| Literature DB >> 35204629 |
Shahad A Ibraheem1, Rozi Mahmud1,2, Suraini Mohamad Saini1,2, Hasyma Abu Hassan1, Aysar Sabah Keiteb3, Ahmed M Dirie4.
Abstract
OBJECTIVE: To compare the diagnostic performance of the automatic breast volume scanner (ABVS) against the handheld ultrasound (HHUS) in the differential diagnosis of benign and malignant breast lesions.Entities:
Keywords: BI-RADS; automatic breast volume scanner; breast cancer; handheld ultrasound
Year: 2022 PMID: 35204629 PMCID: PMC8870745 DOI: 10.3390/diagnostics12020541
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1Bar charts for Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) analysis for 21 studies of diagnostic accuracy.
Quality of studies included in diagnostic accuracy analysis for risk of bias and applicability concerns.
| Study | Risk Of Bias | Applicability Concerns | |||||
|---|---|---|---|---|---|---|---|
| Patient Selection | Index Test | Reference Standard | Flow and Timing | Patient Selection | Index Test | Reference Standard | |
| Depretto et al., 2020 |
|
|
|
|
|
|
|
| Jia et al., 2020 |
|
|
|
|
|
|
|
| Tutar et al., 2020 |
|
|
|
|
|
|
|
| Yun et al., 2019 |
|
|
|
|
|
|
|
| Zhang et al., 2019 |
|
|
|
|
|
|
|
| Niu et al., 2019 |
|
|
|
|
|
|
|
| Choi et al., 2018 |
|
|
|
|
|
|
|
| Zhang et al., 2018 |
|
|
|
|
|
|
|
| Schmachtenberg et al., 2017 |
|
|
|
|
|
|
|
| Hellgren et al., 2016 |
|
|
|
|
|
|
|
| Kim et al., 2016 |
|
|
|
|
|
|
|
| Jeh al., 2015 |
|
|
|
|
|
|
|
| Chio et al., 2014 |
|
|
|
|
|
|
|
| Kim et al., 2013 |
|
|
|
|
|
|
|
| Chen et al., 2013 |
|
|
|
|
|
|
|
| Lin et al., 2012 |
|
|
|
|
|
|
|
| Wang et al., 2012 |
|
|
|
|
|
|
|
| Wojcinski et al., 2011 |
|
|
|
|
|
|
|
| Chang et al., 2011 |
|
|
|
|
|
|
|
| Shin et al., 2011 |
|
|
|
|
|
|
|
Low Risk High Risk ? Unclear Risk.
Figure 2Flow Chart of Study Selection.
Characteristics of the included studies.
| Author | Study Design/Objectives/Participants | Screening Method | Findings | Outcome |
|---|---|---|---|---|
| Tutar et al., 2020 | To prospectively compare between the automated breast volumetric scanning (ABVS) with the hand-held bilateral whole breast ultrasound (HHUS) in lesion detection, and characterization. | ABVS with integrated 14L5BV linear transducer (15.4 cm) HHUS with 14L5 (5–14 MHz) or 9L4 (4–9 MHz) linear transducer |
Recall rate was 46/340 (13.05%) for ABVS, and 4/340 (1.18%) for HHUS. HHUS had higher results of true negatives (BI-RADS 1–2) while ABVS had higher results of false positives ( ABVS had a positive predictive value of 4.17% while HHUS had 50%. In comparison to HHUS, ABVS had irregular nodules of ( 59.7% of the women mentioned that if they had a choice, they would have chosen HHUS. 10.6% of the women experienced severe pain from the use of ABVS. |
The use of ABVS in lesion detection is as good as HHUS. ABVS had a higher recall rate and lower positive predictive value. This could end up in greater follow-ups, and greater anxiety for the women. Had they been given the choice, more than 50% of the women would have preferred HHUS. |
| Depretto et al., 2020 | To examine the agreement between the hand-held ultrasound (HHUS) and the automated breast ultrasound (ABUS) in monitoring of women with breast cancer history, with regard to the contralateral breast cancer or recurrences or new ipsilateral | The Selenia Dimension mammography system was used in conducting the mammography (Hologic; Bedford, MA). ABVS with a 6–14 MHz frequency. HHUS with a linear 6–15 MHz transducer. |
ABUS and HHUS were substantial for dichotomic assessment (κ = 0.794) and for BI-RADS categories (κ = 0.785). There was a significant difference in assigning the BI-RADS categories ( |
A substantial agreement was achieved between ABUS and HHUS in monitoring of women with breast cancer history. In particular, the ABUS could be used in first-level monitoring of intermediate risk women since it could recognize all cancers detected by HHUS. |
