron-filled polyurethane targets at 70, 75, 80, and 83 wt% iron. Each target was
adhered to the bottom of the test apparatus and a 32.0-mm diameter antenna was
mounted to the digital height gauge. Position data were collected in 0.25 mm
increments above the target surfaces. The mean value of 1100 points collected at
each height during a 30 s interval was determined and converted to millimeters
using calibration data (Figure
3). Results showed that resolution was sensitive to iron
concentration at far distances (>9.0 mm) from the target (Figure 4).
Figure 4.
Resolution results from evaluation of puck samples. Measurement
resolution at high antenna-to-target distances depended upon iron
filler concentration.
Resolution results from evaluation of puck samples. Measurement
resolution at high antenna-to-target distances depended upon iron
filler concentration.
Composite-sheath bonding
Amputee prosthetic sheaths are made primarily from nylon or cotton and typically
include a small percentage of Lycra. Nylon sheaths are typically used to
facilitate donning of close fitting sockets, and cotton sheaths are typically
used to fill excess space in looser fitting sockets. Iron-infused polyurethane
was cured onto swatches of both sheaths. While Shore 10A polyurethane with
75 wt% iron cured successfully when molded in puck form, it showed a tendency to
remain tacky when made in larger swatches. For this reason, the softer
polyurethane was removed from consideration, and only Shore 30A polyurethane was
used in further testing. While the Shore 30A iron-infused polyurethane cured
well, it tended to delaminate from the nylon (Figure 5, left). Examination under the
microscope showed that no strong chemical bonds formed between the polyurethane
and the nylon fibers. The Shore 30A iron-infused polyethylene delaminated less
from the cotton than the nylon. However, the iron composite tended to sit on top
of the cotton fibers rather than encapsulate them, which weakened mechanical
integrity (Figure 5,
right).
Figure 5.
Results from composite-to-sheath bonding evaluations. Left: Nylon
sample showing delamination. Right: Edge of cotton sheath showing
composite sitting on top of the cotton rather than encapsulating the
cotton fibers.
Results from composite-to-sheath bonding evaluations. Left: Nylon
sample showing delamination. Right: Edge of cotton sheath showing
composite sitting on top of the cotton rather than encapsulating the
cotton fibers.Cordova and Rowan invented a means to improve chemical adhesion between polyamide
fibers and polyurethane composites using a coupling agent,
γ-glycidoxypropyltriethoxysilane (GLYEO).[13] GLYEO has three ethoxy groups bound to a quaternary silicon atom with an
epoxide at the end of a propoxy group bound to the same silicon. GLYEO has been
used as a coating for medical purposes both inside the body[14-16] and on the skin[17,18] with no
adverse reactions. The material was created by combining it with water, ammonia,
and isopropyl alcohol in the molar ratios 1.55:0.30:1.63 for each mole of GLYEO.
The water served to hydrolyze the ethoxy groups, but this process was slow, so
the ammonia was added to catalyze the reaction. GLYEO is not water soluble until
it reacts and becomes a silanol. However, it is miscible in alcohol. The alcohol
helped to reduce the extent to which the unreacted GLYEO separated from the
solution. Vigorous stirring was still required to prevent separation. Once the
solution was applied to the fabric, the epoxide groups formed a bond with the
tertiary nitrogen atoms in the nylon. The newly created hydroxides crosslinked
and bonded to isocyanate groups in the polyurethane that was applied on top of
the coating. This coating was not needed for the cotton sheath because the
cellulose chains already had an abundance of hydroxide groups available.Testing of 75 wt% iron-filled polyurethane onto GLYEO-coated nylon swatches
showed better adhesion than on uncoated samples, but still mechanical coupling
was weak. The high viscosity of the iron-filled polyurethane continued to be
problematic. Applying a layer of iron-free polyurethane, which had lower
viscosity than the iron-filled composite, onto GLYEO-coated nylon swatches
first, followed by a layer of 75 wt% iron-filled polyurethane had less
delamination. However, penetration of the iron-free polyurethane into the sheath
material was still not optimal. Reduced viscosity samples of iron-free
polyurethane were created by changing the A:B mixing ratio from that recommended
by the manufacturer (1:1). Part A contained the isocyanate monomers to which
GLYEO bonded, while Part B contained the polyols. A mixing ratio of 1.00:0.75
improved the polymer's ability to withstand delamination. Having an excess of
the isocyanate monomers allowed the polyols to fully react, while providing
plenty of bonding opportunity for GLYEO. The 1.00:0.75 ratio was used for all
subsequent polyurethane, both iron-free and 75 wt% iron-filled, on both nylon
and cotton sheaths. It was important that the base polyurethane layer cure to a
level at which it did not mix with the iron-filled polyurethane, as mixing of
the layers would undermine the base layer's ability to fully wet the fabric, but
at the same time too long a cure time resulted in few chemical bond connections.
