Literature DB >> 30499058

Measurement of Upper Limb Range of Motion Using Wearable Sensors: A Systematic Review.

Corrin P Walmsley1, Sîan A Williams2,3, Tiffany Grisbrook2, Catherine Elliott1,4, Christine Imms5, Amity Campbell2.   

Abstract

BACKGROUND: Wearable sensors are portable measurement tools that are becoming increasingly popular for the measurement of joint angle in the upper limb. With many brands emerging on the market, each with variations in hardware and protocols, evidence to inform selection and application is needed. Therefore, the objectives of this review were related to the use of wearable sensors to calculate upper limb joint angle. We aimed to describe (i) the characteristics of commercial and custom wearable sensors, (ii) the populations for whom researchers have adopted wearable sensors, and (iii) their established psychometric properties.
METHODS: A systematic review of literature was undertaken using the following data bases: MEDLINE, EMBASE, CINAHL, Web of Science, SPORTDiscus, IEEE, and Scopus. Studies were eligible if they met the following criteria: (i) involved humans and/or robotic devices, (ii) involved the application or simulation of wearable sensors on the upper limb, and (iii) calculated a joint angle.
RESULTS: Of 2191 records identified, 66 met the inclusion criteria. Eight studies compared wearable sensors to a robotic device and 22 studies compared to a motion analysis system. Commercial (n = 13) and custom (n = 7) wearable sensors were identified, each with variations in placement, calibration methods, and fusion algorithms, which were demonstrated to influence accuracy.
CONCLUSION: Wearable sensors have potential as viable instruments for measurement of joint angle in the upper limb during active movement. Currently, customised application (i.e. calibration and angle calculation methods) is required to achieve sufficient accuracy (error <  5°). Additional research and standardisation is required to guide clinical application. TRIAL REGISTRATION: This systematic review was registered with PROSPERO ( CRD42017059935 ).

Entities:  

Keywords:  Inertial movement unit; Joint angle; Kinematics; Motion analysis; Upper limb; Wearable sensor

Year:  2018        PMID: 30499058      PMCID: PMC6265374          DOI: 10.1186/s40798-018-0167-7

Source DB:  PubMed          Journal:  Sports Med Open        ISSN: 2198-9761


Key Points

Both commercially available and custom wearable sensors have some evidence of validity in the literature. Although commercial wearable sensors were validated against pseudo gold standards, each study customised the commercial software to do so. Wearable sensors demonstrated errors < 5° for all degrees of freedom at the wrist and elbow joints when compared to a robotic device. The range in error is greater when measured in vivo and compared to a pseudo gold standard. The measured errors are within margins that warrant future use of wearable sensors to measure joint angle in the upper limb.

Background

Clinicians and researchers seek information about the quality and quantity of patients’ movement as it provides useful information to guide and evaluate intervention. Range of motion (ROM), defined as rotation about a joint, is measured in a variety of clinical populations including those with orthopaedic, musculoskeletal, and neurological disorders. Measurement of ROM forms a valuable part of clinical assessment; therefore, it is essential that it is completed in a way that provides accurate and reliable results [1, 2]. In clinical practice, the goniometer is a widely used instrument to measure ROM [2-4]. Despite being considered a simple, versatile, and an easy-to-use instrument, reports of reliability and accuracy are varied. Intra-class correlation coefficients (ICCs) range from 0.76 to 0.94 (intra-rater) [3, 4] and 0.36 to 0.91 (inter-rater) [4] for shoulder and elbow ROM. Low inter-rater reliability is thought to result from the complexity and characteristics of the movement, the anatomical joint being measured, and the level of assessor experience [5, 6]. The goniometer is also limited to measuring joint angles in single planes and static positions; thus, critical information regarding joint angles during dynamic movement cannot be measured. In research settings, three-dimensional motion analysis (3DMA) systems, such as Vicon (Vicon Motion Systems Ltd., Oxford, UK) and Optitrack (NaturalPoint, Inc., Corvallis, OR, USA), are used to measure joint angles during dynamic movement in multiple degrees of freedom (DOF). Such systems are considered the ‘gold standard’ for evaluating lower limb kinematics, with a systematic review reporting errors < 4.0° for movement in the sagittal plane and < 2.0° in the coronal plane; higher values have been reported for hip rotation in the transverse plane (range 16 to 34°) [7]. Measurement in the upper limb is considered more technically challenging due to the complexity of shoulder, elbow, and wrist movements [8]. However, given the demonstrated accuracy in the lower limb, 3DMA systems are used as the ‘ground truth’ when validating new upper limb measurement tools [9]. However, 3DMA does have limitations. Most notably, these systems are typically immobile, expensive, require considerable expertise to operate, and therefore rarely viable for use with clinical populations [10, 11]. Wearable sensors, or inertial measurement units, are becoming increasingly popular for the measurement of joint angle in the upper limb [12]. In this review, we were interested in wearable sensors that contained accelerometers and gyroscopes, with or without a magnetometer, to indirectly derive orientation. The software typically utilised three main steps: (i) calibration, using two approaches: (1) system, also referred to as ‘factory calibration’ (offset of the hardware on a flat surface), and (2) anatomical calibration including both static (pre-determined pose) and dynamic (pre-determined movement) [10, 13]; (ii) filtering, using fusion algorithms including variations of the Kalman filter (KF) [14, 15]; and (iii) segment and angle definition, using Euler angle decompositions and/or Denavit-Hartenberg Cartesian coordinates. Wearable sensors are an increasingly popular surrogate for laboratory-based 3DMA due to their usability, portability, size, and cost. Systematic reviews have detailed their use during swimming [16] and whole body analysis [17] and in the detection of gait parameters and lower limb biomechanics [18]. However, their validity and reliability must be established and acceptable prior to their application [19]. Accuracy of the wearable sensors is dependent on the joint and movement being measured; therefore, a systematic review specific to the upper limb is required. This study aimed to establish the evidence for the use of wearable sensors to calculate joint angle in the upper limb, specifically: What are the characteristics of commercially available and custom designed wearable sensors? What populations are researchers applying wearable sensors for and how have they been used? What are the established psychometric properties for the wearable sensors?

Methods

This systematic review was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines [20] and registered with the International Prospective Register of Systematic Reviews on 23 March 2017 (CRD42017059935).

Search Terms and Data Bases

Studies and conference proceedings were identified through searches in scientific data bases relevant to the fields of biomechanics, medicine, and engineering, from their earliest records to November 1, 2016: MEDLINE via PROQUEST, EMBASE via OVID, CINAHL via EBSCO, Web of Science, SPORTDiscus, IEEE, and Scopus. Reference lists were searched to ensure additional relevant studies were identified. The search was updated on 9 October 2017 to identify new studies that met the inclusion criteria. The following search term combinations were used: (“wearable sens*”OR “inertial motion unit*” OR “inertial movement unit*” OR “inertial sens*” OR sensor) AND (“movement* analysis” OR “motion analysis*” OR “motion track*” OR “track* motion*” OR “measurement system*” OR movement) AND (“joint angle*” OR angle* OR kinematic* OR “range of motion*”) AND (“upper limb*” OR “upper extremit*” OR arm* OR elbow* OR wrist* OR shoulder* OR humerus*). Relevant MeSH terms were included where appropriate, and searches were limited to title, abstract, and key words. All references were imported into Endnote X6 (Thomson Reuters, Carlsbad, CA, USA), and duplicates were removed.