| Jia et al., 2020 | To determine, both in combination with mammography and separately, the diagnostic performance of the automated breast ultrasound system (ABUS) and the handheld ultrasound (HHUS) in dense breasts Chinese women. | The mammograms were acquired using the Fujifilm FDR MS-2500 (Fujifilm Crop., Tokyo, Japan), GE Sengraphe DS (GE Healthcare, WI, USA), and Hologic Selenia (Hologic, MA, USA). The HHUS images were obtained using the Aixplorer system (Supersonic Imagine, Aix-en-Provence, France), GE LOGIQ9 (GE Healthcare, WI, USA), iU22 Ultrasound System (Philips Medical System, WA, USA), and S2000 (Siemens Medical Solutions, CA, USA). All ABUS scans were acquired using the Invenia ABUS (GE Healthcare, WI, USA). |
In mammography-negative dense breasts, the rate for incremental cancer detection was 42.8 for each 1000 ultrasound examinations. The combination of HHUS or ABUS with mammography produced a sensitivity of 99.1% (219/221), and the specificities were 84.9% (608/716) and 86.9% (622/716), respectively. The combination of the HHUS with mammography produced a 0.92 area under the curve, while a combination of ABUS with mammography produced one at 0.93. An agreement that is statistically significant in breast cancer detection between HHUS and ABUS was observed (percent agreement = 0.94, κ = 0.85). |
As adjuncts to mammography, both the HHUS and ABUS can substantially increase the rate of breast cancer detection in dense breasts women; a strong correlation exists between them. With various benefits of the ABUS over HHUS, for instance reproducibility and less operator dependence, and the commonness of dense breasts, the use of ABUS in the early detection of breast cancer, especially in areas that have limited resources, has shown great potential. |
| Yun et al., 2019 | In terms of the assessment of Breast Imaging Reporting and Data System (BI-RADS) category—to assess the reliability of suspicious breast masses examination performed using the automated breast ultrasound (ABUS) as opposed to the handheld breast ultrasound (HHUS). To examine factors that affect categorization discrepancies. | ABUS exams HHUS images were acquired using the linear transducer at 7–15 MHz |
There was an overall good agreement in all cases between HHUS and ABUS (79.3%, kappa = 0.61, It was revealed in the logistic regression analysis that differences in the categorization of BI-RADS were associated with the suspicious presence of microcalcification on the mammography (odds ratio [OR], 4.63; 95% confidence interval (CI), 1.83 to 11.71; |
The examination of suspicious breast masses under the categorization of BI-RADS have demonstrated good agreement between HHUS and ABUS. The presence of an irregular shape on US and the accompaniment of suspicious microcalcifications on mammography were factors linked to the yielding of a lower level of suspicion in the ABUS compared to the HHUS regarding the assessment of BI-RADS category. |
| Zhang et al., 2019 | To investigate the diagnostic performance of the automated breast ultrasound system (ABUS), for women 40 years or older for breast cancer, compared to mammography (MG) and hand-held ultrasonography (HHUS). | ABVS with 6–14 MHz linear broadband transducer HHUS with 14L5 (5–14 MHz) linear transducer | 75 cases were malignant while 519 were benign or normal. The sensitivity, specificity, accuracy and Youden index were 97.33%, 89.79%, 90.74% and 0.87 for HHUS; 90.67%, 92.49%, 92.26% and 0.83 for ABUS; 84.00%, 92.87%, 91.75% and 0.77 for MG, respectively. Compared to HHUS, ABUS had a significantly superior specificity ( HHUS had highest area under the receiver operating characteristic curve at 0.936, followed by ABUS at 0.916, and MG at 0.884. The difference was not statistically significant ( |
