Keisuke Ishii1, Hiroyuki Oka2, Yuji Honda3, Daisuke Oguro3, Youichiro Konno4, Kousuke Kumeta4, Shouta Nishihara4, Hazuki Matsuyama4, Ichiro Kaneko5, Yasuo Takeuchi6, Yoshinobu Watanabe1, Hirotaka Kawano7, Naoshi Ogata3.
Abstract
[Purpose] Recently, a photo-based smartphone application for angle measurement-"Grid line imaging application Professional"-was developed to evaluate joint disease treatments. The aim of this study was to determine the accuracy and reliability of the application. [Participants and Methods] We measured the knee joint of a mannequin using an application and a universal goniometer. Twelve examiners measured eight knee joints of mannequins at different arbitrary angles using the application and a universal goniometer. Correlations between the application and universal goniometer measurements were examined using scatter plots and correlation coefficients. Systematic errors of the application were visually confirmed using the Bland-Altman method. Intra-class correlation coefficients were used to evaluate the inter-examiner reliability of the application.Entities:
Keywords: Angle measurement; Reliability; Smartphone application
Year: 2021 PMID: 34083881 PMCID: PMC8165359 DOI: 10.1589/jpts.33.417
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Fig. 1.Measurement of knee joint angle using a smartphone application.
The knee joint angle measurements of APP in inter-examiner reproducibility tests
| MAN1 | MAN2 | MAN3 | MAN4 | MAN5 | MAN6 | MAN7 | MAN8 | |
| Average | 2.8 | 24.5 | 36.1 | 46.4 | 54.3 | 58.3 | 70.5 | 94.0 |
| SD | 1.9 | 2.5 | 2.8 | 3.5 | 4.6 | 3.1 | 2.4 | 2.5 |
APP: smartphone application; MAN: knee joint angle of mannequin; SD: standard deviation.
The knee joint angle measurements of UG in inter-examiner reproducibility tests
| MAN1 | MAN2 | MAN3 | MAN4 | MAN5 | MAN6 | MAN7 | MAN8 | |
| Average | 2.2 | 24.8 | 35.5 | 48.0 | 52.5 | 58.9 | 70.6 | 91.7 |
| SD | 1.6 | 2.4 | 4.7 | 2.2 | 2.4 | 3.3 | 3.7 | 3.6 |
UG: universal goniometer; MAN: knee joint angle of mannequin; SD: standard deviation.
Fig. 2.Scatter plot of identical knee angle measurements by the APP and the UG.
APP: smartphone application; UG: universal goniometer.
Fig. 3.Bland-Altman plot of identical knee angle measurements by the UG and the APP.
APP: smartphone application; UG: universal goniometer.
ICC of inter-examiner reliability (single measure value)
| ICC | 95% CI | p value | |
| APP | 0.987 | 0.976 to 0.998 | <0.001 |
| UG | 0.986 | 0.966 to 0.997 | <0.001 |
APP: smartphone application; UG: universal goniometer; ICC: intra-class correlation coefficients; CI: confidence interval.
ICC of inter-examiner reliability (average measured value)
| ICC | 95% CI | p value | |
| APP | 0.999 | 0.998 to 1.000 | <0.001 |
| UG | 0.999 | 0.997 to 1.000 | <0.001 |
APP: smartphone application; UG: universal goniometer; ICC: intra-class correlation coefficients; CI: confidence interval.
The knee joint angle measurements of APP in intra-examiner reproducibility tests
| Tester 1 | Tester 2 | Tester 3 | Tester 4 | Tester 5 | Tester 6 | |||||||
| 1st | 2nd | 1st | 2nd | 1st | 2nd | 1st | 2nd | 1st | 2nd | 1st | 2nd | |
| MAN1 | 13 | 13 | 10 | 9 | 13 | 17 | 14 | 20 | 15 | 10 | 10 | 10 |
| MAN2 | 45 | 48 | 44 | 46 | 48 | 54 | 49 | 54 | 50 | 52 | 50 | 45 |
| MAN3 | 64 | 62 | 60 | 58 | 68 | 65 | 65 | 67 | 65 | 60 | 60 | 60 |
| MAN4 | 38 | 38 | 33 | 39 | 35 | 40 | 45 | 44 | 38 | 43 | 40 | 35 |
| MAN5 | 53 | 53 | 52 | 53 | 57 | 56 | 60 | 59 | 54 | 52 | 55 | 55 |
| MAN6 | 18 | 15 | 19 | 17 | 17 | 24 | 22 | 18 | 14 | 12 | 14 | 20 |
APP: smartphone application; MAN: knee joint angle of mannequin.
ICC of intra-examiner reliability of APP
| ICC | 95% CI | p value | |
| Single measure value | 0.982 | 0.965 to 0.991 | <0.001 |
| Average measured value | 0.991 | 0.982 to 0.995 | <0.001 |
APP: smartphone application; ICC: intra-class correlation coefficients; CI: confidence interval.