Literature DB >> 19423257

A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT). Part II: On 3D model accuracy.

Xin Liang1, Ivo Lambrichts, Yi Sun, Kathleen Denis, Bassam Hassan, Limin Li, Ruben Pauwels, Reinhilde Jacobs.   

Abstract

AIM: The study aim was to compare the geometric accuracy of three-dimensional (3D) surface model reconstructions between five Cone Beam Computed Tomography (CBCT) scanners and one Multi-Slice CT (MSCT) system.
MATERIALS AND METHODS: A dry human mandible was scanned with five CBCT systems (NewTom 3G, Accuitomo 3D, i-CAT, Galileos, Scanora 3D) and one MSCT scanner (Somatom Sensation 16). A 3D surface bone model was created from the six systems. The reference (gold standard) 3D model was obtained with a high resolution laser surface scanner. The 3D models from the five systems were compared with the gold standard using a point-based rigid registration algorithm.
RESULTS: The mean deviation from the gold standard for MSCT was 0.137 mm and for CBCT were 0.282, 0.225, 0.165, 0.386 and 0.206 mm for the i-CAT, Accuitomo, NewTom, Scanora and Galileos, respectively.
CONCLUSION: The results show that the accuracy of CBCT 3D surface model reconstructions is somewhat lower but acceptable comparing to MSCT from the gold standard. Copyright (c) 2009 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19423257     DOI: 10.1016/j.ejrad.2009.04.016

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  53 in total

1.  Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site.

Authors:  A Dawood; J Brown; V Sauret-Jackson; S Purkayastha
Journal:  Dentomaxillofac Radiol       Date:  2011-12-19       Impact factor: 2.419

2.  [Rhinosinusitis guidelines--unabridged version: S2 guidelines from the German Society of Otorhinolaryngology, Head and Neck Surgery].

Authors:  B A Stuck; C Bachert; P Federspil; W Hosemann; L Klimek; R Mösges; O Pfaar; C Rudack; H Sitter; M Wagenmann; R Weber; K Hörmann
Journal:  HNO       Date:  2012-02       Impact factor: 1.284

3.  Comparison of cone-beam and conventional multislice computed tomography for image-guided dental implant planning.

Authors:  Paul W Poeschl; Nina Schmidt; Godoberto Guevara-Rojas; Rudolf Seemann; Rolf Ewers; Harald T Zipko; Kurt Schicho
Journal:  Clin Oral Investig       Date:  2012-03-14       Impact factor: 3.573

Review 4.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

5.  Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: a tale of two CTs.

Authors:  R Patcas; G Markic; L Müller; O Ullrich; T Peltomäki; C J Kellenberger; C A Karlo
Journal:  Dentomaxillofac Radiol       Date:  2012-05-03       Impact factor: 2.419

6.  The effect of scan parameters on cone beam CT trabecular bone microstructural measurements of the human mandible.

Authors:  N Ibrahim; A Parsa; B Hassan; P van der Stelt; I H A Aartman; D Wismeijer
Journal:  Dentomaxillofac Radiol       Date:  2013-10-16       Impact factor: 2.419

7.  MR-Imaging of teeth and periodontal apparatus: an experimental study comparing high-resolution MRI with MDCT and CBCT.

Authors:  Chiara Gaudino; Raluca Cosgarea; Sabine Heiland; Réka Csernus; Bruno Beomonte Zobel; Mirko Pham; Ti-Sun Kim; Martin Bendszus; Stefan Rohde
Journal:  Eur Radiol       Date:  2011-07-31       Impact factor: 5.315

8.  Accuracy of anatomical landmark identification using different CBCT- and MSCT-based 3D images: an in vitro study.

Authors:  Jürgen Medelnik; Klaus Hertrich; Stefanie Steinhäuser-Andresen; Ursula Hirschfelder; Elisabeth Hofmann
Journal:  J Orofac Orthop       Date:  2011-09-08       Impact factor: 1.938

9.  3D dento-maxillary osteolytic lesion and active contour segmentation pilot study in CBCT: semi-automatic vs manual methods.

Authors:  K Vallaeys; A Kacem; H Legoux; M Le Tenier; C Hamitouche; R Arbab-Chirani
Journal:  Dentomaxillofac Radiol       Date:  2015-05-21       Impact factor: 2.419

Review 10.  Measuring and Establishing the Accuracy and Reproducibility of 3D Printed Medical Models.

Authors:  Elizabeth George; Peter Liacouras; Frank J Rybicki; Dimitrios Mitsouras
Journal:  Radiographics       Date:  2017-08-11       Impact factor: 5.333

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.