Design and evaluation of a grid reciprocation scheme for use in digital breast tomosynthesis.

Tushita Patel1, Helen Sporkin2, Heather Peppard3, Mark B Williams1,2,3.   

Abstract

This work describes a methodology for efficient removal of scatter radiation during digital breast tomosynthesis (DBT). The goal of this approach is to enable grid image obscuration without a large increase in radiation dose by minimizing misalignment of the grid focal point (GFP) and x-ray focal spot (XFS) during grid reciprocation. Hardware for the motion scheme was built and tested on the dual modality breast tomosynthesis (DMT) scanner, which combines DBT and molecular breast tomosynthesis (MBT) on a single gantry. The DMT scanner uses fully isocentric rotation of tube and x-ray detector for maintaining a fixed tube-detector alignment during DBT imaging. A cellular focused copper prototype grid with 80 cm focal length, 3.85 mm height, 0.1 mm thick lamellae, and 1.1 mm hole pitch was tested. Primary transmission of the grid at 28 kV tube voltage was on average 74% with the grid stationary and aligned for maximum transmission. It fell to 72% during grid reciprocation by the proposed method. Residual grid line artifacts (GLAs) in projection views and reconstructed DBT images are characterized and methods for reducing the visibility of GLAs in the reconstructed volume through projection image flat-field correction and spatial frequency-based filtering of the DBT slices are described and evaluated. The software correction methods reduce the visibility of these artifacts in the reconstructed volume, making them imperceptible both in the reconstructed DBT images and their Fourier transforms.

Entities:  

Keywords:  anti-scatter grid; digital tomosynthesis; image filtering; power spectrum; reciprocation

Year:  2016        PMID: 28855746      PMCID: PMC5573244          DOI: 10.1117/12.2216248

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  13 in total

1.  Noise power spectra of images from digital mammography detectors.

Authors:  M B Williams; P A Mangiafico; P U Simoni
Journal:  Med Phys       Date:  1999-07       Impact factor: 4.071

2.  Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.

Authors:  Kenneth A Fetterly; Beth A Schueler
Journal:  Phys Med Biol       Date:  2007-07-30       Impact factor: 3.609

3.  A study of grid artifacts formation and elimination in computed radiographic images.

Authors:  Chih-Yang Lin; Wen-Jeng Lee; Shyh-Jye Chen; Ching-Hwa Tsai; Jei-Han Lee; Chia-Hung Chang; Yu-Tai Ching
Journal:  J Digit Imaging       Date:  2006-12       Impact factor: 4.056

4.  A novel technique to suppress grid line artifacts.

Authors:  David M Gauntt; Gary T Barnes
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

5.  Grid line artifact formation: A comprehensive theory.

Authors:  David M Gauntt; Gary T Barnes
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

6.  Calculation of dose and contrast for two mammographic grids.

Authors:  D R Dance; J Persliden; G A Carlsson
Journal:  Phys Med Biol       Date:  1992-01       Impact factor: 3.609

7.  The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si/CsI(TI) full-field flat panel detector.

Authors:  Ann-Katherine Carton; Raymond Acciavatti; Johnny Kuo; Andrew D A Maidment
Journal:  Med Phys       Date:  2009-03       Impact factor: 4.071

8.  Grid artifact reduction for direct digital radiography detectors based on rotated stationary grids with homomorphic filtering.

Authors:  Dong Sik Kim; Sanggyun Lee
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

9.  Development and Monte Carlo analysis of antiscatter grids for mammography.

Authors:  John M Boone; Olga V Makarova; Vladislav N Zyryanov; Cha-Mei Tang; Derrick C Mancini; Nikolaie Moldovan; Ralu Divan
Journal:  Technol Cancer Res Treat       Date:  2002-12

10.  Grid and slot scan scatter reduction in mammography: comparison by using Monte Carlo techniques.

Authors:  John M Boone; J Anthony Seibert; Cha-Mei Tang; Steve M Lane
Journal:  Radiology       Date:  2002-02       Impact factor: 11.105

View more
  1 in total

1.  A novel total variation based ring artifact suppression method for CBCT imaging with two-dimensional antiscatter grids.

Authors:  Timur Alexeev; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2019-03-20       Impact factor: 4.071

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.