Literature DB >> 12128111

Implementation and utility of a daily ultrasound-based localization system with intensity-modulated radiotherapy for prostate cancer.

Jody Morr1, Thomas DiPetrillo, Jen San Tsai, Mark Engler, David E Wazer.   

Abstract

PURPOSE: To evaluate the clinical feasibility of daily computer-assisted transabdominal ultrasonography for target position verification in the setting of intensity-modulated radiotherapy (IMRT) for prostate cancer. METHODS AND MATERIALS: Twenty-three patients with clinically localized prostate cancer were treated using a sequential tomotherapy IMRT technique (Peacock) and daily computer-assisted transabdominal ultrasonography (BAT) for target localization. Patients were instructed to maintain a full bladder and were placed in the supine position using triangulation tattoos and a leg immobilizer to minimize pelvic rotation. The BAT ultrasound system is docked to the treatment collimator and electronically imports the CT simulation target contours and isocenter. The system is able to use the machine isocenter as a reference point to overlay the corresponding CT contours onto the ultrasound images captured in the transverse and sagittal planes. A touch screen menu is used to maneuver the CT contours in three dimensions such that they match the ultrasound images. The system then displays the three-dimensional couch shifts required to produce field alignment. Data were prospectively collected to measure the frequency by which useful ultrasound images were obtained, the amount of time required for localization/setup, and the direction/magnitude of the positional adjustments.
RESULTS: Of the 23 patients, the BAT ultrasound system produced images of sufficient quality to perform the overlay of the CT contours in 19 patients such that positional verification could be reliably performed. Poor image quality was associated with patient inability to maintain a full bladder, large body habitus, or other anatomic constraints. Of the 19 assessable patients, a total of 185 treatment alignments were performed (mean 8.8/patient). For all cases, the average time required for the daily ultrasound imaging and positional adjustments was 11.9 min. After the initial 5 cases, the user experience skills improved such that the time required for image verification/positional adjustments decreased to a mean of 5.6 min. The average right-left, AP, and cranial-caudal adjustment was 2.6 +/- 2.1 mm, 4.7 +/- 2.7 mm, and 4.2 +/- 2.8 mm, respectively. Positional adjustments >10 mm were infrequent and related primarily to misidentification of the target structures on the ultrasound image, patient movement, or improper registration of the triangulation tattoos.
CONCLUSION: Daily computer-assisted BAT ultrasound positional verification of the prostate can be successfully performed through the acquisition of high-quality images in most patients with only a modest increase in treatment setup time. Positional data obtained with this system resulted in clinically meaningful adjustments in daily setup for sequential IMRT that would not be otherwise apparent from other verification modalities.

Entities:  

Mesh:

Year:  2002        PMID: 12128111     DOI: 10.1016/s0360-3016(02)02820-1

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  14 in total

1.  The impact of androgen deprivation therapy on setup errors during external beam radiation therapy for prostate cancer.

Authors:  Cem Onal; Yemliha Dolek; Yurday Ozdemir
Journal:  Strahlenther Onkol       Date:  2017-04-13       Impact factor: 3.621

2.  Is it essential to use fiducial markers during cone-beam CT-based radiotherapy for prostate cancer patients?

Authors:  Berna A Yildirim; Cem Onal; Yemliha Dolek
Journal:  Jpn J Radiol       Date:  2016-10-11       Impact factor: 2.374

3.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

4.  Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate.

Authors:  Douglas J Moseley; Elizabeth A White; Kirsty L Wiltshire; Tara Rosewall; Michael B Sharpe; Jeffrey H Siewerdsen; Jean-Pierre Bissonnette; Mary Gospodarowicz; Padraig Warde; Charles N Catton; David A Jaffray
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-03-01       Impact factor: 7.038

5.  Dosimetric comparison of image guidance by megavoltage computed tomography versus bone alignment for prostate cancer radiotherapy.

Authors:  Jörn Kalz; Florian Sterzing; Kai Schubert; Gabriele Sroka-Perez; Jürgen Debus; Klaus Herfarth
Journal:  Strahlenther Onkol       Date:  2009-04-16       Impact factor: 3.621

6.  Reduced rectal toxicity with ultrasound-based image guided radiotherapy using BAT (B-mode acquisition and targeting system) for prostate cancer.

Authors:  Markus Bohrer; Peter Schröder; Grit Welzel; Hansjörg Wertz; Frank Lohr; Frederik Wenz; Sabine Kathrin Mai
Journal:  Strahlenther Onkol       Date:  2008-12-24       Impact factor: 3.621

7.  Current concepts on imaging in radiotherapy.

Authors:  Michela Lecchi; Piero Fossati; Federica Elisei; Roberto Orecchia; Giovanni Lucignani
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-31       Impact factor: 9.236

8.  Comparison of daily couch shifts using MVCT (TomoTherapy) and B-mode ultrasound (BAT System) during prostate radiotherapy.

Authors:  Steven H Lin; Elizabeth Sugar; Terrance Teslow; Todd McNutt; Habeeb Saleh; Danny Y Song
Journal:  Technol Cancer Res Treat       Date:  2008-08

9.  Tomotherapy as a tool in image-guided radiation therapy (IGRT): theoretical and technological aspects.

Authors:  S Yartsev; T Kron; J Van Dyk
Journal:  Biomed Imaging Interv J       Date:  2007-01-01

10.  Reconstruction of the treatment area by use of sinogram in helical tomotherapy.

Authors:  Akihiro Haga; Keiichi Nakagawa; Calvin Maurer; Ken Ruchala; Edward Chao; Dylan Casey; Satoshi Kida; Dousatsu Sakata; Masahiro Nakano; Taiki Magome; Yoshitaka Masutani
Journal:  Radiat Oncol       Date:  2014-11-28       Impact factor: 3.481

View more