The aim of this study was to evaluate the validity of the Veloflex infrared dynamic angle-meter (Veloflex-IDA) and the intra- and inter-rater reliability when measuring the ranges of motion (ROMs) of the upper limb joints. Thirty-five healthy and 20 symptomatic participants were evaluated. Twelve upper limb movements were measured in two sessions with the Veloflex-IDA, which is a device composed of a camera that tracks the trajectory of retro-reflective markers. In addition, a goniometer was used in the first session to evaluate concurrent validity. Validity and agreement were evaluated by Pearson correlation coefficient (r) and Bland-Altmann plots. Intra- and inter-rater reliability were evaluated using intra-class correlation (ICC), standard error of measurement (SEM), and minimal detectable change (MDC). Both instruments showed excellent correlation for all movements (r range from 0.992 to 0.999). The intra- and inter-rater reliability were excellent (ICC range from 0.95 to 0.99 and 0.90 to 0.98, respectively). Intra-rater reliability showed SEMs <1.38% and <5.19% and inter-rater reliability SEMs <2.26% and <5.22% for asymptomatic and symptomatic, respectively. Veloflex-IDA is a valid and reliable alternative to measure the upper limb joints' ROM and it can be used in clinical practice and research after basic training.Entities:
Keywords: biomedical technology; optoelectronic device; range of motion; upper limb joints
Year: 2019 PMID: 31684077 PMCID: PMC6912542 DOI: 10.3390/jcm8111851
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Figure 1Measurement of shoulder joint range of motion for extension.
Characteristics of the participants.
| Asymptomatic ( | Symptomatic ( | |
|---|---|---|
| Age (years) | 35.3 (12.8) | 44.7 (17.5) |
| BMI (kg/m2) | 25.7 (6.2) | 24.7 (1.5) |
| Gender | Males ( | Males ( |
| Pain (0–10/10) | 0 | 5.1 (1.9) |
| Duration of symptoms (weeks) | - | 6.2 (3.6) |
| Shoulder disorder | - | |
| Elbow disorder | - | |
| Wrist disorder | - |
Figure 2Bland–Altman plots for Veloflex infrared dynamic angle-meter (Veloflex-IDA) and goniometer during (A) shoulder flexion; (B) shoulder extension; (C) shoulder internal rotation; (D) shoulder external rotation; (E) shoulder abduction; (F) elbow flexion; (G) wrist flexion; (H) wrist extension; (I) wrist pronation; (J) wrist supination; (K) wrist ulnar deviation; (L) wrist radial deviation.
Validity between Veloflex-IDA and goniometer to measure the range of motion of the upper limb joints.
| Movement | Goniometer (SD) | Veloflex-IDA (SD) | Pearson Coefficient | LoA − (%) | LoA + (%) | Mean Difference (%) | SD Difference (%) |
|---|---|---|---|---|---|---|---|
| ShF | 166.87° (5.80°) | 167.07° (5.67°) | 0.986 | −1.70° (1.02%) | 2.11° (1.26%) | 0.20° (0.10%) | 0.97° (0.60%) |
| ShE | 37.78° (8.28°) | 37.78° (8.29°) | 0.999 | −0.69° (1.81%) | 0.68° (1.81%) | −0.01° (0.01%) | 0.35° (1.10%) |
| ShIR | 65.44° (6.07°) | 65.49° (6.14°) | 0.999 | −0.59° (0.89%) | 0.70° (1.06%) | 0.05° (0.10%) | 0.32° (0.50%) |
| ShER | 84.71° (6.085) | 84.71° (6.01°) | 0.997 | −0.86° (1.01%) | 0.86° (1.01%) | −0.01° (0.01%) | 0.44° (0.5%) |
| hAB | 175.50° (5.15°) | 175.54° (5.22°) | 0.997 | −0.70° (0.39%) | 0.79° (0.44%) | 0.04° (0.01%) | 0.38° (0.20%) |
| ElbF | 143.35° (3.84°) | 143.36° (3.84°) | 0.995 | −0.77° (0.53%) | 0.79° (0.55%) | 0.01° (0.01%) | 0.39° (0.30%) |
| WrF | 78.18° (4.30°) | 78.93° (3.89°) | 0.958 | −1.70° (2.17%) | 3.21° (4.10%) | 0.75° (1%) | 1.25° (1.6%) |
| WrE | 66.37° (5.65°) | 66.24° (5.71°) | 0.99 | −1.69° (2.54%) | 1.44° (2.17%) | −0.12° (0.2%) | 0.80° (1.2%) |
| WrPr | 86.83° (3.93°) | 86.87° (3.93°) | 0.996 | −0.64° (0.73%) | 0.72° (0.83%) | 0.04° (0.10%) | 0.34° (0.40%) |
| WrSup | 85.93° (3.77°) | 86.08° (4.01°) | 0.967 | −1.87° (2.17%) | 2.18° (2.53%) | 0.15° (0.20%) | 1.03° (1.20%) |
| WrUlDev | 41.49° (3.63°) | 41.51° (3.57°) | 0.996 | −0.60° (1.43%) | 0.64° (1.54%) | 0.02° (0.1%) | 0.31° (0.8%) |
| WrRadDev | 20.02° (3.08°) | 19.87° (3.08°) | 0.993 | −0.90° (4.47%) | 0.58° (2.92%) | −0.15° (−0.7%) | 0.38° (1.9%) |
SD: standard deviation; LoA: limit of agreement; ShF: shoulder flexion; ShE: shoulder extension; ShIR: shoulder internal rotation; ShER: shoulder external rotation; ShAB: shoulder abduction; ElbF: elbow flexion; WrF: wrist flexion; WrE: wrist extension; WrPr: wrist pronation; WrSup: wrist supination; WrUlDev: wrist ulnar deviation; WrRadDev: wrist radial deviation.