An ideal level was reached by allowing the base layer to cure in a 50℃ oven for
30 min before applying the filled polymer.
External coating
While the developed two-layer polyurethane adhered well to the nylon and cotton
sheaths, it cured with a resin-rich, high friction surface that increased shear
stresses against the socket, reducing wear resistance of the sheath. To reduce
friction and wear, coatings were created to apply over the outside of the
iron-filled polyurethane layer. Modified versions of the GLYEO solution were
generated. One coating included addition of iron as a hard, low friction
inorganic phase. The hydroxide groups on GLYEO crosslinked among themselves and
created a mixture of covalent and double hydrogen bonds with the metallic
particles. A second coating replaced the iron with fumed silica. Fumed silica is
an amorphous silica (SiO2) formed by pyrogenically reacting silicon
tetrachloride with hydrogen and oxygen gas. The resulting particles had a highly
branched structure with a large surface area, providing substantially more
bonding area than the spherical iron particles. Because silica and silanes like
GLYEO are chemically similar, bonding was done by oxygen bonded to silicon in
both cases, thus the bonds created were stronger than with iron.[19]The iron and silica coatings were tested on nylon sheaths prepared with the
GLYEO, iron-free polyurethane, and 75 wt% iron-filled polyurethane layers
described above. The iron coating was tested at concentrations of 37 wt%,
41 wt%, and 45 wt%. The fumed silica was tested at concentrations of 22 vol%, 38
vol%, 53 vol%, and 69 vol% (Fused-silica is low density thus vol% was used
instead of wt%). The iron 45 wt%, fumed silica 22 vol%, 53 vol%, and 69 vol% all
produced unfavorable results and were not tested further (Table 2).
Table 2.
Results from testing candidate external coatings.
Coating
Concentration
Bench testing
After 3-day use as an “athletic sock”
Iron
37 wt%
Acceptable
12.00% Signal degradation after 3 days
41 wt%
Acceptable
0.51% Signal degradation after 3 days, with lower than
desired viscosity
45 wt%
Unacceptable: Coating oversaturated, leading to an iron
film that rubbed off
NA
Fumed silica
22 vol%
Unacceptable: Coating had a rough texture, with lower
than desired viscosity
NA
38 vol%
Acceptable
0.30% Signal degradation after 3 days; viscosity within
the desired range
53 vol%
Unacceptable: Coating had a flakey texture with large
agglomerates
NA
69 vol%
Unacceptable: Coating had a flakey texture with large
agglomerates
NA
Results from testing candidate external coatings.To test the remaining three candidate coatings (iron 37 wt%, iron 41 wt%, fumed
silica 38 vol%) under repetitive mechanical loading, we created nylon sheaths
molded into the form of an “athletic sock.” Nylon sheaths were stretched over a
silicone mold of a foot, and iron-filled polyurethane was brushed onto the
plantar surface at the heels and metatarsal heads, locations subject to high
pressures and shear stresses during ambulation, using the procedures described
above. The coatings were applied and signal strength was tested. The “athletic
socks” were worn by able-bodied individuals inside their shoes for three days
and then retested for signal strength. It is recognized that plantar stresses
are typically higher than stresses at the residual limb–prosthetic socket
interface;[20,21] however, this testing provided a starting point for
evaluation. There was a large decrease in signal strength for the iron 37 wt%
sample (12.00%). The iron 41 wt% sample showed a smaller decrease in signal
strength but the viscosity was lower than desired. The fumed silica 38 vol%
sample showed the lowest signal degradation (0.30%) (Table 2) and was used in subsequent
testing.
