INTRODUCTION: Cortical bone is an important anatomical structure and its thickness needs to be determined prior to many dental procedures to ensure treatment success. Imaging modalities are necessarily used in dentistry for treatment planning and dental procedures. Three-dimensional image reconstruction not only provides visual information but also enables accurate measurement of anatomical structures; thus, it is necessarily required for maxillofacial examination and in case of skeletal problems in this region. AIMS: This study aimed to assess the ability of three Cone Beam Computed Tomography (CBCT) systems including Cranex 3D, NewTom 3G and 3D Promax for Three-Dimensional (3D) image reconstruction of the cortical plate with variable thicknesses.Entities:
Keywords: Cone-Beam Computed Tomography (CBCT); Cortical bone; Image processing, Computer-assisted; Imaging, Three-dimensional
Year: 2018 PMID: 30288182 PMCID: PMC6142658 DOI: 10.2174/1874210601812010586
Source DB: PubMed Journal: Open Dent J ISSN: 1874-2106
Comparison and level of agreement between different CBCT systems for 3D visualization of cortical bone with thicknesses 0.5 mm or less.
| System 1 | System 2 | ≤ 0.5 mm | |
|---|---|---|---|
| Kappa |
| ||
| Soredex FOV:4×6 cm2 | Soredex FOV: 6×8cm2 | 0.800 | <0.001 |
| Planmeca FOV: 8×8cm2 | 0.029 | 0.555 | |
| NewTom FOV:5×5cm2 | 0.400 | 0.046 | |
| NewTom FOV:8×5cm2 | 0.241 | 0.070 | |
| NewTom FOV:8×8cm2 | 0.125 | 0.206 | |
| Soredex FOV: 6×8 cm2 | Planmeca FOV: 8×8cm2 | 0.043 | 0.47 |
| NewTom FOV:5×5cm2 | 0.438 | 0.032 | |
| NewTom FOV:8×5cm2 | 0.341 | 0.026 | |
| NewTom FOV:8×8cm2 | 0.182 | 0.121 | |
| Planmeca FOV: 8×8 cm2 | NewTom FOV:5×5cm2 | 0.043 | 0.47 |
| NewTom FOV:8×5cm2 | -0.079 | 0.512 | |
| NewTom FOV:8×8cm2 | -0.071 | 0.648 | |
| NewTom FOV:5×5 cm2 | NewTom FOV:8×5cm2 | 0.341 | 0.026 |
| NewTom FOV:8×8cm2 | 0.182 | 0.121 | |
| NewTom FOV:8×5 cm2 | NewTom FOV:11×13cm2 | 0.654 | 0.001 |
Results of Kappa statistic (P-value) for assessing the intra and inter observer agreements.
| System | Intra-Observer Agreement for the First Observer | Intra-Observer Agreement for the Second Observer | Inter-Observer Agreement |
|---|---|---|---|
| Soredex FOV: 4×6cm2 | 1.000 (<0.001) | 1.000 (<0.001) | 1.000 (<0.001) |
| Soredex FOV: 6×8cm2 | 0.926 (<0.001) | 0.920 (<0.001) | 1.000 (<0.001) |
| Planmeca FOV: 8×8cm2 | 0.970 (<0.001) | 0.968 (<0.001) | 0.969 (<0.001) |
| NewTom FOV: 5×5cm2 | 0.800 (<0.001) | 1.000 (<0.001) | 0.804 (<0.001) |
| NewTom FOV: 8×5cm2 | 0.961 (<0.001) | 0.959 (<0.001) | 0.963 (<0.001) |
| NewTom FOV: 8×8cm2 | 1.000 (<0.001) | 0.898 (<0.001) | 0.866(<0.001) |
Descriptive statistics for cortical thickness in its different categories.
| Cortical Thickness(mm) | N | Mean | Std. Deviation | Minimum | Maximum |
| 0-0.5 | 24 | .400000 | .0884652 | .2000 | .5000 |
| 0.6-1 | 24 | .729167 | .1122078 | .6000 | .9000 |
| 1.1-1.5 | 24 | 1.295833 | .1781039 | 1.1000 | 1.5000 |
| Total | 72 | .808333 | .3945937 | .2000 | 1.5000 |
Frequency (percentage) of visibility of cortical bone.
| System | ≤ 0.5 mm | 0.6-1 mm | >1 mm |
|---|---|---|---|
| Soredex FOV: 4×6cm2 | 18 (75%) | 24 (100%) | 24 (100%) |
| Soredex FOV: 6×8cm2 | 16 (66.7%) | 24 (100%) | 24 (100%) |
| Planmeca FOV: 8×8cm2 | 1 (4.2%) | 22 (91.7%) | 24 (100%) |
| NewTom FOV: 5×5cm2 | 16 (66.7%) | 24 (100%) | 24 (100%) |
| NewTom FOV: 8×5cm2 | 7 (29.2%) | 24 (100%) | 24 (100%) |
| NewTom FOV: 8×8cm2 | 4 (16.7%) | 23 (95.8%) | 24 (100%) |