| Literature DB >> 30400527 |
Noel S Ha1,2, Saman Sadeghi3, R Michael van Dam4,5.
Abstract
Radiopharmaceuticals labeled with short-lived positron-emitting or gamma-emitting isotopes are injected into patients just prior to performing positron emission tomography (PET) or single photon emission tomography (SPECT) scans, respectively. These imaging modalities are widely used in clinical care, as well as in the development and evaluation of new therapies in clinical research. Prior to injection, these radiopharmaceuticals (tracers) must undergo quality control (QC) testing to ensure product purity, identity, and safety for human use. Quality tests can be broadly categorized as (i) pharmaceutical tests, needed to ensure molecular identity, physiological compatibility and that no microbiological, pyrogenic, chemical, or particulate contamination is present in the final preparation; and (ii) radioactive tests, needed to ensure proper dosing and that there are no radiochemical and radionuclidic impurities that could interfere with the biodistribution or imaging. Performing the required QC tests is cumbersome and time-consuming, and requires an array of expensive analytical chemistry equipment and significant dedicated lab space. Calibrations, day of use tests, and documentation create an additional burden. Furthermore, in contrast to ordinary pharmaceuticals, each batch of short-lived radiopharmaceuticals must be manufactured and tested within a short period of time to avoid significant losses due to radioactive decay. To meet these challenges, several efforts are underway to develop integrated QC testing instruments that automatically perform and document all of the required tests. More recently, microfluidic quality control systems have been gaining increasing attention due to vastly reduced sample and reagent consumption, shorter analysis times, higher detection sensitivity, increased multiplexing, and reduced instrumentation size. In this review, we describe each of the required QC tests and conventional testing methods, followed by a discussion of efforts to directly miniaturize the test or examples in the literature that could be implemented for miniaturized QC testing.Entities:
Keywords: lab-on-a-chip; microfluidics; pharmaceuticals; positron emission tomography (PET); quality control (QC) testing; radiopharmaceuticals; sensor; single photon emission computed tomography (SPECT)
Year: 2017 PMID: 30400527 PMCID: PMC6190332 DOI: 10.3390/mi8110337
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 2.891
Figure 1Production of positron emission tomography (PET)/single photon emission tomography (SPECT) radiotracers for clinical imaging involves generation of the radioisotope (via cyclotron or generator), radiosynthesis, purification (via HPLC or solid-phase extraction), formulation (via evaporative or solid-phase extraction methods), followed by quality control (QC) testing to ensure safety of the formulated radiotracer prior to injection.
Figure 2Examples of microfluidic pH measurement systems. (A) (Left) Microfluidic device to mix a sample with a universal pH indicator and then measure optical absorbance in a 3.1 mm long detection cell. (Right) Representative absorbance measurements at 551 nm as a function of sample pH. Adapted with permission from [31] Copyright © 2014 The Chemical and Biological Microsystems Society. (B) (Top) pH-sensing polymer microbeads are prepared by introducing the pH indicator during the cross-linking process. (Bottom) The beads are immobilized using optical tweezers or tethered to the surface, and the color of each bead reflects the pH of the local environment. Adapted from [34] with permission of The Royal Society of Chemistry. (C) Microfluidic device with internal channel surface coated with pH-sensitive fluorescent molecules (Oregon Green 514). The fluorescence intensity is correlated with pH. Adapted from [36] with permission of The Royal Society of Chemistry. (D) (Left) polydimethylsiloxane (PDMS) microfluidic channel with polyaniline coating that changes color in response to pH. The zoomed in image shows the response to a pH gradient along the channel. Inset SEM image shows the structure of the coating layer. (Right) Absorbance spectra of the polyaniline coating when exposed to solutions of different pH. Adapted from [37] with permission of The Royal Society of Chemistry. (E) Micro-cantilever undergoes deflection in response to pH-dependent swelling of a hydrogel polymer coating. Deflection is monitored with a laser beam. Adapted with permission from [40] Copyright © 2003 Springer. (F) Thin-film electrodes integrated into a PDMS chip allow pH to be measured based on potential between working and reference electrodes. Adapted with permission from [41] Copyright © 2006 Elsevier. (G) Flow-based microfluidic pH measurement system using ion-sensitive field-effect transistor (ISFET) sensors, one acting as a reference and one as a working electrode. Modulation of flow rates moves the liquid junction and allows measurement of the sample solution (SS) after an initialization process using a baseline solution (BLS). Adapted from [42] with permission from MDPI AG.