Study Selection Criteria and Data Extraction

The title and abstracts were screened independently by two reviewers (CW and AC). Full texts were retrieved if they met the inclusion criteria: (i) included human participants and/or robotic devices, (ii) applied/simulated use of wearable sensors on the upper limb, and (iii) calculated an upper limb joint angle. The manuals of commercial wearable sensors were located, with information extracted when characteristics were not reported by study authors. Studies were excluded based on the following criteria: (i) used a single wearable sensor, (ii) included different motion analysis systems (i.e. WiiMove, Kinetic, and smart phones), (iii) used only an accelerometer, (iv) calculated segment angle or position, (v) studied the scapula, or (vi) were not published in English. Two reviewers (CW and AC) extracted data independently to a customised extraction form. Discrepancies were discussed, and a third reviewer (TG) was involved when consensus was not reached. Extracted parameters of the wearable sensor characteristics included custom and commercial brands, the dimensions (i.e. height and weight), components used (i.e. accelerometer, gyroscope, and magnetometer), and the sampling rate (measured in hertz (Hz)). Sample characteristics included the number of participants, their age, and any known clinical pathology. To determine if authors of the included studies customised aspects of the wearable sensors system, the following parameters were extracted: the type of calibration (i.e. system and anatomical), the fusion algorithms utilised, how anatomical segments were defined, and how joint angle was calculated. To understand the validity and reliability of the wearable sensors, information about the comparison system, marker placement, and psychometric properties were extracted. The mean error, standard deviation (SD), and root mean square error (RMSE) reported in degrees were extracted where possible from the validation studies. The RMSE represents the error or difference between the wearable sensor and the comparison system (e.g. 3DMA system). The larger the RMSE, the greater the difference (in degrees) between the two systems. Further, to report on the validity of the wearable sensors, studies that did not delineate error between the wearable sensor and soft tissue artefact (movement of the markers with the skin) by not using the same segment tracking were not further analysed. Reliability was assessed using ICCs, with values < 0.60 reflecting poor agreement, 0.60–0.79 reflecting adequate agreement, and 0.80–1.00 reflecting excellent agreement [21]. The following parameters were used to guide the interpretation of measurement error, with < 2.0° considered acceptable, between 2.0 and 5.0° regarded as reasonable but may require consideration when interpreting the data, and > 5.0° of error was interpreted with caution [7].

Assessment of Risk of Bias and Level of Evidence

Due to the variability between research disciplines (i.e. health and engineering) in the way that studies were reported, and the level of detail provided about the research procedures, the available assessments of risk of bias and levels of evidence were not suitable for this review. Therefore, the following criteria were used to evaluate the quality of the reporting in the included studies: The aim of the study was clear and corresponded to the results that were reported. The study design and type of paper (i.e. conference proceeding) were considered. Number of participants included in the study was considered in relation to the COSMIN guidelines which indicate that adequate samples require 50–99 participants [19].

Results

The initial search (2016) identified 1759 studies eligible for inclusion, with an additional 432 studies identified 12 months later (2017). A total of 66 studies met the inclusion criteria (Fig. 1). Eight studies reported on the  validation against a robotic device, and 22 reported on validation against a motion analysis system with human participants. One study assessed the reliability of the wearable sensors, with the remaining 35 studies using wearable sensors as an outcome measure in an experimental design.
Fig. 1

A PRISMA diagram of the search strategy

A PRISMA diagram of the search strategy

Characteristics and Placement of the Wearable Sensors

The characteristics of the wearable sensors are summarised in Table 1. A total of seven customised wearable sensors and 13 commercial brands were identified. The level of detail provided for the placement of the wearable sensors on the upper limb varied significantly, as did the mode of attachment (Table 1).
Table 1