ABUS yielded an equivalent diagnostic performance for breast cancer as MG and HHUS, and therefore can be potentially utilized as an alternative technique for the diagnosis of breast cancer. |
| Niu et al., 2019 | To examine the diagnostic potential of the automated breast ultrasound (ABUS) system in differentiating malignant and benign breast masses in comparison to handheld ultrasound (HHUS). | A 6–14-MHz linear broadband transducer ABUS A 6–18-MHz linear transducer (18 L6) HHUS |
Pathological results confirmed 599 masses in total, in 398 women. 496 of the 599 masses were benign while the remaining 103 were malignant. No significant differences were found between HHUS and ABUS in terms of positive predictive value (46.46% versus 46.12%), diagnostic accuracy (80.6% versus 80.1%), negative predictive value (95.67% versus 97.96%), and specificity (80.24% versus 77.62%). Significant differences were found in sensitivity (82.52% versus 92.23%; The correlation of the maximum diameter was relatively greater between ABUS and the pathological results (r = 0.885) in comparison to between HHUS and the pathological results (r = 0.855). However, the difference was not significant ( |
In distinguishing between benign and malignant breast masses, automated breast US is better than HHUS, particularly regarding specificity. |
| Choi et al., 2018 | To examine the hand-held ultrasound (HHUS) compared to the automated breast volume scanner (ABVS), based on the fifth edition of BI-RADS ultrasound | The 15-cm-wide linear array transducer with 5–14 MHz was used for ABVS examinations The HHUS examinations involved the use of the ACUSON S2000 ultrasound system (or the ACUSON Sequoia 512 |
There was moderate to good interobserver agreement in ABVS and HHUS (κ = 0.53–0.67 and 0.55–0.70, respectively), with the exception for associated features for BI-RADS lexicons (κ = 0.31 and 0.36, respectively). Irregular shape, posterior features (combined or shadowing), and a non-circumscribed margin were individually linked to in both the ABVS and HHUS, malignancy. The existence of calcification on ABVS (odds ratio (OR), 95% confidence interval (CI): 2.09, 1.11–3.94), and non-parallel orientation on HHUS (OR: 95% CI: 2.04, 1.10–3.78) were individually linked to malignancy. No significant differences were found between HHUS and ABVS in sensitivity (84.2% vs. 84.2%), specificity (83.9% vs. 80.5%), or AUC (0.90 vs. 0.88). |
Based on the fifth BI-RADS edition, there is no statistically significant difference between ABVS and HHUS in terms of diagnostic performance and interobserver variability. |
| Zhang et al., 2018 | To assess the clinical performance of ABUS in comparison to mammography (MG) and the handheld ultrasound (HHUS), for breast cancer detection. | ABVS with 6–14 MHz linear broadband transducer. HHUS was performed using the GE LOGIQ9, Aixplorer system, iU22 Ultrasound System and s2000. MG images included Fujifilm FDR MS-2500, Hologic Selenia, GE Sengraphe DS, dan. |
620 (31.4%) and 1,353 (68.6%) of these were classified as BI-RADS categories 4–5 and 1–3, respectively. The Kappa value and the agreement rate between the ABUS and HHUS were 0.860 ( In terms of consistency between the results of pathology and imaging, 78.6% of women classified as BI-RADS 4–5 using ABUS later were diagnosed as having cancer or precancerous lesions. This was 7.2% higher compared to women classified using HHUS. The false-negative rates of HHUS and ABUS for BI-RADS 1–2 was much lower than those of MG and were nearly identical. |
A good diagnostic reliability was observed for ABUS. ABUS is thus a promising alternative in detecting breast cancer in China due to its lower dependence on the operator and its performance in detecting breast cancer in women with high-density breasts. |