Reliability of Veloflex-IDA to measure the range of motion of the upper limb joints in symptomatic participants.
| Intra-Rater | Inter-Rater | |||||||
|---|---|---|---|---|---|---|---|---|
| Movement | Retest (SD) | ICC (95% CI) | SEM (SEM%) | MDC | Test (SD) | ICC (95% CI) | SEM (SEM%) | MDC |
| ShF | 166.87° (5.80°) | 0.99 (0.98 to 0.99) | 5.32° (4.65%) | 1.31° | 166.62° (6.54°) | 0.96 (0.93 to 0.98) | 1.09° (0.66%) | 3.04° |
| ShE | 37.69° (8.39°) | 0.99 (0.99 to 0.99) | 0.98° (2.87%) | 1.44° | 37.69° (7.98°) | 0.98 (0.97 to 0.99) | 0.85° (2.26%) | 2.36° |
| ShIR | 65.53° (5.86°) | 0.99 (0.98 to 0.99) | 2.08° (4.02%) | 1.38° | 65.46° (5.95°) | 0.98 (0.97 to 0.99) | 0.66° (1.01%) | 1.85° |
| ShER | 84.60° (5.86°) | 0.99 (0.98 to 0.99) | 2.81° (4.24%) | 1.46° | 83.73° (6.07°) | 0.98 (0.97 to 0.99) | 0.65° (0.77%) | 2.74° |
| ShAB | 175.08° (5.19°) | 0.98 (0.97 to 0.99) | 5.95° (4.93%) | 1.50° | 174.41° (5.06°) | 0.96 (0.93 to 0.98) | 0.91° (0.52%) | 2.75° |
| ElbF | 142.62° (3.92°) | 0.95 (0.91 to 0.98) | 2.22° | 142.65° (3.88°) | 0.92 (0.85 to 0.96) | 1.04° (0.73%) | 2.53° | |
| WrF | 78.18° (4.30°) | 0.97 (0.95 to 0.99) | 2.14° (3.08%) | 1.75° | 78.39° (3.91°) | 0.95 (0.90 to 0.98) | 0.85° (1.07%) | 2.35° |
| WrE | 65.82° (6.17°) | 0.98 (0.96 to 0.99) | 2.47° (5.19%) | 2.25° | 66.09° (6.09°) | 0.98 (0.96 to 0.99) | 0.76° (1.15%) | 2.11° |
| WrPr | 85.01° (3.95°) | 0.96 (0.92 to 0.98) | 1.14° (1.40%) | 2.17° | 85.77° (3.99°) | 0.90 (0.81 to 0.96) | 1.19° (1.39%) | 2.52° |
| WrSup | 85.01° (3.77°) | 0.95 (0.89 to 0.97) | 0.88° (1.03%) | 2.41° | 85.40° (3.43°) | 0.94 (0.88 to 0.97) | 0.85° (1.03%) | 2.61° |
| WrUlDev | 40.71° (3.62°) | 0.96 (0.92 to 0.98) | 1.01° (2.98%) | 1.91° | 41.34° (4.22°) | 0.96 (0.92 to 0.98) | 0.76° (1.84%) | 2.12° |
| WrRadDev | 19.62° (2.95°) | 0.96 (0.91 to 0.98) | 0.84° (4.82%) | 1.74° | 20.18° (2.94°) | 0.96 (0.91 to 0.98) | 0.61° (3.05%) | 1.69° |
SD: standard deviation; ICC: intraclass correlation coefficient; CI: confidence interval; SEM: standard error of measurement; MDC: minimum detectable change; ShF: shoulder flexion; ShE: shoulder extension; ShIR: shoulder internal rotation; ShER: shoulder external rotation; ShAB: shoulder abduction; ElbF: elbow flexion; WrF: wrist flexion; WrE: wrist extension; WrPr: wrist pronation; WrSup: wrist supination; WrUlDev: wrist ulnar deviation; WrRadDev: wrist radial deviation.
