BACKGROUND AND OBJECTIVE: To compare the postoperative refractive outcomes between IOLMaster biometry (Carl Zeiss Meditec, Inc., Dublin, CA) and immersion ultrasound biometry for axial length measurements. PATIENTS AND METHODS: Refractive outcomes in 354 eyes were compared using the IOLMaster and the immersion ultrasound biometry. Predicted refraction was determined using manual keratometry and the SRK-T formula with personalized A-constant. RESULTS: The axial lengths measured using the IOLMaster and immersion ultrasound were 24.49 ± 2.11 and 24.46 ± 2.11 mm, respectively, and the difference was significant (P < .05). The mean errors were 0.000 ± 0.578 D with the IOLMaster, and 0.000 ± 0.599 D with the immersion ultrasound, but the difference was not significant. The mean absolute error was smaller with the IOLMaster than with immersion ultrasound (0.463 ± 0.341 vs 0.479 ± 0.359 D), but the difference was not significant. CONCLUSION: IOLMaster biometry yields highly accurate results in cataract surgery. However, if the IOLMaster is unavailable, immersion ultrasound biometry with personalized intraocular lens constants is an acceptable alternative. Copyright 2012, SLACK Incorporated.
BACKGROUND AND OBJECTIVE: To compare the postoperative refractive outcomes between IOLMaster biometry (Carl Zeiss Meditec, Inc., Dublin, CA) and immersion ultrasound biometry for axial length measurements. PATIENTS AND METHODS: Refractive outcomes in 354 eyes were compared using the IOLMaster and the immersion ultrasound biometry. Predicted refraction was determined using manual keratometry and the SRK-T formula with personalized A-constant. RESULTS: The axial lengths measured using the IOLMaster and immersion ultrasound were 24.49 ± 2.11 and 24.46 ± 2.11 mm, respectively, and the difference was significant (P < .05). The mean errors were 0.000 ± 0.578 D with the IOLMaster, and 0.000 ± 0.599 D with the immersion ultrasound, but the difference was not significant. The mean absolute error was smaller with the IOLMaster than with immersion ultrasound (0.463 ± 0.341 vs 0.479 ± 0.359 D), but the difference was not significant. CONCLUSION: IOLMaster biometry yields highly accurate results in cataract surgery. However, if the IOLMaster is unavailable, immersion ultrasound biometry with personalized intraocular lens constants is an acceptable alternative. Copyright 2012, SLACK Incorporated.