| Schmachtenberg et al., 2017 | To determine the diagnostic value of the automated breast volume scanning (ABVS) in comparison to the handheld ultrasonography (HHUS) by using the breast magnetic resonance imaging (MRI) as the gold standard. | ABVS with −14 MHz linear broadband transducer. HHUS with 14L5 (5–14 MHz) linear transducer MRI |
HHUS detected 54 lesions, MRI detected 72 lesions, and ABVS detected 59 lesions. No significant difference was found between HHUS and ABVS regarding sensitivity (100% vs. 93.3%), specificity (83.3% vs. 83.3%), diagnostic accuracy (89.7% vs. 87.2%), positive predictive value (78.9% vs. 77.8%), and negative predictive value (100% vs. 95.2%). In terms of lesion localization (same quadrant), the agreement was 91.2% for MRI and HHUS, and 94.3% for MRI and ABVS. The assessment of size of lesion was (+/−3 mm) correct in 80% (ABVS) and 79.4% (HHUS) compared to MRI lesion size. The correlation of measurement of size was moderately higher for ABVS-MRI (r = 0.89) than for HHUS-MRI (r = 0.82), with |
ABVS is a probable option to HHUS. Although ABVS has limitations in assessing the axillary lymph nodes, and is lacking in elastography or Doppler capacities, which occasionally give significant additional information in HHUS, ABVS has the advantages of better reproducibility and operator independence. |
| Hellgren et al., 2016 | To compare the specificity and sensitivity of ABVS with the handheld breast US in detecting breast cancer under the situation of recall post mammography screening. | ABVS with 14 MHz frequency. HHUS using a linear L17-5 transducer or a L12-5. |
Twenty-six cancers were discovered in 25 women. Both ABVS and handheld US, used for suspicious mammographic finding in breasts ( ABVS and the handheld US, used in breasts with negative mammography (n¼103), yielded the sensitivity of 100% (1/1). The specificity of the ABVS was 94.1% (96/102) while the specificity was 100% (102/102) for the handheld US. |
The ABVS has the potential to replace the handheld US for the investigation of women recalled from mammography screening due to dubious mammographic findings. |
| Kim et al., 2016 | To compare the diagnostic performance of the automated breast volume scanner (ABVS) and the handheld ultrasound (US) as a second-look US techniques subsequent to preoperative breast magnetic resonance imaging (MRI) | MRI scan. ABVS imaging that was performed contained a 5–14-MHz wide-aperture linear transducer. HHUS with 7–15-MHz and 6–14-MHz linear array transducers. |
For the second-look examination, the ABVS has a higher detection rate compared to the handheld US (94.7% vs. 86.8%; Out of 76 lesions in total, only 1 was discovered by the handheld US, only 7 were identified by the ABVS, while neither the handheld US nor the ABVS could detect the 3 lesions. Both the handheld US and the ABVS had a lower ability in detecting non-mass lesions compared to the ability in detecting mass-type lesions ( |
As a method for pre-operational assessment of breast cancer patients, ABVS has a higher efficiency compared to handheld US for a second-look US examination subsequent to preoperative breast MRI. In non-mass lesion detection, both techniques have limitations. |
| Jeh et al., 2015 | To compare the clinical utility of HHUS and ABUS in breast lesion diagnosis and detection. | ABVS with a 5–14 MHz wide-aperture linear probe. HHUS using a 7–15 MHz and a 6–14 MHz linear transducer. | ABUS overall detection rate was 83.0% while HHUS overall detection rate was 94.2%. Ten lesions, of which were microcalcifications (nine benign lesions and one malignant), could neither be detected by ABUS nor HHUS. Out of 194 lesions detected by HHUS, ABUS detected 169 while the other 25 benign lesions were left undetected. ABUS could less frequently detect smaller-sized lesions, including those of lower final-assessment category and benign appearance ( |
