| Literature DB >> 32024117 |
Cristina Carmona-Pérez1,2, Juan Luis Garrido-Castro3,4, Francisco Torres Vidal3, Sandra Alcaraz-Clariana2,5, Lourdes García-Luque2, Francisco Alburquerque-Sendín4,5, Daiana Priscila Rodrigues-de-Souza5.
Abstract
OBJECTIVE: This study aimed to determine the validity and reliability of Inertial Measurement Units (IMUs) for the assessment of craniocervical range of motion (ROM) in patients with cerebral palsy (CP).Entities:
Keywords: inertial sensors; kinematics; pediatric neurological disease
Year: 2020 PMID: 32024117 PMCID: PMC7168926 DOI: 10.3390/diagnostics10020080
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1Devices and procedure of assessment. (A) Inertial Measurement Unit (IMU) Shimmer3 ® (Dublin, Ireland); (B) Cervical Range of Motion (CROM) 3 ® device (Lindstrom, MN, USA); (C) positioning of IMU and CROM to assess craniocervical range of motion (first assessment on the first day); (D) positioning of IMU to assess craniocervical range of motion (second assessment on the first day, and assessment on the second day).
Demographics and clinical characteristics of the subjects.
| CP Group ( | Control Group ( | ||
|---|---|---|---|
|
| 9.2 (3.2) | 8.7 (3.3) | 0.594 |
|
| 14/9 | 14/9 | |
|
| 28.3 (12.7) | 34.6 (16.8) | 0.161 |
|
| 1.32 (0.20) | 1.36 (0.20) | 0.503 |
|
| 15.5 (3.4) | 17.5 (3.5) | 0.049* |
|
| - | - | |
|
| - | - | |
|
| - | - | |
|
| - | - | |
|
| - | - |
Quantitative data are expressed as mean (standard deviation). Abbreviations: GMFCS, Gross Motor Function Classification System; BMI, body mass index. * indicates p < 0.05.
Concurrent validity between the first IMU assessment performed on the first day of measurements and CROM by groups.
| Spatial Plane | IMU First Assessment on the First Day (Standard Deviation) | CROM Assessment (Standard Deviation) | Pearson r ( | Student’s t-test ( |
|---|---|---|---|---|
|
| ||||
|
| 133.3 (24.6) | 129.0 (22.4) | 0.969 (<0.001) | −3.333 (0.003) |
|
| 153.5 (19.9) | 145.3 (14.7) | 0.601 (0.003) | −2.396 (0.026) |
|
| 91.3 (25.7) | 96.1 (23.3) | 0.916 (<0.001) | 2.236 (0.036) |
|
| ||||
|
| 137.0 (24.5) | 133.3 (23.9) | 0.992 (<0.001) | −5.771 (<0.001) |
|
| 147.0 (19.0) | 146.1 (16.1) | 0.786 (<0.001) | −0.371 (0.714) |
|
| 90.7 (17.5) | 94.9 (18.9) | 0.949 (<0.001) | 3.413 (0.002) |
Abbreviations: IMU, Inertial Measurement Unit; CROM, Cervical Range of Motion; CI, confidence interval; CP, cerebral palsy. Evaluation data are expressed in degrees.
Figure 2Bland–Altman plots for craniocervical ranges of CP group measured by IMU first assessment on the first day and the CROM in (A) flexion-extension range, (B) rotational range, (C) side-bending range. The red line indicates the mean bias, whereas the blue lines refers to its upper and lower limits (mean ± 1.96 standard deviation). All plots were developed as follows: the Y axis corresponds to the differences between the paired values of both methods (IMU-CROM), whereas the X axis represents the respective value of the average of both (IMU + CROM/2).
Figure 3Bland–Altman plots for craniocervical ranges of the Control group measured by IMU first assessment on the first day and the CROM in (A) flexion-extension range, (B) rotational range, (C) side-bending range. The red line indicates the mean bias, whereas the blue lines refer to its upper and lower limits (mean ± 1.96 standard deviation). All plots were developed as follows: the Y axis corresponds to the differences between the paired values of both methods (IMU-CROM), whereas the X axis represents the respective value of the average of both (IMU + CROM/2).
Intra-day and inter-day reliability of the IMU by groups.
|
| ||||
|
|
|
|
|
|
|
| ||||
|
| 138.8 (26.5) | 0.900 (0.762, 0.958) | 8.0 | 18.6 |
|
| 151.3 (20.1) | 0.821 (0.600, 0.920) | 8.3 | 19.4 |
|
| 91.3 (23.7) | 0.925 (0.822, 0.968) | 6.7 | 15.5 |
|
| ||||
|
| 135.0 (24.7) | 0.893 (0.750, 0.955) | 7.9 | 18.4 |
|
| 147.6 (20.3) | 0.902 (0.750, 0.961) | 5.4 | 12.6 |
|
| 86.3 (16.1) | 0.913 (0.772, 0.965) | 5.0 | 11.4 |
|
| ||||
|
| ||||
|
| ||||
|
| 140.8 (26.4) | 0.873 (0.680, 0.947) | 9.0 | 21.1 |
|
| 156.2 (19.5) | 0.533 (0.117, 0.803) | 13.3 | 30.9 |
|
| 94.4 (24.8) | 0.890 (0.743, 0.953) | 8.3 | 19.3 |
|
| ||||
|
| 140.2 (29.5) | 0.831 (0.602, 0.928) | 11.0 | 25.6 |
|
| 153.2 (19.1) | 0.846 (0.601, 0.935) | 6.5 | 15.1 |
|
| 93.1 (16.5) | 0.864 (0.653, 0.946) | 6.1 | 14.2 |
Abbreviations: IMU, Inertial Measurement Unit; ICC, Intraclass Correlation Coefficient; CI, confidence interval; SEM, Standard Error of Measurement; MDC, Minimum Detectable Change. Evaluation data are expressed in degrees.