Localized application of composite polymer
Thin nylon and cotton prosthetic sheaths were coated with the layers described
above at locations of interest to monitor limb–socket displacements. The primary
locations of interest included an inferior region (to monitor pistoning), and
the anterior distal, posterior distal, anterior proximal, and posterior proximal
areas (to monitor sagittal plane rotation).To fabricate the composite sheath, a nylon (Knit-Rite, Super Stretch, Kansas
City, Kansas) or cotton (Paceline Interface Sheath, Matthews, North Carolina)
sheath was stretched over a foam positive of comparable size to the residual
limb to be monitored. A polyvinyl acetate (PVA) bag was placed inside of the
sheath to prevent the composite coating from seeping through to the underside of
the sheath. Areas on the sheath where the composite coating was not to be
applied were identified and coated with uncured PVA. After the PVA cured
(∼15 min), vinyl tape (ULine 96GK Electrical Tape, Pleasant Prairie, Wisconsin)
was placed over the PVA coated areas.A GLYEO coating was brushed onto the exposed areas of the sheath, and the
iron-free polyurethane was immediately added on top of it, using a stiff brush
to work the polyurethane into the weave. Excess polymer was removed using a soft
foam brush, and the mold was placed in the oven at 50℃ for 30 min.Once the layers were set, iron-filled polyurethane was applied using a brush and
the positive mold was pressed into a preheated negative silicone mold (Figure 6). Then the entire
structure was returned to the oven. Preheating the negative silicone mold to 50℃
before adding the uncured polymer replaced the step of preheating the polymer. A
10.8 kg mass was placed on top of the positive mold to increase pressure and
facilitate formation of a smooth surface. After 60 min, the positive was
demolded from the negative mold and the fumed silica-filled GLYEO coating was
applied. The coating was dried in the fume hood for 10 min, placed in the oven
for 2 h at 50℃ and then allowed to cure for 20 h. The tape was removed, and the
PVA and bag were removed using soap and water.
Figure 6.
Magnetic composite sheath formation. A coated sheath on a foam
positive about to be inserted into the negative mold.
Magnetic composite sheath formation. A coated sheath on a foam
positive about to be inserted into the negative mold.
Summary of fabrication process
The final fabrication process, summarized in Appendix 1, started with pulling a
sheath over a foam positive for a socket, masking the areas intended to remain
free of composite material with PVA, and then curing for 15 min. The exposed
areas of the sheath were then coated with unfilled GLYEO and immediately coated
again with unfilled polyurethane. The sheath was allowed to cure for 30 min at
50℃. The filled polyurethane layer, which was 75 wt% iron, was then applied to
the sheath, which was pressed into a pre-heated negative mold and allowed to
cure at 50℃ for 1 h. The positive was removed from the mold and GLYEO/silica
coating, containing 38 vol% silica, was applied, dried in the fume hood for
10 min, and allowed to cure for 2 h at 50℃. The sheath was allowed to cure at
room temperature for 20 h.The time devoted to active preparation of the sheaths was approximately 2.5 h;
the total elapsed time from start to finish was 25.9 h. A total of about 1.5 h
was required to apply the two GLYEO layers, 30 min for the two polyurethane
layers, and 25 min to apply the mask to the sheath.Production of the sheaths required a means to degas the polyurethane. We used an
unheated vacuum oven (Model 3606, Lab-Line Instruments, Melrose Park, Illinois)
with a vacuum pump (Model 0523, Gast Manufacturing, Benton Harbor, Michigan).
While the temperature requirements were low, the exterior dimensions of the mold
were larger than many benchtop ovens could accommodate (27 cm tall by 20 cm in
diameter) so we used an incubator (Imperial III, Lab-Line) to heat the mold
during the curing process. Before curing, GLYEO emits fumes which can irritate
the airways and when hydrolyzed it releases ethanol, a flammable gas. Thus, the
reaction was carried out in a fume hood.
Target-antenna misalignment testing
We conducted tests to determine to what extent minor sheath misalignment affected
sensor results. It is known that conductive targets produce strong signals when
the antenna is fully covered by the target material,[22] but the performance of magnetic targets was less clear. To test this
issue, a 10.0-cm square magnetic target was adhered to the lab bench. An antenna
mounted on the testing fixture arm 3.0 mm above the target was moved in
horizontal steps across the target while the signal was recorded.The signal decreased approximately 11% on average from its maximal value when the
edges of the target and antenna were even with each other (Figure 7). An 11% signal loss
corresponded to a displacement of approximately 0.44 mm.
Figure 7.
Results from testing target antenna misalignment: decay in signal
amplitude due to edge proximity. The antenna diameter was 3.2 cm.
Based on these results, during clinical use the center of the
32.0-mm diameter antenna was required to be positioned at least
5.1 cm from the target edge. This meant that the target positioned
on a liner was required to overlap at least 3.5 cm with the edge of
the antenna.
Results from testing target antenna misalignment: decay in signal
amplitude due to edge proximity. The antenna diameter was 3.2 cm.
