S Bergiers1, H S Hothi1, J Henckel1, A Eskelinen2, J Skinner3, A Hart4.
Abstract
OBJECTIVES: Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event.Entities:
Keywords: Material loss; Metal-on-metal; Retrieval; Wear analysis
Year: 2018 PMID: 30581557 PMCID: PMC6269595 DOI: 10.1302/2046-3758.711.BJR-2018-0143.R1
Source DB: PubMed Journal: Bone Joint Res ISSN: 2046-3758 Impact factor: 5.853
Summary of the comparison between patient, implant, and surgical characteristics of the two groups
| Category | Pre-2007 (n = 45) | 2007 onwards (n = 47) | p-value |
|---|---|---|---|
| 0.67[ | |||
| Female | 24 ( | 28 ( | |
| Male | 21 ( | 19 ( | |
| Mean age, yrs ( | 61.3 (6.2) | 61.8 (8.1) | 0.77[ |
| Mean time revision, mths ( | 96 (14) | 50 (20) | < 0.001[ |
| 0.04[ | |||
| Corail | 20 ( | 32 ( | |
| Summit | 25 ( | 15 ( | |
| Mean inclination, ° ( | 47.6 (8.1) | 43.5 (7.6) | 0.02[ |
| Mean horizontal offset, mm ( | 44.8 (6.0) | 43.0 (7.4) | 0.23[ |
| Mean vertical offset, mm ( | 79.6 (8.8) | 75.0 (11.1) | 0.04[ |
| 0.08[ | |||
| No | 11 ( | 20 ( | |
| Yes | 34 ( | 27 ( |
Fisher’s exact test
Unpaired t-test
Statistically significant
Comparison of the blood metal ion levels of the two groups
| Parameter | Pre-2007 | 2007 onwards | p-value[ |
|---|---|---|---|
| Median cobalt (Co), ppb (range) | 7.70 (1.00 to 101.60) | 7.23 (0.60 to 130.00) | 0.4288 |
| Median chromium (Cr), ppb (range) | 2.50 (0.50 to 59.60) | 5.70 (0.56 to 90.00) | 0.4182 |
| Median Co/Cr ratio (range) | 2.71 (0.83 to 10.20) | 1.42 (0.66 to 5.98) | 0.0006[ |
Unpaired t-test
Statistically significant
Comparison of the bearing and taper wear rates of the two study groups, reporting the median and interquartile range (IQR) for each group
| Outcome | Pre-2007 | 2007 onwards |
|---|---|---|
| Median bearing wear rate, mm3/yr (IQR) | 2.3 (1.6 to 9.7) | 3.8 (2.8 to 8.1) |
| Median taper wear rate, mm3/yr (IQR) | 0.81 (0.10 to 1.64) | 0.19 (0.03 to 0.75) |
Fig. 1Graph showing the comparison of the bearing wear rate of the ‘pre-2007’ and ‘2007 onwards’ groups.
Fig. 2Graph showing the comparison of the taper wear rate of the ‘pre-2007’ and ‘2007 onwards’ groups.
Summary of the regression analyses performed to compare the study outcomes between the two groups, both before and after adjusting for potentially confounding variables
| Adjustments | Ratio (95% CI)[ | p-value[ |
|---|---|---|
| None | 1.22 (0.77 to 1.92) | 0.39 |
| Time to revision, design, edge wear | 0.97 (0.48 to 1.98) | 0.94 |
| + Inclination, vertical offset | 0.84 (0.38 to 1.87) | 0.67 |
| None | 0.52 (0.33 to 0.82) | 0.005[ |
| Time revision, design, edge wear | 0.72 (0.34 to 1.49) | 0.37 |
| + Inclination, vertical offset | 0.70 (0.31 to 1.59) | 0.39 |
The differences in wear rate between groups are reported as ratios, due to the log transformation of the outcomes, specifically, the wear rate from 2007 onwards relative to the wear rate in the pre-2007 period
Regression analysis
Statistically significant
CI, confidence interval
Fig. 3Box plots representing the time to revision for both groups of implants.
Fig. 4Bar chart plotting the year of primary implantation against the mean time to revision.