| Literature DB >> 31054217 |
Satoshi Tanabe1, Satoru Utsunomiya2, Eisuke Abe3, Hiraku Sato4, Atsushi Ohta3, Hironori Sakai5, Takumi Yamada5, Motoki Kaidu3, Hidefumi Aoyama3.
Abstract
PURPOSE: We evaluated the setup accuracy of a three-degree-of-freedom fiducial marker (3DOF-FM)-based setup compared to a soft tissue (ST)-based setup in hypofractionated intensity-modulated radiotherapy (IMRT) for prostate cancer.Entities:
Keywords: cone-beam computed tomography (CBCT); fiducial marker; hypofractionated radiotherapy; image-guided radiotherapy (IGRT); intensity-modulated radiotherapy (IMRT); prostate cancer
Mesh:
Year: 2019 PMID: 31054217 PMCID: PMC6560240 DOI: 10.1002/acm2.12603
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Patient and treatment characteristics
| Data | n = 16 |
|---|---|
| Age, yrs; median (range) | 72 (62–85) |
| Initial PSA, ng/ml; median (range) | 11.2 (5.5–20.9) |
| Gleason score: | |
| ≤6 | 2 |
| 7 | 10 |
| ≥8 | 4 |
| Clinical stage (7th UICC): | |
| T1c | 2 |
| T2a | 6 |
| T2b | 0 |
| T2c | 8 |
| D'Amico risk group: | |
| Low | 1 |
| Intermediate | 4 |
| High | 11 |
| Radiotherapy dose (Gy): | |
| 62 Gy in 20 fractions | 6 |
| 70 Gy in 28 fractions | 10 |
PSA, prostate specific antigen; UICC, Union for International Cancer Control.
Figure 1Flowchart of the study protocol.
Figure 2The frequency histogram for the differences between ST‐based and FM‐based translational shifts: (a) anterior‐posterior (AP), (b) superior‐inferior (SI), and (c) left‐right (LR) axes.
Percentage of number data of the shifts in each axis for the fiducial marker‐based setup and soft tissue‐based setup
| Shifts | Soft tissue‐based setup | ||
|---|---|---|---|
| ≤ −2 mm | within 2 mm | ≥ 2 mm | |
| Fiducial marker‐based setup | |||
| AP | |||
| ≤ −2 mm | 4.8% | 5.0% | 0% |
| within 2 mm | 0.8% | 41% | 13% |
| ≥ 2 mm | 0% | 3.5% | 32% |
| SI | |||
| ≤ −2 mm | 9.0% | 4.0% | 0% |
| within 2 mm | 1.8% | 50% | 5.0% |
| ≥ 2 mm | 0% | 7.5% | 23% |
| LR | |||
| ≤ −2 mm | 0% | 0.3% | 0% |
| within 2 mm | 1.8% | 97% | 0.3% |
| ≥ 2 mm | 0% | 0% | 0.3% |
AP, anterior‐posterior; SI, superior‐inferior; LR, left‐right.
Figure 3The frequency histogram for the rotational shifts detected by the ExacTrac x‐ray system after the BS‐based setup: (a) AP, (b) SI, and (c) LR axes.
Figure 4Two‐dimensional correlations (a–c) and Bland‐Altman error analysis (d–f) between the ST‐based setup and FM‐based shifts: (a) AP, (b) SI, and (c) LR axes.