Based on these results, during clinical use the center of the
32.0-mm diameter antenna was required to be positioned at least
5.1 cm from the target edge. This meant that the target positioned
on a liner was required to overlap at least 3.5 cm with the edge of
the antenna.Based on these results, and consideration of the possibility for misalignment of
the donned liner, the target positioned on a liner was required to overlap at
least 3.5 cm with the edge of the antenna. This specification required the
target to be at least 10.2 cm across.We also tested the tensile stiffness of the composite sheath, using testing
procedures developed for evaluation of prosthetic liner materials.[23] A mechanical testing machine (5940 load frame, Instron, Norwood,
Massachusetts) with a 2000 N load cell was used.
Pressure testing and mechanical property characterization
Because the composite areas were deformable under the loads encountered within a
socket, it was important to determine the effect of compression on the recorded
signal. Compressive stress was applied using the mechanical testing machine. A
custom fixture made of acetal was designed (Figure 8) to insulate the antenna from
the metal in the load frame. Pressures up to 250 kPa were applied to a composite
sheath sample, while the signal was recorded. A magnitude of 250 kPa was used
based on the prior research measuring interface stresses on people with limb loss.[24] Since the target did not rest completely flat on the base, an 11 kPa
pre-load was applied. The test was replicated at four different
antenna-to-target distances: 1.5, 2.5, 4.0, and 4.5 mm. The signal shift was
calculated for each pressure by subtracting the frequency ratio with only the
pre-load applied from the ratio measured with the pressure applied divided by
the pre-load frequency ratio.
Figure 8.
Custom test fixture for characterizing effect of compression on the
magnetic composite material. The antenna was placed within a cavity
in the base so that only the composite target was compressed.
Custom test fixture for characterizing effect of compression on the
magnetic composite material. The antenna was placed within a cavity
in the base so that only the composite target was compressed.Compression increased the amplitude of the signal (Figure 9), and the magnitude of the
increase was dependent on the target's proximity to the antenna, with smaller
distances inducing greater error. However, even at the smaller distances
measured (1.5 mm, 2.5 mm), the increase was small, less than 0.12% signal shift
at 250 kPa.
Figure 9.
Compression testing results from a magnetic composite sheath with a
nylon substrate. Percent signal shift increased with stress and with
distance from the target. However, the increases were small.
Compression testing results from a magnetic composite sheath with a
nylon substrate. Percent signal shift increased with stress and with
distance from the target. However, the increases were small.From material testing machine data, the magnetic composite sheath reduced in
thickness to 0.16 mm under a compressive load of 250 kPa. This localized
thickness reduction was not expected to have clinical impact during prosthesis
use since it is less than the deformation an elastomeric liner experiences
during use. The tensile stiffness of the magnetic composite sheath with a nylon
substrate was 1.61 MPa.
Field testing on people with limb loss
Participants were included in this study if they had a trans-tibial amputation at
least 18 months prior and regularly used a definitive prosthesis at least 4 h
per day without assistive aides (e.g. a cane). Candidate participants were
excluded if they were currently experiencing skin breakdown or other soft tissue
injury. Approval from the University of Washington Institutional Review Board
(approval #42899) and written informed consent from participants were obtained
before any human subject testing procedures were initiated.Both participants were male and had their amputations as a result of traumatic
injury. Participant characteristics are listed in Table 3.
Participant characteristics.PTB: patellar tendon bearing; TPE: thermoplastic elastomer.The wear resistance of the sheaths was evaluated by measuring signal strength
before and after sheaths were worn by study participants. Signal strength was
measured using the following technique. An antenna was sandwiched between a
carbon fiber tile and a 1.8-mm thick plastic plate, simulating its condition
within a socket. This assembly was affixed to the lab bench. A 44.0-mm thick
foam spacer was placed inside the sheath to isolate panels on opposite sides
from contacting each other during testing. Measurement points were marked at the
lateral and medial sides of the distal and proximal ends of the panels so that
positions could be replicated during post-wear tests. For each sensing location
on the target (medial distal, lateral distal, medial proximal, lateral proximal
on both anterior and posterior surfaces (eight measurement sites total)), a
weight was placed on the exposed surface of the composite sheath, inducing a
compressive stress of approximately 25 kPa and a measurement was taken for 5 s.
The signal mean, for each 5 s measurement, was subtracted from the mean value
without weight to calculate signal amplitude. With the initial values recorded,
the participant was sent home with the sheath and asked to wear it as part of
his or her normal routine. After two to four weeks, the participant returned the
sheath to the lab, and the marked locations were retested using the same
method.Results showed signal losses ranging from 0.1% to 1.7%, except for the anterior
medial distal site for the four-week test, which was 2.8% (Figure 10).
Figure 10.
Signal loss from field use of the magnetic composite sheath. One
participant used the sheath for two weeks and the other for four
weeks.