Summary of the descriptive characteristics of the wearable sensors

StudyBrandNo. of sensors usedDimensions (mm)L × W × HWeight (grams)WirelessComponentsSample rate (Hz)Method of attachmentParticipants
First authorConference/full textPopulation N Mean age ± SD (years)
AccGyrMag
Muller et al. [22]FullXsens—MTw Awinda247 × 30 × 13*16*Y*DS tapeHealthy125
Bouvier et al. [23]FullXsens—MTw434.5 × 57.8 × 14.527Y60DS tape and elasticHealthy1029 ± 3.4
Robert-Lachaine et al. [24]FullXsens—MVN1750*N30VelcroHealthy1226.3 ± 4.4
Robert-Lachaine et al. [25]FullXsens—MVN1750*N30VelcroHealthy1226.3 ± 4.4
Eckardt et al. [26]FullXsens—MVN1750*N120Body suitHealthy2020.2 ± 5.7
Eckardt et al. [27]FullXsens—MVN1750*N120Body suitHealthy1023.4 ± 5.3
Alvarez et al. [28]FullXsens—MTx438 × 53 × 21*30*N50Velcro and elasticRobot and healthy1
Quinones et al. [29]ConXsens—MTx738 × 53 × 21*30*N50SCI1537.4 ± 7.3
Gil-Agudo et al. [30]FullXsens—MTx538 × 53 × 21*30*N25Healthy130
Alvarez et al. [31]FullXsens—MTx440 × 55 × 2230*50ElasticRobot and healthy2
Bai et al. [32]ConXsens—MTx338 × 53 × 20.930N100
Bai et al. [33]ConXsens—MTx238 × 53 × 21*30*120VelcroHealthy1
Zhang et al. [34]FullXsens—MTx338 × 53 × 21*30*100Healthy4
Rodriques-Anglese et al. [35]ConXsens—MTx238 × 53 × 21*30*N100Robot and healthy1
Cutti et al. [36]FullXsens—MT9B439 × 54 × 2838N100DS tape and elasticHealthy123
Zhou et al. [37]FullXsens—MT9B2N25VelcroHealthy420–40
Zhou et al. [38]FullXsens—MT9B2N25Healthy1
Perez et al. [39]FullXsens—MTi458 × 58 × 22*5050FabricHealthy1
Miezal et al. [15]FullXsens3120Healthy130
Miguel-Andres et al. [40]FullXsens3N75Velcro and DS tapeHealthy1029.3 ± 2.21
Luinge et al. [41]FullXsens2NDS tape and leukoplastHealthy1
Morrow et al. [42]FullADPM Opal643.7 × 39.7 × 13.7*< 25*Y80StrapSurgeons645 ± 7
Rose et al. [43]FullADPM Opal643.7 × 39.7 × 13.7*< 25*Y128StrapSurgeons14
Bertrand et al. [44]ConADPM Opal348 × 36 × 13< 22YDS tapeAstronauts2
Fantozzi et al. [45]FullADPM Opal743.7 × 39.7 × 13.7*< 25*Y128VelcroSwimmers826.1 ± 3.4
Kirking et al. [46]FullADPM Opal343.7 × 39.7 × 13.7*22DS tape and strapHealthy5
Ricci et al. [47]FullADPM Opal643.7 × 39.7 × 13.7*< 25*Y128VelcroRobot
El-Gohary et al. [48]FullADPM Opal343.7 × 39.7 × 13.7*< 25a128VelcroRobot
Ricci et al. [49]ConADPM Opal543.7 × 39.7 × 13.7*< 22Y128VelcroHealthy4 and 47 ± 0.3 and 27 ± 1.9
El-Gohary et al. [50]FullADPM Opal243.7 × 39.7 × 13.7*< 25*128^VelcroHealthy8
El-Gohary et al. [51]ConADPM Opal243.7 × 39.7 × 13.7*< 25*YStrapHealthy1
Mazomenos et al. [52]FullShimmer 2r2Y50Custom holders and elasticHealthy and stoke18 and 425–50 and 45–73
Tran et al. [53]ConShimmer 2r2Y18StrapHealthy1
Daunoravicene et al. [54]FullShimmer351.2StrapStroke1460.8 ± 12.5
Bertomu-Motos et al. [55]FullShimmer251 × 34 × 14*YStrapHealthy4 and 5021–51 and 20–72
Meng et al. [56]ConShimmer251 × 34 × 14*Y20VelcroSpherical coordinate system and healthy1
Peppoloni et al. [57]ConShimmer351 × 34 × 14*Y100VelcroHealthy1
Ruiz-Olaya et al. [58]FullInvenSenseMPU9150 chip2N50StrapsHealthy3
Callejas –Curervo et al. [59]FullInvenSenseMPU9150 chip2N30DS tapeRobot and healthy3
Li et al. [60]FullInvenSense MPU9150 chip2NStroke and Healthy35 and 11
Gao et al. [61]ConInvenSenseMPU9150 chip226.2 × 39.2 × 14.8YHealthy125
Lambretcht et al. [62]FullInvenSenseMPU9150 chip412 × 12 × 6N50Healthy1
Peppoloni et al. [63]ConInvenSenseMPU9150 chip4VelcroHealthy1
Eom et al. [64]FullInvenSenseMPU6050 chip2YStrapsRobot and goniometer
Roldan-Jimenez et al. [65]FullInterSense InertiaCube3326.2 × 39.2 × 14.817NDS tape and elastic cohesive bandageHealthy1518–35
Roldan-Jimenez et al. [66]FullInterSenseInertiaCube3426.2 × 39.2 × 14.817N1000DS tape and elastic cohesive bandageHealthy1124.7 ± 4.2
Nguyen et al. [67]ConBioKin WMS2Y200StrapsHealthy1520–60
Karunarathne et al. [68]ConBioKin WMS2YStrapsHealthy4
Ligorio et al. [69]FullYEI Technology2N220VelcroHealthy1528 ± 3
Vignais et al. [70]FullCAPTIV Motion560 × 35 × 1932Ya64StrapsHealthy541.2 ± 11
Chen et al. [71]ConL-P Research Motion Sensor B2839 × 39 × 8*12YGoniometer
Matsumoto et al. [72]FullNoraxon Myomotion1337.6 × 52 × 18.1< 34200Healthy and stoke10 and 132.2 ± 9.3 and 27
Schiefer et al. [73]FullCUELA1350VelcroHealthy2037.4 ± 9.9
Balbinot et al. [74]FullArduMuV3 chip9Y20Straps
Huang et al. [75]FullMSULS430 × 35 × 1250FabricHealthy and stoke11 and 2253 ± 8 and 62 ± 10
Salam et al. [76]FullCustom344.45 × 44.45Y150Cricketers10
Chang et al. [77]FullCustom2NRobot
Borbely et al. [78]ConCustom2N200Velcro1
Kumar et al. [79]FullCustom1466.6 × 28.2 × 18.1*22*Y*25Custom holders and VelcroHealthy and un-healthy19 and 1924.6 ± 6.7 and 68.4 ± 8.9
Lee et al. [80]FullCustom766.6 × 28.2 × 18.122Y25StrapsGoniometer and stroke568
Cifuentes et al. [81]ConCustom243 × 6060StrapsHealthy9
Kanjanapas et al. [82]FullCustom2N100OrthosisHealthy125
Zhang et al. [83]Con2YHealthy1
Lin et al. [84]Full2YStrapsStroke2552.2 ± 10.2 and 62.2 ± 7.1
El-Gohary et al. [85]Con2
Hyde et al. [86]FullRobot

Table 1 is organised by the brand of the wearable sensor followed by the date that the study was published. This allows direct comparison to be made within the brand of the wearable sensors and trends to be identified between more recently published studies

Abbreviations: Gms grams, Y yes, N no, Acc accelerometer, Gyr gyroscope, Mag magnetometer, Hz hertz (unit of frequency), SD standard deviation, SCI spinal cord injury, PD Parkinson’s disease, Full full text, Con conference paper, mm millimetre, DS double sided

Key:

Wireless—the wearable sensor system was considered wireless if the wearable sensors did not have wires connecting them to an external source, even if that external source was also mounted on the subject

Sample rate—the number of data samples collected per second by the wearable sensor measured in hertz (Hz) which is the unit of frequency

Custom—defined as a newly developed wearable sensor or modifications have occurred to the pre-existing hardware of the wearable sensor

Symbols:

*The information was obtained from the manufacturer procedure manual or other referenced papers

^The sample rate was down sampled (reduced) to allow comparison to the MOCAP system

–Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual

Summary of the descriptive characteristics of the wearable sensors Table 1 is organised by the brand of the wearable sensor followed by the date that the study was published. This allows direct comparison to be made within the brand of the wearable sensors and trends to be identified between more recently published studies Abbreviations: Gms grams, Y yes, N no, Acc accelerometer, Gyr gyroscope, Mag magnetometer, Hz hertz (unit of frequency), SD standard deviation, SCI spinal cord injury, PD Parkinson’s disease, Full full text, Con conference paper, mm millimetre, DS double sided Key: Wireless—the wearable sensor system was considered wireless if the wearable sensors did not have wires connecting them to an external source, even if that external source was also mounted on the subject Sample rate—the number of data samples collected per second by the wearable sensor measured in hertz (Hz) which is the unit of frequency Custom—defined as a newly developed wearable sensor or modifications have occurred to the pre-existing hardware of the wearable sensor Symbols: *The information was obtained from the manufacturer procedure manual or other referenced papers ^The sample rate was down sampled (reduced) to allow comparison to the MOCAP system –Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual

Calibration Methods

Forty-seven studies reported on a calibration procedure prior to data acquisition. System calibration, also commonly known as ‘factory calibration’, was reported on 12 occasions, with two procedures described for the wearable sensors: (i) placement on a flat surface and/or (ii) movement in a pre-determined order while attached to a flat surface [56, 62]. The aim of system calibration was reported to be to align coordinate systems [39, 56] and account for inaccuracies in the orientation of wearable sensor chip relative to its case/packaging [62]. Static anatomical calibration was performed often (n = 34), with dynamic anatomical calibration performed sometimes (n = 10) [23, 30, 36, 41, 45, 49, 57]. Only one study used system calibration alongside both static and dynamic anatomical calibrations to compute joint kinematics [47].

Populations Assessed Using Wearable Sensors

Most studies (n = 52) recruited healthy adults; participants with known pathology were reported in nine studies (Table 1). One study recruited children (< 18 years) [49]. Sample sizes ranged from 1 to 54 participants, with a median sample of 7.6 participants per study. Twenty-nine studies recruited less than five participants, with 20 studies recruiting one single participant.