Figure 3Microfluidic methods for detection of bacteria. (A) A microbial sample with “universal” fluorescent stain is concentrated by capillary electrophoresis to improve detection sensitivity. (BA = “blocking agent”). Adapted with permission from [50] Copyright © 2007 American Chemical Society. (B) Impedance-based bacterial detection chip. Cells are first focused via dielectrophoresis (DEP) to the center of channel for downstream detection. A detectable shift in resonant frequency and phase occur when a single cell or particle passes through the measurement area due to a change in the channel impedance. Adapted from [53] with permission of The Royal Society of Chemistry.
Figure 4Examples of microfluidic devices for bacterial endotoxin test. (A) Commercial LAL assay system (Endosafe®, Charles River Laboratories, Inc.) comprising a portable, hand-held spectrometer (left) and ~25 mm × 75 mm disposable microfluidic cartridge (right). Adapted from [56]. (B) Photograph of microfluidic LAL test chip (left) and block diagram illustrating detection method (right). Sample mixed with chromogenic substrate is loaded into the microchannel and optical detection is performed in the middle of channel. Photograph adapted with permission from [57] Copyright © 2004 Springer. (C) Screen-printed electrodes for electrical-impedance monitoring of standard gel clot LAL assay. Adapted with permission from [58] Copyright © 2013 Elsevier. (D) Biosensor comprising a gold electrode functionalized with endotoxin-specific aptamer. Binding of endotoxin is detected via impedance spectroscopy. Adapted with permission from [60] Copyright © 2012 Elsevier.
Figure 5Examples of miniature chromatography systems. (A) Microfluidic HPLC system. A sliding fluoropolymer element is used to create a high-pressure valve for sample injection. When the element is in the left position, sample can be loaded via the sample input port. When the HPLC pump is activated, the flowing buffer closes the valve (slides the element to the right) and sweeps the injected sample through the integrated monolith column for separation. Adapted with permission from [64] Copyright © 2005 American Chemical Society. (B) Commercial HPLC-chip (Agilent)) includes an integrated micro-rotary injection valve with sample enrichment column serving as the injection loop, for injecting samples into the integrated LC column and downstream electrospray emitter to nebulize and transfer the sample to a mass spectrometric detector. Adapted from [65]. (C) Microchip electrophoresis setup with a volumetric sample injection chip, a separation capillary, and an optical detection chip. Inset shows representative electropherogram showing baseline separation of a mixture of 4 compounds. Adapted with permission from [74] Copyright © 2017 Elsevier.
Figure 6Examples of radiation detection methods (A) (left) Setup of microfluidic chip and silicon photomultiplier (SiPM) array for measurement of radioactivity and half-life of a sample. (right) Photograph of the final chip setup on the SiPM array. Adapted from [107] with permission of the Royal Society of Chemistry. (B) Device for positron detection from blood samples in a microchannel bonded to a silicon PIN photodiode. Adapted from [108] with permission of the Royal Society of Chemistry. (C) Dual silicon PIN photodiode detection of a sample in a channel or tubing. Photodiode 1 responds to both positrons and gamma rays, while photodiode 2 responds only to gamma radiation, allowing subtraction of the background gamma signal. Adapted with permission from [109]. Copyright © 2016 Springer. (D) Scintillator based detection. A scintillator was placed in close proximity to a PDMS microfluidic chip and scintillation light from positron interactions was detected with a CCD camera. Adapted with permission from [110] Copyright © 2007 IEEE. (E) Radioluminescence imaging. Scintillator light output is observed via sensitive microscope as radioactive decays occur in sample. Adapted with permission from [111] Copyright © 2015 John Wiley & Sons, Inc. (F) Cerenkov imaging. (left) Cerenkov radiation is emitted as energetic particles travel through the liquid or chip material. (right) Bright-field and Cerenkov images of 200 µm-wide microchannel filled with [18F]FDG solution. Adapted from [112] with permission of IOP PUBLISHING, Ltd. (G) Beta-box imaging. (top left) Microfluidic chip containing multiple sample chambers is placed on the detector; (bottom left) photograph of position sensitive avalanche photodiode (PSAPD) detector; (right) resulting image when sample chambers are filled with varying concentrations of [18F]fluoride. Adapted from [113] with permission of IOP PUBLISHING, Ltd. (H) System for measuring radioactivity in a silica-based monolith column placed above a Medipix sensor. Adapted from [114] with permission of the Royal Society of Chemistry. (I) Microfluidic channel filled with liquid scintillator that acts an array of scintillation detectors connected by liquid waveguides to photodetectors. Adapted with permission from [115] Copyright © 2010 Elsevier. (J) Autoradiography (phosphor imaging system). (left) Autoradiography image of adsorbed radiolabeled peptide on the channel surfaces of a plastic microfluidic chip. Adapted from [116] with permission of The Royal Society of Chemistry. (right) Autoradiography image of [18F]fluoride trapped in the serpentine channel of an electrochemical cell. Adapted with permission from [117] Copyright © 2013 Elsevier.