Signal loss from field use of the magnetic composite sheath. One
participant used the sheath for two weeks and the other for four
weeks.
Discussion
Wearable sensors that measure displacements between residual limbs and sockets of
prosthesis users must be durable and long lasting to be usable in clinical care.
Field data collected over weeks or months may help a patient and prosthetist
identify a deteriorating socket fit and the need for modification or replacement.
Data from such a wearable sensor might also be used as a feedback signal in
automatically controlled adjustable socket systems, which may improve patient
outcomes.Results from the present study demonstrate that the use of ferromagnetic targets,
instead of conductive targets as used previously,[10] improved performance of inductive sensors for limb–socket displacement
measurement. Conductive target performance relies on eddy currents induced on the
target surface by current flowing through the sensor's antenna, and thus requires a
continuous, defect-free layer of conductive material. Ferromagnetic targets operate
on a magnetic permeability principle, thus can be discontinuous, allowing them to
take the form of particles suspended within a polymeric matrix, a configuration that
is more wear tolerant. The change in oscillation frequency generated by the presence
of the iron-seeded composite polymer overwhelmed other magnetic or electric fields,
thus the system was not sensitive to the wearer touching metal objects or stray
fields in the user's environment.Chemical strategies helped tackle mechanical bonding and surface texture issues in
the present development effort. The base layer of unfilled polyurethane helped to
prevent delamination because its lower viscosity allowed it to penetrate and
encapsulate the nylon fibers. GLYEO improved adhesion by creating a chemical bond
between the nylon and polyurethane. The epoxide groups were able to form bonds with
amides in the nylon while the silanol groups, produced by hydrolysis, bound to
di-isocyantes in the prepolymer. The opposite NCO groups allowed those bound to
GLYEO to be incorporated into the polyurethane matrix. NCO is the chemical formula
for an isocyanate functional group in which carbon is double-bonded to nitrogen and
oxygen, and nitrogen is also bound to the host molecule. In this case, the molecule
was methylbenzene with an additional isocyanate group. GLYEO was also used to create
a low-friction surface (on the external surface of the sheath). The silanol groups
crosslinked with each other, binding to the di-isocyanates as before, and binding to
oxygen atoms on the surface of the silica particles. Compared to both GLYEO and
polyurethane, silica was very hard and its presence reduced the polymer's tendency
to stick to textured surfaces that the target came in contact with, decreasing its
coefficient of friction.The developed magnetic composite sheaths performed better than conductive targets in
field testing. Conductive fabric targets caused at least a 50% signal reduction
within two weeks,[10] while in the present study less than 3% of the signal was lost over four
weeks. We suspect that the reason the anterior distal location in the present study
showed the highest signal loss at four weeks was because this site tends to be
subjected to high pressures and shear stresses during ambulation,[13] which plastically deformed the composite in the sheath, causing areas of
local thinning. Additional testing would need to be conducted to determine if the
effect worsens under longer wear durations, such as the 6 + month longevity of
prosthetic liners.The tensile stiffness of the magnetic composite sheath did not prove problematic for
participants in this investigation in terms of discomfort. However, it is noted that
the measured tensile stiffness of 1.61 MPa is higher than that of polyurethane or
silicone elastomeric liners commonly used in clinical prosthetics practice, but
comparable to that reported for some thermoplastic elastomer liners.[25] If they were worn directly on the skin instead of over the outside of the
liner, the sheaths may be less comfortable. None of the materials in the sheath,
once cured, would be expected to cause chemically induced skin irritation.
Potentially, iron particles could be seeded directly into the elastomeric material
of the liner during fabrication, possibly reducing the mechanical stiffness issue
and improving convenience of the technology. A next step in development of the
limb–socket displacement sensor is to conduct clinical trials to determine if
deterioration in socket fit metrics preclude deterioration in socket comfort
metrics, such as the Socket Comfort Score and subsections of the Prosthesis
Evaluation Questionnaire.[26,27] If limb–socket displacements proved to be a useful measure of
prosthetic fit, then a next step would be to communicate collected data to
prosthesis users, via a mobile phone app for example, informing them when a
deterioration in socket fit occurred and that accommodation (e.g. sock change or
socket size adjustment) was needed. Collected data could also be accumulated and
prepared into summary reports for practitioners, augmenting patient self-report at
clinical visits.
Authors: Ethan J Weathersby; Clement J Gurrey; Jake B McLean; Benjamin N Sanders; Brian G Larsen; Ryan Carter; Joseph L Garbini; Joan E Sanders Journal: Sensors (Basel) Date: 2019-09-19 Impact factor: 3.576