Psychometric Properties of Wearable Sensors

Validity

Validation studies were split into two categories: (i) studies that compared the wearable sensor output to simulated upper limb movement on a robotic device (Table 2) and (ii) studies that compared wearable sensors output to a 3DMA system on a human participant (Table 3). The term ‘error’ is used to describe the difference between the capture systems; however, we acknowledge that comparisons between the wearable sensors and a robotic device are the only true measures of error.
Table 2

List of the 8 articles organised by first author and containing information related to the validation of wearable sensors for the measurement of joint angle for simulated movements of the upper limb when compared to a robotic device

First authorAim of the studyBrand of wearable sensorsDescription of robotic deviceSensor fusion algorithmCalibrationSegment(s)DOFsSimulated movementsRMSEMean error (SD)
SystemStaticDynamic
Callejas–Cuervo et al. [59]System validationInvensense MPU-9150Industrial robotic arm (ABB IRB 120)KFElbow1DOFFlex/ext2.12–2.44°
Chang et al. [77]System validationCustomRehabotics Medical Technology CorporationFinger1DOFFlex/ext5–7°
Alvarez et al. [28]System validationXsensPan and tilt unit (Model PTU-D46)Wrist2DOFFlexLat dev0.06° (9.20)1.05° (2.18)
Alvarez et al. [31]System validationXsensPan and tilt unit (Model PTU-D46)Wrist2DOFFlexLat dev1.8° for each axis, with a max error ± 6°
Rodriguez-Angleseet et al. [35]System validationXsensPlantar robotKFElbow2DOFDid not report discrete statistics
Kirking et al. [46]Validation/comparison of sensor fusion methodsOpalIndustrial Epson C3 robot armUKFShoulderElbowForearmWrist2DOF1DOF1DOF2DOFInt/ext rotFlex/extFlex/extPro/supFlex/extTwist8.1°2.4°2.6°2.1°2.2°3.9°
Modified UKFShoulderElbowForearmWrist2DOF1DOF1DOF2DOFInt/ext rotFlex/extFlex/extPro/supFlex/extTwist3.0°1.6°2.0°1.2°1.5°2.8°
Ricci et al. [47]Validation/comparison of sensor fusion methodsOpalLWR 4+ (KUKA GmbH)KFShoulderElbowForearmWrist7DOFUnable to determine exact values from box plot
GNFShoulderElbowForearmWrist7DOF
El-Gohary et al. [48]Validation/comparison of sensor fusion methodsOpalNot describedUKFShoulderElbowForearmWrist2DOF1DOF1DOF2DOFIn/ext rotFlex/extFlex/extPro/supFlex/extTwistSlowMedFast
7.8°0.8°0.9°1.3°1.1°1.7°3.0°1.6°2.0°1.2°1.5°2.8°5.9°2.5°2.8°1.1°1.8°2.2°
EKFShoulderElbowForearmWrist2DOF1DOF1DOF2DOFIn/ext rotFlex/extFlex/extPro/supFlex/extTwist8.8°1.2°1.3°0.8°1.2°1.8°8.6°1.9°2.1°1.4°1.9°3.7°9.7°2.5°3.1°1.4°2.9°3.4°

Abbreviations: RMSE root mean square error, SD standard deviation, CMC coefficient of multiple correlation, KBF Kalman-based filter, KF Kalman filter, EKF extended Kalman filter, UKF unscented Kalman filter, WLS weighted least squares, Flex flexion, Ext extension, Pro pronation, Sup supination, Ab abduction, Ad adduction, Dev deviation, Rad radial, Uln ulnar, In internal, Ex external, Rot rotation, Elev elevation, Dep depression, DOF degrees of freedom, C customised, M manufacture

–Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual

Table 3

List of the selected 22 articles organised by first author and containing information related to the validation of wearable sensors for the measurement of joint angle in upper limb when compared to a three-dimensional motion analysis system