The detection of all malignant lesions by HHUS were similarly made by ABUS. ABUS failed to detect a few smaller benign lesions. |
| Choi et al., 2014 | ABVS with 5–14 MHz with a 9 MHz centre frequency. HHUS with a bandwidth of 5–12 MHz and a linear array transducer. |
ABVS had a recall rate of 2.57 per 1000 (48/1866) while HHUS was 3.57 per 1000 (132/3700), with a substantial difference ( Cancer detection yield was 3.8 per 1000 for ABVS, and 2.7 per 1000 for HHUS. The diagnostic accuracy, with a statistical significance of ( The specificity of HHUS and ABVS were 96.7% and 97.8%, respectively ( |
The performance of ABVS is as good as HHUS in detecting lesions. ABVS could lead to greater anxiety for the women due to more follow-ups since it has a lower positive predictive value and higher recall rate. If given the option, over 50% of the women would prefer HHUS. | |
| Kim et al., 2013 | To compare between the automated whole breast ultrasound (AWUS) and the hand-held breast ultrasound (HHUS) in terms of their detection performance. To evaluate the variability of interobserver in interpreting AWUS. | Mammograms. ABVS with a 5–14 MHz frequency. HHUS using a 7–15 MHz linear probe and a 6–14 MHz linear probe. |
The malignancy detection rate for HHUS was 98.0%, while that of the three readers for AWUS were 90.0%, 88.0% and 96.0%. In HHUS, the specificity and sensitivity were 62.5% and 98.0%, 90.0% and 87.5% for reader 1, 88.0% and 81.3% for reader 2, and 96.0% and 93.8% for reader 3, in AWUS No significant difference was found in the sensitivity, specificity, and detection performance of the radiologists ( Fair to good interobserver agreement was found for size, location of breast masses, ultrasonographic features, and categorization. |
AWUS is assumed as beneficial for breast lesion detection. There is no significant difference between HHUS and AWUS in terms of specificity, detection rate, and sensitivity. However, AWUS demonstrated high degree of interobserver agreement. |
| Chen et al., 2013 | To clarify ABVS value compared to HHUS in the differentiation of malignant and benign breast masses | ABVS with integrated (15.4 cm) 14L5BV linear transducer HHUS with 9L4 (4–9 MHz) or 14L5 (5–14 MHz) linear transducers. |
No differences were shown between the ABVS and HHUS regarding accuracy (88.1% vs. 87.2%), specificity (86.2% vs. 87.5%), sensitivity (92.5% vs. 88.0%), false-positive rate (13.8% vs. 12.5%), false-negative rate (11.8% vs. 7.5%), negative predictive value (96.3% vs. 94.3%), and positive predictive value (74.7% vs. 75.6%) ( |
There is no difference in diagnostic accuracy between HHUS and ABVS in discriminating malignant or benign breast masses. |
| Lin et al., 2012 | To examine the clinical utility of automated breast volume scanner (ABVS) compared to the handheld ultrasound for detecting and providing diagnosis of breast lesions. | ABVS with a wide aperture 14L5BV linear array transducer and central frequency of transducer varied from 9 to 11 MHz. HHUS that uses the 18L6HD linear array transducer at 10 MHz grayscale central frequency. |
95 breast lesions were detected by both automated breast volume scanner and handheld ultrasound. Both handheld ultrasound and ABVS demonstrated high specificity (85.0%, and 95.0% respectively), and high sensitivity (both 100%). For breast neoplasms, a higher diagnostic accuracy was demonstrated by handheld ultrasound (91.4%) than ABVS (97.1%). No significant difference was demonstrated in maximum diameters of 2D, ABVS, and pathology ( There was better correlation with pathology (r = 0.616) than 2D (r = 0.468) for ABVS. |
The automated breast volume scanner is a promising modality in breast imaging with the benefits of operator-independence, high diagnostic accuracy, whole-breast visualization, and greater prediction of lesion size. |
