Accuracy of measuring the cortical bone thickness adjacent to dental implants using cone beam computed tomography.

Touraj Razavi1, Richard M Palmer, Jonathan Davies, Ron Wilson, Paul J Palmer.   

Abstract

OBJECTIVE: To assess the accuracy of measuring the cortical bone thickness adjacent to dental implants using two cone beam computed tomography (CBCT) systems.
MATERIAL AND METHODS: Ten 4 x 11 mm Astra Tech implants were placed at varying distances from the cortical bone in two prepared bovine ribs. Both ribs were scanned in a reproducible position using two different CBCT scanners. Ten examiners each carried out four measurements on all 10 implants using the two CBCT systems: vertical distance between the top of the implant and the alveolar crest (IT-AC), and thickness of the cortical bone from the outer surface of the implant threads at 3, 6 and 9 mm from the top of the implant. Ground sections were prepared and bone thickness was measured using a light microscope and a graticule to give a gold standard (GS) measurement.
RESULTS: The examiner's measurements were significantly different between CBCT systems for the vertical and thickness dimensions (P<0.001) while measuring the cortical bone thickness between 0.3 and 3.7 mm. Within that range, i-CAT NG measurements were consistently underestimated in comparison with the GS. Accuitomo 3D60 FPD measurements closely approximated the GS, except when cortical bone thickness was <0.8 mm. The mean percentage errors from the GS at 3, 6 and 9 mm measurement levels were 68%, 28% and 18%, respectively, for i-CAT NG and 23%, 5% and 6%, respectively, for Accuitomo 3D60 FPD.
CONCLUSIONS: Within the limitations of this study, it was concluded that i-CAT NG (voxel size 0.3) may not produce sufficient resolution of the thin cortical bone adjacent to dental implants and, therefore, the measurements may not be accurate; whereas, Accuitomo 3D60 FPD (voxel size 0.125) may produce better resolution and more accurate measurement of the thin bone.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20636726     DOI: 10.1111/j.1600-0501.2009.01905.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  36 in total

1.  Accuracy of cone-beam computed tomography to assess the configuration and extent of ligature-induced peri-implantitis defects. A pilot study.

Authors:  Vladimir Golubovic; Ilja Mihatovic; Jürgen Becker; Frank Schwarz
Journal:  Oral Maxillofac Surg       Date:  2012-04-04

2.  Dose reduction of cone beam CT scanning for the entire oral and maxillofacial regions with thyroid collars.

Authors:  X M Qu; G Li; G C H Sanderink; Z Y Zhang; X C Ma
Journal:  Dentomaxillofac Radiol       Date:  2012-07       Impact factor: 2.419

3.  Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode.

Authors:  K Kamburoglu; E Kolsuz; S Murat; H Eren; S Yüksel; C S Paksoy
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

4.  Comparison of mesiodistal space measurements of single-implant sites on panoramic and oblique images generated by cone-beam CT.

Authors:  Mustafa Alkhader; Malik Hudieb
Journal:  Surg Radiol Anat       Date:  2013-08-27       Impact factor: 1.246

5.  Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography.

Authors:  V Hirschinger; S Hanke; U Hirschfelder; E Hofmann
Journal:  J Orofac Orthop       Date:  2015-03       Impact factor: 1.938

6.  Accuracy of linear measurements around dental implants by means of cone beam computed tomography with different exposure parameters.

Authors:  Lauren O L Bohner; Pedro Tortamano; Juliana Marotti
Journal:  Dentomaxillofac Radiol       Date:  2017-03-23       Impact factor: 2.419

7.  Accuracy of peri-implant bone thickness and validity of assessing bone augmentation material using cone beam computed tomography.

Authors:  Dongyun Wang; Andreas Künzel; Vladimir Golubovic; Ilya Mihatovic; Gordon John; Zhuofan Chen; Jürgen Becker; Frank Schwarz
Journal:  Clin Oral Investig       Date:  2012-10-12       Impact factor: 3.573

8.  Accuracy of CBCT images in the assessment of buccal marginal alveolar peri-implant defects: effect of field of view.

Authors:  K Kamburoğlu; S Murat; C Kılıç; S Yüksel; H Avsever; A Farman; W C Scarfe
Journal:  Dentomaxillofac Radiol       Date:  2014-03-20       Impact factor: 2.419

9.  Buccal cortical bone thickness at miniscrew placement sites in patients with different vertical skeletal patterns.

Authors:  Ilknur Veli; Tancan Uysal; Asli Baysal; Irfan Karadede
Journal:  J Orofac Orthop       Date:  2014-10-26       Impact factor: 1.938

10.  Cone-beam computed tomography and microtomography for alveolar bone measurements.

Authors:  Nathália Ferrare; André Ferreira Leite; Hugo César Pinto Marques Caracas; Ricardo Bentes de Azevedo; Nilce Santos de Melo; Paulo Tadeu de Souza Figueiredo
Journal:  Surg Radiol Anat       Date:  2013-02-12       Impact factor: 1.246

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.