First AuthorAim of the studyBrand of SensorsSensor fusion algorithmPlacement of sensorsComparison systemUsed same segment trackingTask(s)Anatomical Segment(s)Degrees of FreedomMovementsMean error (SD)RMSECorrelation coefficientsCalibration
SystemStaticDynamic
Robert Lachaine et al. [24]Validate protocolXsensKFS1: Upper armS2: ForearmS3: HandOptotrakYesElbow flex/ext, pro/sup; wrist flex/ext, ul/rad deviation, rotation and manual handling tasksShoulderElbowWrist3DOF3DOF3DOFFlex/extAb/adRotationFlex/extAb/adPro/supFlex/extRad/ul devRotationOptotrak ISB to Xsens ISB
3.0°2.9°2.5°2.9°2.0°2.6°3.8°2.8°3.6°
Ligorio et al. [69]Validate calibration methodYEI technologyViconYesFlex/ext and pro/supElbow2DOF2DOF 2DOFFlex/extPro/supFlex/extPro/supFlex/extPro/supMethod A
8.5–11.1°11.9–13.3°
Method B
3.4–3.6°6.8–7.6°
Method C – Proposed
3.1–3.3°3.8–4.0°
Fantozzi et al. [45]Validate protocolOpalKBFS1: Flat portion of the sternum.S2: Laterally on the humerus above the centre and posteriorly.S3: Distal forearm above the ulnar and radial styloid.S4: Back of the hand.Stereo-photogrammetric system (SMART-DX 7000)YesSimulated front crawlShoulderElbowWrist3DOF2DOF2DOFFlex/extAb/adIn/ext rotFlex/extPro/supFlex/extRad/ul dev5.0° (4–6)10.0° (7–11)7.0° (5–8)15° (12–17)10.0° (7–11)5.0° (4–5)3.0° (2–4)0.990.970.990.950.930.950.90
Simulated breaststrokeShoulderElbowWrist3DOF2DOF2DOFFlex/extAb/adIn/ext rotFlex/extPro/supFlex/extRad/ul dev5.0° (3–7)3.0° (3–4)8.0° (5–10)6.0° (5–10)5.0° (4–7)4.0° (3–5)-0.990.990.980.970.980.93
Gil-Agudo et al. [30]Validate protocolXsensKFS1: TrunkS2: Back of the headS3: Right armS4: Distal forearmS5: Hand.CODAYesShoulder rot, flex/ext and ab/ad; elbow flex/ext and pro/sup, wrist flex/ext and ul/rad deviation.ShoulderElbowWrist3DOF2DOF2DOFFlex/extAb/adIn/ext rotFlex/extPro/supFlex/extRad/ul dev0.76° (4.4)0.69° (10.47)0.65° (5.67)0.54° (2.63)5.16° (4.5)3.47° (9.43)2.19° (4.64)
Miezal et al. [15]Validate sensor fusion/algorithmXsensEKF, WLSNot describedNatural Point Optitrack system 13 camerasYesEight-shaped movements at varied speeds, smooth parts imitating reaching and steering in the case of real-slow, and agile parts with quick starts and stops, as well as, parts reminding of sportive movements, such as boxing, in the case of real fastShoulderElbowWrist1DOF1DOF1DOFChaintracker (real fast w/mag)
9.38° (5.79)11.91° (6.27)7.37° (4.60)
ShoulderElbowWrist1DOF1DOF1DOFChaintracker (real slow w/mag)
4.76° (2.24)8.83° (4.64)4.72° (2.61)
ShoulderElbowWrist1DOF1DOF1DOFOptitracker (real fast w/mag)
1.88° (0.91)2.22° (1.38)2.28° (1.15)
Shoulder ElbowWrist1DOF1DOF1DOFOptitracker (real fast w/mag)
1.27° (0.81)2.16° (1.35)2.32° (1.37)
Lambretcht et al. [62]Validate sensor fusion/algorithmCustomDMP algorithmS1: SternumS2: Upper armS3: Distal forearmS4: HandOptotrakYesReaching movementsShoulderElbowWrist3DOF2DOF2DOFAzimuthElevInt rotFlexProFlex/ExtDev4.9°1.2°2.9°7.9°1.5°5.5°2.6°0.990.990.990.990.990.970.94
Zhang et al. [34]Validate sensor fusion/algorithmXsensUKFS1: SternumS2: Lateral side above the elbow S3: Lateral and flat side of the forearm near the wristBTS SMART-D optoelectronic tracking systemYesMove the upper limb arbitrarily.ShoulderElbow3DOF2DOFFlex/extAb/adInt/ext rotFlex/extPro/supIndependent Estimation
0.070° (0.083)0.023° (0.042)0.061° (0.061)0.052° (0.155)0.321° (0.265)0.11°0.04°0.08°0.16°0.41°0.990.990.990.810.96
ShoulderElbow3DOF2DOFFlex/extAb/adInt/ext rotFlex/extPro/supConstraints method
0.040° (0.039)0.013° (0.018)0.029° (0.032)0.046° (0.100)0.155° (0.143)0.05°0.02°0.04°0.11°0.21°0.990.990.990.880.96
ShoulderElbow3DOF2DOFFlex/extAb/adInt/ext rotFlex/extPro/supPapers proposed method
0.028° (0.029)0.007° (0.013)0.035° (0.036)0.054° (0.093)0.168° (0.153)0.04°0.01°0.05°0.10°0.22°0.990.990.990.890.96
Morrow et al. [42]Validate protocolOpalBilateral:S1: Lateral aspect upper armsS2: ForearmsRaptor 12 Digital Real-time Motion Capture SystemNoPeg transfer task using straight laparoscopic surgical instruments.ShoulderElbow1DOF1DOFElevationFlexion3.0° (2.1)2.2° (1.6)6.8° (2.7)8.2° (2.8)
Callegas-Cuerro et al. [59]Validate protocolInvensense MPU-9150KFS1: External arm aligned with the humerus.S2: Between the radial styloid and ulnar styloid, aligned with external part of the hand.Qualisys Oqus 5NoFlex/extElbow1DOFFlex/ext< 3.0° to < 5.0°2.44%
Meng et al. [56]Validate protocolShimmerKFNot describedVicon Mocap SystemNo(1) Raise shoulder. (2) Move shoulder right then left. (3) Clockwise axial rotation to its max, then rotate the upper arm counter clockwise. (4) Elbow extension move into flexion.ShoulderElbow3DOF2DOFFlex/extAb/adIn/ext rotFlex/extPro/sup0.50° (1.79)0.18° (1.34)0.16° (1.96)1.86° (1.85)1.22° (2.87)1.85°1.35°1.96°2.62°3.12°
Cifuentes et al. [81]Validate protocolCustomS1: ArmS2: ForearmOptical tracking systemNoReaching and grasping from the rest position with the forearm on the table, at angle of approximately 90° with respect to the arm before reaching and grasping an object, and then returning it to starting position.Elbow1DOFFlex/extNo discrete data reported only figures of continuous data
Muller et al. [22]Validate sensor fusion/algorithmXsensKF*S1: Thorax.S2: Lateral side of the armS3: Posterior side of the wristViconNo(1) Flex/ext in a horizontal plane with the shoulder abducted 90° flex/ext in a sagittal plane while standing with the elbow close to the trunk. (2) Flex/ext in a sagittal plane with the spine bent forward 90° and the upper arm aligned horizontally and parallel to the ground sup/pro with the elbow flexed 90°ElbowElbow2DOF2DOFFlex/extPro/supFlex/extPro/supProposed algorithm
2.7°3.8°
Manual alignment
3.8°8.7°
Bertomu-Motos et al. [55]Validate sensor fusion/algorithmShimmerEKFS1: ShoulderS2: Upper armOptitrackNoThe activity consisted of taking a box from the perimeter and placing it in the centre of the screen.Shoulder5DOFUnclearWithout compensation Filter
5.24° (3.38)0.5° (1.6)3.6° (2.1)1.8° (1.0)1.60° (0.6)
Shoulder5DOFUnclearCompensation filter
1.69° (2.1)1.1° (0.8)5.9° (2.3)2.6° (1.7)0.9° (1.2)
Karunarathne et al. [68]Validate sensor fusion/algorithmBioKin WMSKF*S1: Near the elbowS2: WristViconNoLifting a water bottleElbow1DOFFlex/extHigh-pass filte—gyroscope
10.18°
Elbow1DOFFlex/extLow-pass filter—accelerations
18.30°
Elbow1DOFFlex/extTradition complementary filter
10.30°
Elbow1DOFFlex/extAdaptive complementary filter
8.77°
El-Gohary et al. [50]Validate Sensor fusion/algorithmOpalUKFS1: Upper armS2: ForearmVicon motion analysis systemNoSingle movements: Shoulder flex/ext, ab/ad, Elbow flex/ext and forearm sup/pro.ShoulderElbow2DOF2DOFFlex/extAb/adFlex/extPro/sup5.5°4.4°6.5°0.95°0.980.990.980.95
Complex tasks: (1) touching nose and (2) reaching for doorShoulderElbow1DOF1DOF9.8°8.8°6.5°5.5°0.940.95
El-Gohary et al. [51]Validate Sensor fusion/algorithmOpalUKFS1: Between the shoulder and elbowS2: Near the wristEagle Analog System, Motion AnalysisNoSingle movements at different speeds: Shoulder flex/ext, ab/ad, Elbow flex/ext, sup/proShoulderElbow2DOF2DOFFlex/extAb/adFlex/extPro/supNormal speed
0.970.940.920.96
ShoulderElbow2DOF2DOFFlex/extAb/adFlex/extPro/supFast speed
0.940.910.890.93
Perez et al. [39]Validate sensor fusion/algorithmXsensS1: BackS2: 18 cm from acromionS3: 25 cm from epicondyleS4: 5.5 cm from distal radio-cubital joint.BTS SMART-D optoelectronic tracking systemNoSingle movements: Shoulder flex/ ext, horizontal ab/ad, and internal rotation. Elbow flex, pro/sup and wrist flex/ext.ShoulderElbowWrist3DOF2DOF1DOFFlex/extAb/adIn rotFlexPro/supFlex/ext13.4°17.2°60.4°5.8°24.1°11.6°0.990.710.990.980.960.98
Pouring water from a glass jar into a glassShoulderElbowWrist3DOF2DOF1DOFFlex/extAb/adIn rotFlex/extPro/supFlex/ext13.8°7.4°28.8°18.6°11.7°26.8°0.990.900.850.970.920.92
Zhou et al. [37]Validate sensor fusion/algorithmXsensKFS1: Lateral aspect of upper arm between the lateral epicondyle and the acromion process (5 cm from the AP)S2: Wrist centre on the palmer aspectCODANoReaching, shrugging, forearm rotationElbow2DOFFlex/extRot0.4° (2.34)0.06° (4.82)2.4°4.8°
Luinge et al. [41]Validate sensor fusion/algorithmXsensKFS1: Lateral upper arm near the elbowS2: Dorsal side of the forearm near the wrist.ViconNo(1) Mimicking eating routines (pouring a glass eating soup, eating spaghetti, eating meat, drinking). (2) Mimicking morning routines (splashing water on face and drying it using a towel, applying deodorant, buttoning a blouse, combing hair, brushing teeth).Elbow2DOFNo discreet data reported
Peppoloni et al. [57]Validate kinematic modelShimmerUKFS1: Scapula beside the angulus acromialisS2: Lateral side of the upper arm above the elbow.S3: Lateral side of forearm a few centimetres far from the wrist.ViconNoSingle movements:Scapula elev/dep, ante-position/retro-position. Shoulder flex/ext, ab/ad, and int/ext rotation. Elbow flex/ext, pro/sup.7DOF model
ScapulaShoulderElbow2DOF3DOF2DOFElev/depProf/retrFlex/extAb/adIn/ext rotFlex/extPro/sup6.19°3.43°8.19°10.68°8.79°5.00°9.61°0.650.740.940.630.970.990.85
5DOF model
ShoulderElbow3DOF2DOFFlex/extAb/adIn/ext rotFlex/extPro/sup7.03°6.03°4.95°9.93°11.29°0.950.870.990.980.85
Robert-Lachaine et al. [25]Validate calibration methodXsensKFOptotrakNoSingle plane movementsNo discrete data reported
Bouvier et al. [23]Validate calibration methodXsensKFS1: SternumS2: Central third of upper arm laterally (or slightly posterior)S3: Dorso-distally on the forearmS4: Dorsum handEagle 4 Optoelectric systemNoMove through 9 calibration trials for each joint.ShoulderElbowWrist3DOF2DOF2DOFFlex/extAb/AdWheelFlex/extPro/supFlex/extAb/sd20.46°14.76°14.21°13.9°0.840.940.930.68