Reliability of Veloflex-IDA to measure the range of motion of the upper limb joints in symptomatic participants.
| Intra-Rater | Inter-Rater | |||||||
|---|---|---|---|---|---|---|---|---|
| Shoulder Disorder ( | Test (SD)/Retest (SD) | ICC (95% CI) | SEM (SEM%) | MDC | Test (SD) | ICC (95% CI) | SEM (SEM%) | MDC |
| ShF | 114.49° (53.28°)/114.65° (54.00°) | 0.99 (0.98 to 0.99) | 5.32° (4.65%) | 14.77° | 114.76° (53.68°) | 0.99 (0.99 to 0.99) | 5.37° (4.68%) | 14.88° |
| ShE | 34.08° (9.80°)/33.93° (9.82°) | 0.99 (0.97 to 0.99) | 0.98° (2.87%) | 2.71° | 34.08° (9.50°) | 0.99 (0.97 to 0.99) | 0.95° (2.79%) | 2.63° |
| ShIR | 51.74° (20.78°)/51.87° (21.64°) | 0.99 (0.98 to 0.99) | 2.08° (4.02%) | 5.76° | 51.98° (21.27°) | 0.99 (0.98 to 0.99) | 2.13° (4.09%) | 5.89° |
| ShER | 66.21° (28.10°)/66.21° (28.1°5) | 0.99 (0.99 to 0.99) | 2.81° (4.24%) | 7.79° | 66.36° (27.87°) | 0.99 (0.99 to 0.99) | 2.73° (4.11%) | 7.56° |
| ShAB | 120.49° (59.46°)/120.63° (60.35°) | 0.99 (0.99 to 0.99) | 5.95° (4.93%) | 16.48° | 120.49° (60.17°) | 0.99 (0.99 to 0.99) | 6.02° (4.99%) | 16.68° |
|
| ||||||||
| WrF | 69.40° (21.36°)/68.66° (21.09°) | 0.99 (0.98 to 0.99) | 2.14° (3.08%) | 5.92° | 68.66° (21.08°) | 0.99 (0.98 to 0.99) | 2.11° (3.07%) | 5.84° |
| WrE | 47.52 (24.58°)/46.64° (24.19°) | 0.99 (0.99 to 0.99) | 2.47° (5.19%) | 6.84° | 46.88° (24.48°) | 0.99 (0.99 to 0.99) | 2.45° (5.22%) | 6.79° |
| WrPr | 80.98 (11.38°)/80.06° (10.81°) | 0.99 (0.98 to 0.99) | 1.14° (1.40%) | 3.15° | 80.33° (11.27°) | 0.99 (0.97 to 0.99) | 1.13° (1.40%) | 3.12° |
| WrSup | 84.87 (5.09°)/83.88° (5.13°) | 0.97 (0.89 to 0.99) | 0.88° (1.03%) | 2.44° | 83.85° (4.93°) | 0.95 (0.85 to 0.99) | 1.10° (1.31%) | 3.05° |
| WrUlDev | 33.70 (10.04°)/35.82° (10.79°) | 0.99 (0.97 to 0.98) | 1.01° (2.98%) | 2.78° | 35.70° (10.52°) | 0.99 (0.97 to 0.99) | 1.05° (2.95%) | 2.91° |
| WrRadDev | 17.37 (5.92°)/16.64° (5.57°) | 0.98 (0.92 to 0.99) | 0.84° (4.82%) | 2.32° | 16.73° (5.27°) | 0.96 (0.85 to 0.99) | 1.05° (6.30%) | 2.92° |
SD: standard deviation; ICC: intraclass correlation coefficient; CI: confidence interval; SEM: standard error of measurement; MDC: minimum detectable change; ShF: shoulder flexion; ShE: shoulder extension; ShIR: shoulder internal rotation; ShER: shoulder external rotation; ShAB: shoulder abduction; ElbF: elbow flexion; WrF: wrist flexion; WrE: wrist extension; WrPr: wrist pronation; WrSup: wrist supination; WrUlDev: wrist ulnar deviation; WrRadDev: wrist radial deviation.