Abstract
PURPOSE: The purpose of this study was to assess the reliability of automated breast ultrasound (ABUS) examinations of suspicious breast masses in comparison to handheld breast ultrasound (HHUS) with regard to Breast Imaging Reporting and Data System (BI-RADS) category assessment, and to investigate the factors affecting discrepancies in categorization.Entities:
Keywords: Automated breast ultrasound; Breast Imaging Reporting and Data System; Breast neoplasms; Hand-held ultrasound
Year: 2018 PMID: 30999717 PMCID: PMC6595129 DOI: 10.14366/usg.18055
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Characteristics of the study group
| Characteristic | Value |
|---|---|
| Age (yr) | 49.0±9.7 (35-82) |
| Tumor size (mm) | 18.6±11.5 (4-66) |
| Clinical findings | |
| No symptom | 77 (57.0) |
| Lump | 54 (40.0) |
| Nipple discharge | 4 (3.0) |
Values are presented as mean±SD (range) or number (%).
Cross-tabulation of Breast Imaging Reporting and Data System categorization by automated versus handheld breast ultrasound
| ABUS category | HHUS category | ||||
|---|---|---|---|---|---|
| 4A | 4B | 4C | 5 | Total | |
| All cases (n=135) | |||||
| 4a | 81 | 9 | 4 | 1 | 95 |
| 4b | 1 | 9 | 4 | 1 | 15 |
| 4c | 1 | 0 | 4 | 3 | 8 |
| 5 | 0 | 0 | 4 | 13 | 17 |
| Total | 83 | 18 | 16 | 18 | 135 |
| Malignancy (n=49) | |||||
| 4A | 7 | 4 | 3 | 1 | 15 |
| 4B | 1 | 3 | 4 | 1 | 9 |
| 4C | 1 | 0 | 4 | 3 | 8 |
| 5 | 0 | 0 | 4 | 13 | 17 |
| Total | 9 | 7 | 15 | 18 | 49 |
ABUS, automated breast ultrasound; HHUS, handheld breast ultrasound.
Fig. 1.Microinvasive ductal carcinoma of the left breast in a 56-year-old woman.
A, B. Craniocaudal and mediolateral mammography images indicate grouped fine pleomorphic microcalcifications in the 3H direction of the left breast. C, D. On automated breast ultrasound imaging, a subtle, indistinct, irregular, hypoechoic mass with microcalcifications in the corresponding location was noted, which was classified as Breast Imaging Reporting and Data System category 4B. E. On handheld breast ultrasound imaging, a relatively more prominent, spiculated, irregular, heterogeneous, echoic mass was found in the same location, which was classified as Breast Imaging Reporting and Data System category 5.
Odds ratios of mammographic and ultrasonographic findings for a lower Breast Imaging Reporting and Data System categorization being assigned by automated than by handheld breast ultrasound
| Total | Lower grade on ABUS, n (%) | Univariable analysis | Multivariable analysis | |||
|---|---|---|---|---|---|---|
| OR (95% CI) | P-value | OR (95% CI) | P-value | |||
| Mammographic finding | ||||||
| Density | ||||||
| B | 23 | 4 (17.4) | 1 | - | ||
| C | 68 | 11 (16.2) | 0.87 (0.26-2.89) | 0.816 | - | - |
| D | 44 | 8 (18.2) | 1.01 (0.28-3.57) | 0.989 | - | - |
| Microcalcifications | ||||||
| None | 102 | 11 (10.8) | 1 | - | ||
| Present | 33 | 12 (36.4) | 4.63 (1.83-11.71) | 0.001 | - | - |
| Ultrasonographic finding | ||||||
| Size (cm) | ||||||
| <1 | 32 | 5 (15.6) | 1 | - | ||
| >1 and ≤2 | 61 | 7 (11.5) | 0.69 (0.21-2.27) | 0.539 | - | - |
| >2 and ≤5 | 39 | 11 (28.2) | 2.02 (0.64-6.33) | 0.229 | - | - |
| >5 | 3 | 0 | 0.71 (0.03-1.588) | 0.832 | - | - |
| Shape | ||||||
| Oval, round | 46 | 2 (4.3) | 1 | 1 | ||
| Irregular | 89 | 21 (23.6) | 5.59 (1.43-21.83) | 0.013 | 3.75 (0.78-17.97) | 0.099 |
| Orientation | ||||||
| Parallel | 124 | 20 (16.1) | 1 | - | ||
| Nonparallel | 11 | 3 (27.3) | 2.10 (0.55-7.95) | 0.275 | - | - |
| Margin | ||||||
| Circumscribed | 51 | 5 (9.8) | 1 | - | ||
| Irregular | 71 | 14 (19.7) | 2.13 (0.74-6.12) | 0.160 | - | - |
| Indistinct | 13 | 4 (30.8) | 4.00 (0.96-16.70) | 0.057 | - | - |
| Echo pattern | ||||||
| Isoechoic | 16 | 1 (6.3) | 1 | - | ||
| Complex solid and cystic | 7 | 2 (28.6) | 4.70 (0.49-44.38) | 0.177 | - | - |
| Hypoechoic | 94 | 16 (17.0) | 2.17 (0.37-12.60) | 0.387 | - | - |
| Heterogeneous | 18 | 4 (22.2) | 3.21 (0.44-23.27) | 0.249 | - | - |
| Posterior echogenicity | ||||||
| None | 105 | 18 (17.1) | 1 | - | ||
| Enhancement | 16 | 2 (12.5) | 0.82 (0.19-3.42) | 0.780 | - | - |
| Shadowing | 11 | 3 (27.3) | 1.95 (0.51-7.45) | 0.330 | - | - |
| Combined | 3 | 0 | 0.68 (0.03-13.64) | 0.798 | - | - |
| Calcification | ||||||
| None | 115 | 14 (12.2) | 1 | 1 | ||
| Present | 20 | 9 (45.0) | 5.78 (2.08-16.04) | 0.001 | 5.37 (1.75-16.46) | 0.003 |