| Literature DB >> 30340561 |
Hyo Kyung Lee1,2, Mee Kum Kim3,4.
Abstract
BACKGROUND: The purpose of this study is to compare the biometric parameters and intraocular lens (IOL) power calculation by a new swept-source optical coherence tomography (SS-OCT) biometer with those by a partial coherence interferometry (PCI) biometer.Entities:
Keywords: IOLMaster 700; Intraocular lens power calculation; Optical biometer; Swept-source optical coherence tomography
Mesh:
Year: 2018 PMID: 30340561 PMCID: PMC6194682 DOI: 10.1186/s12886-018-0936-6
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Comparison of biometric measurements between two devices (n = 150)
| Parameter | PCI biometer | SS-OCT biometer | ||||
|---|---|---|---|---|---|---|
| Mean ± SD | Range | Mean ± SD | Range | Difference | ||
| AL (mm) | 23.99 ± 1.61 | 21.23 30.47 | 23.98 ± 1.60 | 21.19 30.43 | 0.0098 ± 0.03 | < .0001 |
| Km (D) | 44.20 ± 1.53 | 40.99 48.36 | 44.24 ± 1.55 | 40.99 48.10 | −0.0365 ± 0.20 | .03 |
| J0 (D) | −0.007 ± 0.33 | −0.995 0.772 | − 0.005 ± 0.32 | −0.794 0.811 | − 0.0019 ± 0.44 | .96 |
| J45 (D) | 0.008 ± 0.35 | −1.117 0.908 | −0.022 ± 0.33 | −0.789 1.041 | 0.0309 ± 0.46 | .41 |
| ACD (mm) | 3.11 ± 0.41 | 2.14 4.35 | 3.08 ± 0.41 | 2.03 4.15 | 0.0232 ± 0.15 | .06 |
PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography, SD standard deviation, AL axial length, Km mean keratometry, ACD anterior chamber depth
*Paired t-test
Fig. 1Bland-Altman plots for optical parameters. Bland-Altman plots of the differences in AL (a), Km (b), J0 (c), J45 (d) and ACD (e) between the SS-OCT and PCI biometers. The middle line indicates the mean difference. The top and bottom dashed lines show the upper and lower 95% LoAs, respectively (AL = axial length; Km = mean keratometry; ACD = anterior chamber depth; SS-OCT = swept-source optical coherence tomography; PCI = partial coherence interferometry; LoA = limit of agreement)
Comparison of refractive mean errora at 1 month after the surgery
| Formula | PCI biometer | SS-OCT biometer | |
|---|---|---|---|
| Tecnis ZCB00 ( | |||
| SRK-T | − 0.128 ± 0.42 | −0.136 ± 0.40 | .59 |
| Hoffer Q | −0.036 ± 0.46 | −0.028 ± 0.44 | .65 |
| Holladayb | − 0.070 ± 0.44 | −0.157 ± 0.34 | .0003 |
| Haigis | −0.046 ± 0.46 | −0.038 ± 0.41 | .69 |
| AcrySof SA60AT ( | |||
| SRK-T | 0.008 ± 0.53 | − 0.019 ± 0.52 | .31 |
| Hoffer Q | 0.162 ± 0.51 | 0.126 ± 0.52 | .28 |
| Holladayb | 0.127 ± 0.51 | 0.069 ± 0.46 | .14 |
| Haigis | 0.128 ± 0.53 | 0.115 ± 0.47 | .71 |
PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography
*Paired t-test
aRefractive mean error = postoperative spherical equivalent (SE) – predicted SE
bHolladay 1 and 2 formulas were embedded in PCI and SS-OCT biometers, respectively
Comparison of refractive mean absolute errora at 1 month after the surgery
| PCI biometer | SS-OCT biometer | ||
|---|---|---|---|
| Tecnis ZCB00 ( | |||
| SRK-T | 0.339 ± 0.28 | 0.334 ± 0.26 | .69 |
| Hoffer Q | 0.364 ± 0.28 | 0.352 ± 0.26 | .49 |
| Holladayb | 0.353 ± 0.27 | 0.334 ± 0.27 | .40 |
| Haigis | 0.378 ± 0.26 | 0.335 ± 0.24 | .03 |
| AcrySof SA60AT ( | |||
| SRK-T | 0.399 ± 0.35 | 0.405 ± 0.31 | .79 |
| Hoffer Q | 0.409 ± 0.33 | 0.401 ± 0.35 | .77 |
| Holladayb | 0.410 ± 0.33 | 0.355 ± 0.30 | .02 |
| Haigis | 0.417 ± .034 | 0.372 ± 0.31 | .09 |
PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography
*Paired t-test
aRefractive mean absolute error = l postoperative spherical equivalent (SE) – predicted SE l
bHolladay 1 and 2 formulas were embedded in PCI and SS-OCT biometers, respectively
Comparison of examination failure rate between the two devices
| Number of eyes | Failure rate (%) | |||
|---|---|---|---|---|
| Failure | Success | Total | ||
| PCI biometer | 25 | 150 | 175 | 14.29 |
| SS-OCT biometer | 2 | 173 | 175 | 1.14 |
PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography
Types of Cataract in eyes that failed to be examined by the devices
| Type of cataract | Number of eyes | |
|---|---|---|
| PCI biometer | SS-OCT biometer | |
| Cortical opacity | 1 | 0 |
| ASC | 2 | 0 |
| PSC | 18 | 0 |
| ASC with PSC | 2 | 0 |
| Total white cataract | 2 | 2 |
| Total | 25 | 2 |
PCI partial coherence interferometry, SS-OCT swept-source optical coherence tomography, ASC anterior subcapsular cataract, PSC posterior subcapsular cataract