| Literature DB >> 15465147 |
Luc Beaulieu1, Louis-Martin Girouard, Sylviane Aubin, Jean-François Aubry, Lucie Brouard, Lise Roy-Lacroix, Jean Dumont, Daniel Tremblay, Jacques Laverdière, Eric Vigneault.
Abstract
Online prostate positioning using gold markers and a standard video-based electronic portal imaging device is reported. The average systematic (random) errors have been reduced from 2.1 mm (2.7 mm) to 0.5 mm (1.5 mm) in AP direction, 1.1 mm (1.7 mm) to 0.7 mm (1.2 mm) SI and 1.2 mm (1.7 mm) to 0.6 mm (1.3 mm) LR.Mesh:
Substances:
Year: 2004 PMID: 15465147 DOI: 10.1016/j.radonc.2004.06.006
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280