Abbreviations: RMSE root mean square error, SD standard deviation, CMC coefficient of multiple correlation, KBF Kalman-based filter, KF Kalman filter, EKF extended Kalman filter, UKF unscented Kalman filter, WLS weighted least squares, Flex flexion, Ext extension, Pro pronation, Sup supination, Ab abduction, Ad adduction, Dev deviation, Rad radial, Uln ulnar, In internal, Ex external, Rot rotation, Elev elevation, Dep depression, DOF degrees of freedom, C customised, M manufacture

*The information was obtained from the manufacturer procedure manual or other referenced papers

–Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual

List of the 8 articles organised by first author and containing information related to the validation of wearable sensors for the measurement of joint angle for simulated movements of the upper limb when compared to a robotic device Abbreviations: RMSE root mean square error, SD standard deviation, CMC coefficient of multiple correlation, KBF Kalman-based filter, KF Kalman filter, EKF extended Kalman filter, UKF unscented Kalman filter, WLS weighted least squares, Flex flexion, Ext extension, Pro pronation, Sup supination, Ab abduction, Ad adduction, Dev deviation, Rad radial, Uln ulnar, In internal, Ex external, Rot rotation, Elev elevation, Dep depression, DOF degrees of freedom, C customised, M manufacture –Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual List of the selected 22 articles organised by first author and containing information related to the validation of wearable sensors for the measurement of joint angle in upper limb when compared to a three-dimensional motion analysis system Abbreviations: RMSE root mean square error, SD standard deviation, CMC coefficient of multiple correlation, KBF Kalman-based filter, KF Kalman filter, EKF extended Kalman filter, UKF unscented Kalman filter, WLS weighted least squares, Flex flexion, Ext extension, Pro pronation, Sup supination, Ab abduction, Ad adduction, Dev deviation, Rad radial, Uln ulnar, In internal, Ex external, Rot rotation, Elev elevation, Dep depression, DOF degrees of freedom, C customised, M manufacture *The information was obtained from the manufacturer procedure manual or other referenced papers –Information was not reported and/or unclear in the study and/or unable to be obtained from the manufacturer manual

Robot Comparisons

Eight studies reported the error of wearable sensors when compared to simulated upper limb movement on a robotic device (Table 2). A mean error between 0.06 and 1.8° for flexion and 1.05 and 1.8° for lateral deviation of the wrist was reported using Xsens [28, 31]. For elbow flexion/extension, the difference between Invensence and the robotic device was between 2.1 and 2.4° [59]. For finger flexion/extension, RMSEs ranged from 5.0 to 7.0° using a customised wearable sensor system [77]. Three studies reported the error associated with the use of different fusion algorithms. Using the unscented Kalman filter (UKF) to fuse data from Opal wearable sensors, the RMSE range was 0.8–8.1° for 2DOF at the shoulder, 0.9–2.8° for 1DOF at the elbow, 1.1–3.9° for 1DOF of the forearm, and 1.1–2.1° for 2DOF at the wrist [46, 48]. The rotation of the shoulder and twist of the wrist resulted in more error compared to single plane movements of flexion/extension and pronation/supination [46, 48]. When the UKF was compared to a modified UKF, lower RMSEs were found across all 6DOF using the modified UKF [46]. One study investigated the effects that speed of movement had on measurement error. Using Opal wearable sensors, the UKF was compared to the extended Kalman filter (EKF) under three speed conditions: slow, medium, and fast. For slow movements, both fusion algorithms were comparable across all 6DOF (RMSE 0.8–7.8° for the UKF and 0.8–8.8° for the EKF). The UKF resulted in less error across 6DOF for the medium (RMSE 1.2–3.0°) and fast (RMSE 1.1–5.9°) speeds compared to the EKF (RMSE 1.4–8.6°; 1.4–9.7°) [48].

3DMA Comparisons

Twenty-two studies compared the joint angles calculated by wearable sensors, both custom and commercial, to a ‘gold standard’ 3DMA system (Table 3). Studies that used same segment tracking (i.e. motion analysis markers directly on the wearable sensors) were reported in 7 studies. Opal wearable sensors were compared to a 3DMA system during simulated swimming (multiplane movement). The largest difference between the two systems occurred at the elbow (RMSE 6–15°), with the least occurring at the wrist (RMSE 3.0–5.0°) [45]. Xsens was compared to codamotion during single plane movement, with the addition of a dynamic anatomical calibration trial [30]. The largest difference occurred at the elbow (5.16° ± 4.5 to 0.54° ± 2.63), and the least difference at the shoulder (0.65° ± 5.67 to 0.76° ± 4.40) [30]. Xsens was compared to Optotrak with consistent differences between systems across all DOFs of the shoulder (RMSE 2.5–3.0°), elbow (RMSE 2.0–2.9°), and wrist (RMSE 2.8–3.8°) [24]. Three studies investigated the performance of wearable sensors using different fusion methods to amalgamate the data and compared this to a ‘gold standard’ system. Zhang and colleagues [34] compared the accuracy of their own algorithm to two pre-existing algorithms. Comparing Xsens to the BTS Optoelectronic system, their methodology resulted in less error (RMSE = 0.08°, CC = 0.89 to 0.99) across 5DOF compared to the two other methods [34]. The addition of a magnetometer in the analysis of data was also investigated using the EKF- and non-EKF-based fusion algorithm [15]. The latter produced the least difference between the two systems, irrespective of the speed of the movement and whether or not a magnetometer was included. In contrast, the EKF fusion algorithm resulted in the largest difference from the reference system, particularly for fast movements where magnetometer data was included (7.37° ± 4.60 to 11.91° ± 6.27) [15]. The level of customisation to achieve these results is summarised in Table 4.
Table 4

Summary of the software customisation reported by the authors for validation studies that used the same segment tracking

First authorSensor hardwareSoftware
Sensor fusion algorithmCalibrationAnatomical segment definitionKinematic calculation
Robert Lachaine et al. [24]Commercial—Xsens MVNManufacturerManufacturerCustomCustom
Ligorio et al. [69]Commercial—YEI TechnologyCustomCustomCustomCustom
Fantozzi et al. [45]Commercial—ADPM OpalCustomCustomCustomCustom
Gil-Agudo et al. [30]Commercial—Xsens MTxCustomCustomCustomCustom
Miezal et al. [15]Commercial—XsensDid not reportDid not reportCustomCustom
Lambretcht et al. [62]Commercial—InvenSense MPU9150 chipCustomCustomCustomCustom
Zhang et al. [34]Commercial—Xsens MTxCustomManufacturerCustomCustom
Summary of the software customisation reported by the authors for validation studies that used the same segment tracking One study compared the difference between YEI Technology (YEI technology, Portsmouth, OH) wearable sensors and Vicon during three customised calibration methods for the elbow, which resulted in RMSEs that ranged from 3.1 to 7.6° [69].

