Claude B Rieker1, Rolf Schön, Reto Konrad, Gernot Liebentritt, Patric Gnepf, Ming Shen, Paul Roberts, Peter Grigoris.
Abstract
Large-diameter metal-on-metal articulations may provide an opportunity for wear reduction in total hip implants because earlier studies have shown that the formation of a fluid film that completely separates the bearing surfaces is theoretically possible. In such a lubrication mode and under ideal conditions, there is theoretically no amount of wear. Studies have suggested that the two primary parameters controlling the lubrication mode are the diameter and the clearance of the articulation. The goal of the present study was to experimentally investigate the influence of these two parameters on the wear behavior of large-diameter metal-on-metal articulations pertaining to resurfacing hip implants. The results of this in vitro investigation showed that longer running-in periods and higher amounts of running-in wear were associated with larger clearances.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15833451 DOI: 10.1016/j.ocl.2005.02.004
Source DB: PubMed Journal: Orthop Clin North Am ISSN: 0030-5898 Impact factor: 2.472