| Literature DB >> 15791338 |
Abstract
Quantitative data on the effect of the electrode geometry on the signal strength and the signal-to-noise ratio is given. The measurements are affected by the unavoidable presence of stray capacitance. Best results are achieved for short and narrow electrodes arranged in an antiparallel configuration and separated by a minimal gap, which determines the dimensions of the actual detection volume. Limits of detection between 150 and 250 microg l(-1) and separation efficiencies from 13,000 to 17,000 theoretical plates were achieved for six inorganic cations (NH(4)(+), K(+), Ca(2+), Na(+), Mg(2+)and Li(+)) with electrodes of 1 mm width and a detection gap of 0.5 mm (separation channel length: 7.5 cm) when operating the detector at 20 V(pp) and 500 kHz. The analyses of all major inorganic cations in tap and rain water samples were demonstrated for the first time in microchip electrophoresis with contactless conductivity detection.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15791338 DOI: 10.1039/b418845d
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799