Reliability

Adequate to excellent agreement was reported for 2DOF at the shoulder (ICC 0.68–0.81) and poor to moderate agreement for the 2DOF at the elbow (ICC 0.16–0.83). The wrist demonstrated the highest overall agreement with ICC values ranging from 0.65 to 0.89 for 2DOF [73].

Risk of Bias

The sample sizes of the included studies were mostly inadequate, with 30% including single participants (Table 1). Twenty-eight percent of the included studies were conference papers, providing limited information.

Discussion

This systematic review described the characteristics of wearable sensors that have been applied in research and clinical settings on the upper limb, the populations with whom they have been used with, and their established psychometric properties. The inclusion of 66 studies allowed for a comprehensive synthesis of information. Similar to other systematic reviews on wearable sensors, commercial wearable sensors, as opposed to custom designed, were reported in most studies (83%) [17]. One benefit for users of commercial wearable sensors is the user-friendly nature of the associated manufacturer guidelines and processing software, including in-built fusion algorithms and joint calculation methods. However, the studies that utilised commercial hardware often customised aspects of the software (i.e. fusion algorithm, calibration method, anatomical segment definition, and the kinematic calculation). Therefore, the validity and reliability of an entirely commercial system (hardware and software) for use in the upper limb remains unknown. Customisation impacts the clinical utility of the wearable sensor systems, especially if there are no support personnel with appropriate knowledge and expertise. Of the studies reviewed, there was no consensus on the procedures to follow for using wearable sensors on the upper limb. The placement of the wearable sensors varied and, in some cases, was poorly described. Manufacturer guidelines for placement of commercial wearable sensors were not referred to, which lead to apparent differences in placement for studies that utilised the same commercial brand. Multiple fusion algorithms were reported, with no clear outcome about which was best suited to a specific joint or movement. The level of customisation of fusion algorithms makes it difficult to compare between studies, and often, the specifics of the algorithm were not readily available, limiting replication. Similar inconsistencies and a lack of consensus were reported in other systematic reviews investigating use of wearable sensors [16, 87]. Without clear guidelines, measurement error can be introduced and/or exacerbated depending on the procedures followed. The methods of calibration also varied between studies, with a static anatomical calibration the most commonly utilised method (typically adopting a neutral pose, standing with arms by the side and palms facing forward, as recommended by most manufacturers). Dynamic anatomical calibration was often customised to suit the needs of the study and the joint being measured. For example, dynamic anatomical calibration of the elbow varied from repetitions of flexion and extension at various speeds [59], to the rapid movement of the arm from 45° to neutral [42]. Details of the dynamic anatomical calibrations were omitted in some studies, limiting replication. More pertinent for the calculation of joint kinematics is anatomical calibration as compared to system calibration, with the type of calibration (i.e. static or dynamic) and movements of the dynamic anatomical calibration, having a significant impact on the accuracy of wearable sensors [69]. Of the 66 studies included in this review, almost half (45%) were validation studies with the remaining studies using wearable sensors as an outcome measure. Over one third (29%) were conference proceedings in the field of engineering, thus limiting the amount of information available. The median sample size was 7.6 participants per study; only one study was considered to have an adequate sample size for the validation of a measurement tool as per the COSMIN guidelines [19]. The majority (78%) of the results were obtained from healthy adults, with clinical populations (12%) and those under the age of 18 (1.5%) not well represented. Research investigating the use of wearable sensors to measure lower limb kinematics has demonstrated a level of accuracy with clinical populations and children. Errors < 4° were reported for elderly individuals with hemiparesis [88] and RMSEs between 4.6 and 8.8° for children with spastic cerebral palsy [10]. There is potential for wearable sensors to be applied to the upper limb of these populations; however, more research is required to determine the optimal procedures prior to implementation in clinical practice. The validity and reliability of wearable sensors when applied to the upper limb has not been clearly described to date. When compared to a robotic device, the commercial wearable sensors with customised software recorded errors below McGinley’s [7] suggested 5.0° threshold. Less than 3.9° was reported for replica/simulated movements of the wrist in 3DOF [28, 46, 48, 56], < 3.1° for 2DOF at the elbow [46, 48, 56], and < 2.5° for 1DOF (flexion/extension) at the shoulder [48]. Shoulder internal and external rotation resulted in the largest error (3.0–9.7°) [48], and therefore, results for this movement should be interpreted with caution. The next section will discuss ‘in vivo’ studies with 3DMA as a pseudo gold standard. Studies that made a direct comparison between the wearable sensors and 3DMA system (i.e. used the same segment tracking) demonstrated differences that exceeded the suggested 5.0° threshold, with up to 15.0° difference reported for the elbow. However, depending on the software specifications and level of customisation, a difference of < 0.11° (3DOF shoulder), < 0.41° (2DOF elbow), and < 2.6 (2DOF wrist) was achievable. The range in difference observed between the two systems is indicative that wearable sensors are still largely in a ‘developmental phase’ for the measurement of joint angle in the upper limb. Consistent with prior findings, error values were unique to the joint and movement tasks being measured. Most of the tasks involved movements in multiple planes (i.e. reaching tasks), which resulted in more error compared to studies that assessed isolated movement in a single plane (i.e. flexion and extension). Measuring multiple planes of movement poses a further challenge to motion analysis and needs careful consideration when interpreting the results [89].

Limitations

Due to the heterogeneity in the reported studies, a meta-analysis was not appropriate given the variance in sample sizes, movement tasks, different procedures, and statistical analyses used. It was also not possible to apply a standard assessment of quality and bias due to the diversity of the studies. The inclusion of small samples (30% single participant) is a potential threat to validity, with single participant analysis insufficient to support robustness and generalisation of the evidence. The inclusion of conference papers (28%) meant that many papers provided limited detail on the proposed system and validation results. Small sample sizes and the inclusion of mostly healthy adults means the results of this review cannot be generalised to wider clinical populations. In addition, studies that utilised different segment tracking (i.e. 3DMA markers were not mounted on the wearable sensor) were not further analysed as it was not possible to delineate between the sources of error.