Summary of required QC tests, the conventional method(s) used, and typical specifications. In addition, examples of microfluidic approaches to perform each test are listed. Note we have indicated in italics approaches that have not been demonstrated/proven, but may be possible in principle. Abbreviations: HPLC/UV = HPLC with UV absorbance detector, HPLC/RI = HPLC with refractive index detector, etc.
| QC Test | Conventional Method (s) | Typical Acceptance Criteria | Examples of Microfluidic Suitable Approaches |
|---|---|---|---|
| pH | pH indicator strips; electronic pH meter | 4.5 < pH < 8.5 | Absorption spectroscopy of sample + indicator [ Absorption spectroscopy of pH-sensitive surface [ Fluorescence emission from pH-sensitive dye/surface [ Hydrogel-based pH sensing (physical or electrical change) [ Electrochemical cell [ |
| Appearance (color/clarity) | Visual | Clear, colorless, particulate-free | Absorption spectroscopy of sample [ |
| Sterility | Short term: filter integrity test (e.g., bubble point test); long term: Bacterial culture | Long term: No bacterial growth observed | Fluorescent detection of “universal” dye that binds to bacteria [ Electrical impedance detection of individual bacteria [ |
| Bacterial endotoxin | LAL test | 175 EU/V | Variations of LAL assay (absorption spectroscopy detection [ Detection of endotoxin binding to surface [ MCE/fluorescence [ |
| Chemical identity/purity | HPLC/UV | Varies | MCE/UV [ Chip-HPLC/UV |
| Kryptofix 2.2.2 | Color spot test | <50 µg/mL (USP); 2.2 mg/V (EP) | Absorption spectroscopy of sample + indicator [ MCE/UV Chip-HPLC/UV |
| Residual organic solvents | Gas chromatography; HPLC/RI | Varies (e.g., MeCN 4.1 mg/day, EtOH 50 mg/day, DMSO 50 mg/day, DCM 6 mg/day, DMF8.8 mg/day) | MCE/RI Micro-GC |
| Radioactivity concentration | Dose calibrator | Varies | Solid-state detectors (SiPM [ Cerenkov imaging [ Scintillator-based detectors (CCD imaging [ |
| Radionuclidic identity | Half-life measurement with dose calibrator | Varies (e.g., 105-115 min for 18F-labled tracers) | Radiation detector (half-life measurement) [ Potentially some of radiation detectors listed under “Radioactivity concentration” can be used |
| Radionuclidic purity | Gamma spectrometer | Match expected energy spectrum | Potentially some of radiation detectors listed under “Radioactivity concentration” can be used |
| Radiochemical identity and purity | Radio-HPLC; radio-TLC | >95%; (>90% for [18F]FDG) | Porous silica monolith with Medipix positron detector array [ TLC plate with Timepix positron detector array [ MCE/positron detector Chip-HPLC/positron detector |
| Specific activity | Radio-HPLC and dose calibrator | Varies | MCE/UV and radioactivity measurement Chip-HPLC/UV and radioactivity measurement |