| Wang et al., 2012 | To assess the diagnostic value for the discrimination of breast lesions that are benign and malignant, between the automated breast volume scanning (ABVS) and the conventional handheld ultrasonography (HHUS). | ABVS with frequency of 5–14 MHz transducer (15.4 cm). HHUS with 18L6 linear transducer. |
The pathology of 239 breast lesions revealed that 154 (64.4%) were benign and 85 (35.6%) were malignant lesions. There are similarities between ABVS and HHUS regarding specificity (80.5% vs. 82.5%), sensitivity (95.3% vs. 90.6%), accuracy (85.8% vs. 85.3%), positive predictive value (73.0% vs. 74.0%), and negative predictive value (93.3% vs. 94.1%) Only minor differences were demonstrated by the area under the receiver operating characteristic (ROC) curve used to estimate the accuracy between HHUS and ABVS (0.928 and 0.948, respectively). |
In the differentiation of breast lesions that are malignant and benign, HHUS and ABVS had almost identical diagnostic accuracy. ABVS may assist in uncovering small lesions, demonstrating the presence of intraductal lesions, and distinguishing between real lesions and inhomogeneous areas. This technique is feasible for clinical application and is a promising new modality in breast imaging. |
| Wojcinski et al., 2011 | To assess the detectability of breast lesions by independent examiner using only ABVS data of breast lesions that were previously detected using conventional ultrasound. | ABVS with 14L5BV linear transducer (14 MHz, 15.4 cm). HHUS using 18L6 HD linear transducer (5.5–18 MHz). |
The experimental ABVS yielded a sensitivity of 100% in the described setting (95% CI: 73.2–100%), and 66.0% diagnostic accuracy (95% confidence interval (CI: 52.9–79.1). The specificity was maintained at 52.8% (95% CI: 35.7–69.2). Comparison between the ABVS concordance with the gold standard (conventional handheld ultrasound) indicated a fair agreement between both, with the value of Cohen Kappa as an estimation of the inter-rater reliability of κ = 0.37. |
For breast ultrasound, ABVS must continue to be perceived as an experimental technique, thus requiring further evaluation studies. |
| Chang et al., 2011 | To examine, retrospectively the detection performance of malignant and benign breast masses with the utilization of 3D volume data obtained by ABUS, and to determine the variables of lesion that influence detectability | ABVS with 7.5–10 MHz transducer. HHUS with 6–14 MHz linear transducer. |
The sensitivities for the detection of benign and malignant masses were 65.2% (30/46), 95.8% (23/24) for reader 1 ( It is demonstrated in the logistic analysis that mass shape (odds ratio, 95% CI; 3.12, 1.02–9.55), size of the mass (odds ratio, 95% CI; 1.12, 1.02–1.24), and changes in surrounding tissue (odds ratio, 95% CI; 0.11, 0.02–0.47) were the variables linked to the ABUS detectability. |
Significantly higher sensitivity was demonstrated in reader studies using ABUS data for breast masses that are malignant than benign. |
| Shin et al., 2011 | To evaluate, prospectively, the interobserver agreement of five radiologists in detecting lesions, and the characterization in the review of automated ultrasound breast images. | ABVS using a 5–14 MHz with a 9-MHz centre frequency. HHUS using a linear 5–12 MHz transducer. |
Using the automated ultrasound, the identification of 145 lesions were made by a minimum of two observers. Lesions larger than 1.2 cm had a detection rate of 92%. Majority of lesions detected using only handheld ultrasound (11/12, 92%) or automated ultrasound (34/36, 94%) were probably benign masses or cysts. There is high reliability since all intraclass correlation coefficients for size and location of lesion exceeded 0.75. |
The reporting of lesion location and size yielded high reliability. The description of key features and final assessment category demonstrated substantial agreement. |