Renée Keijsers1,2, Elisa L Zwerus1,2,3, Dagmar R M van Lith4, Koen L M Koenraadt5, Pjotr Goossens4, Bertram The1, Michel P J van den Bekerom3, Denise Eygendaal1,2.
Abstract
INTRODUCTION: Range of motion (ROM) is closely monitored before and after surgery for stiff elbow and during rehabilitation. Measurements in the home environment may be helpful to increase involvement and adherence of the patient. Therefore, our objective is to investigate the validity and inter- and intraobserver reliability of 3 alternative methods to assess the ROM by the patient in a home-based situation, in comparison to the universal goniometer (UG). We hypothesize that all 3 alternative methods will be valid alternatives and show a level of reliability equivalent to UG.Entities:
Year: 2018 PMID: 30631776 PMCID: PMC6304590 DOI: 10.1155/2018/7906875
Source DB: PubMed Journal: J Sports Med (Hindawi Publ Corp) ISSN: 2314-6176
Figure 1Validity of different measurement methods compared to UG measurements (ICC) and mean difference (∆) and 95% confidence intervals (CI) (in degrees).
| UG vs. |
|
|
| |
|---|---|---|---|---|
|
| ICC (95% CI) | 0.71 (0.51-0.83) | 0.63 (0.41-0.79) | 0.57 (0.32-0.75) |
|
| ||||
|
| ICC (95% CI) | 0.76 (0.58-0.87) | 0.78 (0.63-0.88) | 0.28 (-0.05-0.55) |
|
| ||||
|
| ICC (95% CI) | 0.44 (0.15-0.66) | 0.45 (0.17-0.67) | 0.67 (0.47-0.82) |
|
| ||||
|
| ICC (95% CI) | 0.50 (0.23-0.70) | 0.47 (0.18-0.68) | 0.61 (0.37-0.77) |
Interobserver reliability (ICC) and mean difference (∆) and 95% CI (in degrees) of UG, photography, and movie and smartphone application measurements.
|
|
|
|
| ||
|---|---|---|---|---|---|
|
| ICC (95% CI) | 0.41 (0.07-0.65) | 0.83 (0.65-0.92) | 0.86 (0.75-0.92) | 0.66 (0.45-0.81) |
|
| |||||
|
| ICC (95% CI) | 0.65 (0.43-0.80) | 0.93 (0.88-0.96) | 0.88 (0.77-0.93) | 0.56 (0.31-0.74) |
|
| |||||
|
| ICC (95% CI) | 0.40 (0.11-0.63) | 0.90 (0.76-0.95) | 0.82 (0.56-0.91) | 0.55 (0.29-0.73) |
|
| |||||
|
| ICC (95% CI) | 0.71 (0.51-0.83) | 0.89 (0.77-0.94) | 0.96 (0.93-0.98) | 0.48 (0.20-0.67) |
Intraobserver reliability (ICC) and mean difference (∆) and 95% CI of UG, photography, and movie and smartphone application measurements.
|
|
|
|
| ||
|---|---|---|---|---|---|
|
| ICC (95% CI) | 0.50 (0.31-0.67) | 0.87 (0.81-0.92) | 0.94(0.89-0.97) | 0.60 (0.36-0.76) |
|
| |||||
|
| ICC (95% CI) | 0.84 (0.75-0.91) | 0.82 (0.73-0.88) | 0.96(0.93-0.98) | 0.45 (0.16-0.66) |
|
| |||||
|
| ICC (95% CI) | 0.71 (0.57-0.81) | 0.72 (0.59-0.81) | 0.94(0.98-0.97) | 0.58 (0.33-0.75) |
|
| |||||
|
| ICC (95% CI) | 0.47 (0.28-0.65) | 0.71 (0.57-0.81) | 0.95(0.92-0.98) | 0.31 (0.02-0.56) |