Literature DB >> 21969427

Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement.

A J Hart1, S A Sabah, A S Bandi, P Maggiore, P Tarassoli, B Sampson, J A Skinner.   

Abstract

Blood metal ions have been widely used to investigate metal-on-metal hip replacements, but their ability to discriminate between well-functioning and failed hips is not known. The Medicines and Healthcare products Regulatory Agency (MHRA) has suggested a cut-off level of 7 parts per billion (ppb). We performed a pair-matched, case-control study to investigate the sensitivity and specificity of blood metal ion levels for diagnosing failure in 176 patients with a unilateral metal-on-metal hip replacement. We recruited 88 cases with a pre-revision, unexplained failed hip and an equal number of matching controls with a well-functioning hip. We investigated the 7 ppb cut-off level for the maximum of cobalt or chromium and determined optimal mathematical cut-off levels from receiver-operating characteristic curves. The 7 ppb cut-off level for the maximum of cobalt or chromium had a specificity of 89% and sensitivity 52% for detecting a pre-operative unexplained failed metal on metal hip replacement. The optimal cut-off level for the maximum of cobalt or chromium was 4.97 ppb and had sensitivity 63% and specificity 86%. Blood metal ions had good discriminant ability to separate failed from well-functioning hip replacements. The MHRA cut-off level of 7 ppb provides a specific test but has poor sensitivity.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21969427     DOI: 10.1302/0301-620X.93B10.26249

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  72 in total

1.  Comparison between component designs with different femoral head size in metal-on-metal total hip arthroplasty; multicenter randomized prospective study.

Authors:  Wataru Ando; Kengo Yamamoto; Takashi Atsumi; Satoshi Tamaoki; Kazuhiro Oinuma; Hideaki Shiratsuchi; Hirohiko Tokunaga; Yutaka Inaba; Naomi Kobayashi; Masaharu Aihara; Kenji Ohzono
Journal:  J Orthop       Date:  2015-06-10

Review 2.  Imaging near orthopedic hardware.

Authors:  Matthew F Koff; Alissa J Burge; Kevin M Koch; Hollis G Potter
Journal:  J Magn Reson Imaging       Date:  2017-02-02       Impact factor: 4.813

Review 3.  The surgical options and clinical evidence for treatment of wear or corrosion occurring with THA or TKA.

Authors:  Charles A Engh; Henry Ho; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

4.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

5.  Repeated metal ion measurements in patients with high risk metal-on-metal hip replacement.

Authors:  Aleksi Reito; Teemu Moilanen; Timo Puolakka; Jorma Pajamäki; Antti Eskelinen
Journal:  Int Orthop       Date:  2014-03-18       Impact factor: 3.075

6.  Outcome of Birmingham hip resurfacing at ten years: role of routine whole blood metal ion measurements in screening for pseudotumours.

Authors:  Aleksi Reito; Timo Puolakka; Petra Elo; Jorma Pajamäki; Antti Eskelinen
Journal:  Int Orthop       Date:  2014-07-17       Impact factor: 3.075

7.  Pleomorphic clinical spectrum of metallosis in total hip arthroplasty.

Authors:  Francesco Pisanu; Carlo Doria; Matteo Andreozzi; Marco Bartoli; Laura Saderi; Giovanni Sotgiu; Paolo Tranquilli Leali
Journal:  Int Orthop       Date:  2018-09-29       Impact factor: 3.075

8.  Metal-on-Metal Hip Joint Prostheses: a Retrospective Case Series Investigating the Association of Systemic Toxicity with Serum Cobalt and Chromium Concentrations.

Authors:  James H Ho; Jerrold B Leikin; Paul I Dargan; John R H Archer; David M Wood; Jeffrey Brent
Journal:  J Med Toxicol       Date:  2017-09-12

Review 9.  Management of metal-on-metal hip implant patients: Who, when and how to revise?

Authors:  Reshid Berber; John A Skinner; Alister J Hart
Journal:  World J Orthop       Date:  2016-05-18

10.  601 metal-on-metal total hip replacements with 36 mm heads a 5 minimum year follow up: Levels of ARMD remain low despite a comprehensive screening program.

Authors:  Amit Atrey; Alister Hart; Nasir Hussain; Jonathon Waite; Andrew J Shepherd; Steve Young
Journal:  J Orthop       Date:  2016-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.