Literature DB >> 23918742

Material loss at the taper junction of retrieved large head metal-on-metal total hip replacements.

Ashley K Matthies1, Radu Racasan, Paul Bills, Liam Blunt, Suzie Cro, Anna Panagiotidou, Gordon Blunn, John Skinner, Alister J Hart.   

Abstract

It has been speculated that material loss, either as corrosion or wear, at the head-stem taper junction is implicated in the high revision rates reported for metal-on-metal total hip replacements. We measured the volume of material loss from the taper and bearing surfaces of retrieved devices, and investigated the associations with blood metal ion levels and the diagnosis of a cystic or solid pseudotumor. The median volumes of material lost from the female and male taper surfaces were 2.0 and 0.29 mm(3) , respectively, while the median volumes of wear from the cup and head bearing surfaces were 1.94 and 3.44 mm(3) , respectively. Material loss from the female taper was similar to that from the acetabular bearing surface (p = 0.55), but significantly less than that from the femoral bearing surface (p < 0.001). Material loss from the male taper was less than that from both bearing surfaces (p < 0.001). Multivariable analysis demonstrated no significant correlations between the volume of material lost from the taper surfaces and either blood cobalt or chromium ions, or the presence of pseudotumor. While a substantial volume of material is lost at the taper junction, the clinical significance of this debris remains unclear.
© 2013 Orthopaedic Research Society.

Entities:  

Keywords:  metal ions; metal-on-metal; retrievals; taper; wear

Mesh:

Substances:

Year:  2013        PMID: 23918742     DOI: 10.1002/jor.22431

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  37 in total

Review 1.  What is the trouble with trunnions?

Authors:  Christina I Esposito; Timothy M Wright; Stuart B Goodman; Daniel J Berry
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

Review 2.  Metallic debris from metal-on-metal total hip arthroplasty regulates periprosthetic tissues.

Authors:  Christoph H Lohmann; Gurpal Singh; Hans-Georg Willert; Gottfried H Buchhorn
Journal:  World J Orthop       Date:  2014-11-18

3.  Repeated metal ion measurements in patients with high risk metal-on-metal hip replacement.

Authors:  Aleksi Reito; Teemu Moilanen; Timo Puolakka; Jorma Pajamäki; Antti Eskelinen
Journal:  Int Orthop       Date:  2014-03-18       Impact factor: 3.075

4.  Is there material loss at the backside taper in modular CoCr acetabular liners?

Authors:  Matthias T Agne; Richard J Underwood; Sevi B Kocagoz; Daniel W MacDonald; Judd S Day; Javad Parvizi; Matthew J Kraay; Michael A Mont; Gregg R Klein; Harold E Cates; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2014-10-16       Impact factor: 4.176

Review 5.  Wear of dual-mobility cups: a review article.

Authors:  Anna Di Laura; Harry Hothi; Clement Battisti; Arianna Cerquiglini; Johann Henckel; John Skinner; Alister Hart
Journal:  Int Orthop       Date:  2016-12-09       Impact factor: 3.075

6.  Does Taper Angle Clearance Influence Fretting and Corrosion Damage at the Head-Stem Interface? A Matched Cohort Retrieval Study.

Authors:  Sevi B Kocagöz; Richard J Underwood; Shiril Sivan; Jeremy L Gilbert; Daniel W Macdonald; Judd S Day; Steven M Kurtz
Journal:  Semin Arthroplasty       Date:  2013-12-01

7.  Comparison of metal ion levels in patients with hip resurfacing versus total hip arthroplasty.

Authors:  Craig W Forsthoefel; Nicholas M Brown; Mark L Barba
Journal:  J Orthop       Date:  2017-08-31

8.  Five Hundred Fifty-five Retrieved Metal-on-metal Hip Replacements of a Single Design Show a Wide Range of Wear, Surface Features, and Histopathologic Reactions.

Authors:  Sang-Hyun Park; Zhen Lu; Robert S Hastings; Patricia A Campbell; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

Review 9.  Do retrieval analysis and blood metal measurements contribute to our understanding of adverse local tissue reactions?

Authors:  Patricia A Campbell; Michael S Kung; Andrew R Hsu; Joshua J Jacobs
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

10.  Trunnionosis: A pain in the neck.

Authors:  Philip S Pastides; Matthew Dodd; Khaled M Sarraf; Charles A Willis-Owen
Journal:  World J Orthop       Date:  2013-10-18
View more