Conclusion

Wearable sensors have become smaller, more user-friendly, and increasingly accurate. The evidence presented suggests that wearable sensors have great potential to bridge the gap between laboratory-based systems and the goniometer for the measurement of upper limb joint angle during dynamic movement. A level of acceptable accuracy was demonstrated for the measurement of elbow and wrist flexion/extension when compared to a robotic device. Error was influenced by the fusion algorithm and method of joint calculation, which required customisation to achieve errors < 2.9° from known angles on a robotic device. Higher error margins were observed in vivo when compared to a 3DMA system, but < 5° was achievable with a high level of customisation. The additional level of customisation that was often required to achieve results with minimal error is particularly relevant to clinicians with limited technical support, and critically, when using a system ‘off the shelf’, the expected level of accuracy may not be comparable to the findings reported in this review. With this technology rapidly evolving, future research should establish standardised protocol/guidelines, and subsequent reliability and validity for use in the upper limb, and in various clinical populations. Direct comparisons with the gold standard (i.e. same segment tracking) is needed to produce results that are most meaningful. We recommend and encourage the use of wearable sensors for the measurement of flexion/extension in the wrist and elbow; however, this should be combined with outcome measures that have demonstrated reliability and validity in the intended population.
  58 in total

Review 1.  Movement biomechanics goes upwards: from the leg to the arm.

Authors:  G Rau; C Disselhorst-Klug; R Schmidt
Journal:  J Biomech       Date:  2000-10       Impact factor: 2.712

2.  Ambulatory measurement of arm orientation.

Authors:  H J Luinge; P H Veltink; C T M Baten
Journal:  J Biomech       Date:  2006-02-07       Impact factor: 2.712

3.  Use of multiple wearable inertial sensors in upper limb motion tracking.

Authors:  Huiyu Zhou; Thomas Stone; Huosheng Hu; Nigel Harris
Journal:  Med Eng Phys       Date:  2007-01-23       Impact factor: 2.242

4.  Ambulatory measurement of shoulder and elbow kinematics through inertial and magnetic sensors.

Authors:  Andrea Giovanni Cutti; Andrea Giovanardi; Laura Rocchi; Angelo Davalli; Rinaldo Sacchetti
Journal:  Med Biol Eng Comput       Date:  2007-12-18       Impact factor: 2.602

5.  Estimation of upper-limb orientation based on accelerometer and gyroscope measurements.

Authors:  Rick A Hyde; Laurence P Ketteringham; Simon A Neild; Rosie S Jones
Journal:  IEEE Trans Biomed Eng       Date:  2008-02       Impact factor: 4.538

6.  Joint angle tracking with inertial sensors.

Authors:  Mahmoud El-Gohary; Sean Pearson; James McNames
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2008

Review 7.  The reliability of three-dimensional kinematic gait measurements: a systematic review.

Authors:  Jennifer L McGinley; Richard Baker; Rory Wolfe; Meg E Morris
Journal:  Gait Posture       Date:  2008-11-13       Impact factor: 2.840

8.  Outcome evaluation in shoulder surgery using 3D kinematics sensors.

Authors:  Brian Coley; Brigitte M Jolles; Alain Farron; Aline Bourgeois; François Nussbaumer; Claude Pichonnaz; Kamiar Aminian
Journal:  Gait Posture       Date:  2006-08-28       Impact factor: 2.840

9.  Advances in wearable technology and applications in physical medicine and rehabilitation.

Authors:  Paolo Bonato
Journal:  J Neuroeng Rehabil       Date:  2005-02-25       Impact factor: 4.262

10.  Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement.

Authors:  David Moher; Alessandro Liberati; Jennifer Tetzlaff; Douglas G Altman
Journal:  BMJ       Date:  2009-07-21
View more
  15 in total

1.  Quantifying Pathological Synergies in the Upper Extremity of Stroke Subjects With the Use of Inertial Measurement Units: A Pilot Study.

Authors:  Miguel M C Bhagubai; Gerjan Wolterink; Anne Schwarz; Jeremia P O Held; Bert-Jan F Van Beijnum; Peter H Veltink
Journal:  IEEE J Transl Eng Health Med       Date:  2020-12-07       Impact factor: 3.316

2.  Concurrent Validity and Reliability of an Inertial Measurement Unit for the Assessment of Craniocervical Range of Motion in Subjects with Cerebral Palsy.

Authors:  Cristina Carmona-Pérez; Juan Luis Garrido-Castro; Francisco Torres Vidal; Sandra Alcaraz-Clariana; Lourdes García-Luque; Francisco Alburquerque-Sendín; Daiana Priscila Rodrigues-de-Souza
Journal:  Diagnostics (Basel)       Date:  2020-02-01

3.  Validity and Reliability of a New Optoelectronic System for Measuring Active Range of Motion of Upper Limb Joints in Asymptomatic and Symptomatic Subjects.

Authors:  Rodrigo Martín-San Agustín; Jose A García-Vidal; German Cánovas-Ambit; Aurelio Arenas-Della Vecchia; Manuel López-Nicolás; Francesc Medina-Mirapeix
Journal:  J Clin Med       Date:  2019-11-02       Impact factor: 4.241

Review 4.  Wearable Health Devices in Health Care: Narrative Systematic Review.

Authors:  Lin Lu; Jiayao Zhang; Yi Xie; Fei Gao; Song Xu; Xinghuo Wu; Zhewei Ye
Journal:  JMIR Mhealth Uhealth       Date:  2020-11-09       Impact factor: 4.773

5.  Wearable Motion Sensor Device to Facilitate Rehabilitation in Patients With Shoulder Adhesive Capsulitis: Pilot Study to Assess Feasibility.

Authors:  Yu-Pin Chen; Chung-Ying Lin; Ming-Jr Tsai; Tai-Yuan Chuang; Oscar Kuang-Sheng Lee
Journal:  J Med Internet Res       Date:  2020-07-23       Impact factor: 5.428

6.  Measures of Interjoint Coordination Post-stroke Across Different Upper Limb Movement Tasks.

Authors:  Anne Schwarz; Janne M Veerbeek; Jeremia P O Held; Jaap H Buurke; Andreas R Luft
Journal:  Front Bioeng Biotechnol       Date:  2021-01-28

7.  The Reliability and Validity of Wearable Inertial Sensors Coupled with the Microsoft Kinect to Measure Shoulder Range-of-Motion.

Authors:  Peter Beshara; Judy F Chen; Andrew C Read; Pierre Lagadec; Tian Wang; William Robert Walsh
Journal:  Sensors (Basel)       Date:  2020-12-17       Impact factor: 3.576

8.  Efficient Multiaxial Shoulder-Motion Tracking Based on Flexible Resistive Sensors Applied to Exosuits.

Authors:  J Luis Samper-Escudero; Aldo F Contreras-González; Manuel Ferre; Miguel A Sánchez-Urán; David Pont-Esteban
Journal:  Soft Robot       Date:  2020-01-06       Impact factor: 8.071

9.  Development of a Post-stroke Upper Limb Rehabilitation Wearable Sensor for Use in Sub-Saharan Africa: A Pilot Validation Study.

Authors:  Charmayne M L Hughes; Alexander Louie; Selena Sun; Chloe Gordon-Murer; Gashaw Jember Belay; Moges Baye; Xiaorong Zhang
Journal:  Front Bioeng Biotechnol       Date:  2019-11-12

10.  Assessment of Upper Limb Movement Impairments after Stroke Using Wearable Inertial Sensing.

Authors:  Anne Schwarz; Miguel M C Bhagubai; Gerjan Wolterink; Jeremia P O Held; Andreas R Luft; Peter H Veltink
Journal:  Sensors (Basel)       Date:  2020-08-24       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.