Literature DB >> 30286620

Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.

Stefan Pleus1, Nina Jendrike1, Annette Baumstark1, Jochen Mende1, Cornelia Haug1, Guido Freckmann1.   

Abstract

INTRODUCTION: Blood glucose monitoring systems (BGMS) should provide sufficient analytical quality to allow adequate therapy for diabetes patients. Besides system accuracy, measurement precision is an important aspect of a BGMS' analytical quality.
METHODS: Based on ISO 15197:2013/EN ISO 15197:2015, system accuracy, measurement repeatability, and intermediate measurement precision were assessed. ISO 15197:2013 system accuracy criteria require that ⩾95% of individual BGMS' test strip lot results shall fall within ±15 mg/dl or ±15% of corresponding comparison method results (at glucose concentrations <100 mg/dl and ⩾100 mg/dl, respectively), and that ⩾99% of results fall within consensus error grid (CEG) zones A and B. Measurement repeatability was assessed using venous blood samples, whereas intermediate measurement precision was assessed using control solution samples. Standard deviation (SD) and coefficient of variation (CV) were calculated for glucose concentrations <100 mg/dl and ⩾100 mg/dl, respectively. Precision acceptance criteria are not specified by ISO 15197:2013.
RESULTS: All three BGMS fulfilled system accuracy criteria with 96% to 98% of individual test strip lot's results falling within the acceptable accuracy limits. All measurement results fell within CEG zones A and B. For measurement repeatability, SD was ⩽3.3 mg/dl, and CV was ⩽3.9% for the investigated BGMS. Assessment of intermediate measurement precision showed SD ⩽1.3 mg/dl and CV ⩽3.0%.
CONCLUSION: All three BGMS fulfilled system accuracy criteria of ISO 15197:2013. In absence of acceptance criteria, precision results were found to be consistent with the manufacturer's labeling of the investigated devices.

Entities:  

Keywords:  ISO 15197; intermediate measurement precision; measurement repeatability; system accuracy

Year:  2018        PMID: 30286620      PMCID: PMC6313280          DOI: 10.1177/1932296818804837

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  17 in total

1.  Partial pressure of oxygen in capillary blood samples from the fingertip.

Authors:  Guido Freckmann; Christina Schmid; Annette Baumstark; Stefan Pleus; Manuela Link; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-11-01

2.  Accuracy and User Performance Evaluation of a New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device.

Authors:  Timothy S Bailey; Jane F Wallace; Scott Pardo; Mary Ellen Warchal-Windham; Bern Harrison; Robert Morin; Mark Christiansen
Journal:  J Diabetes Sci Technol       Date:  2017-02-01

3.  Performance of 10 systems for self-monitoring of blood glucose by trained healthcare professionals and in the hands of the users.

Authors:  Una Ø Sølvik; Marianne Risa; Camilla E Jacobsen; Grete Monsen; Sverre Sandberg
Journal:  Clin Chem       Date:  2015-03-03       Impact factor: 8.327

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.

Authors:  Annette Baumstark; Nina Jendrike; Christina Liebing; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-01-01

6.  Evaluation of Four Blood Glucose Monitoring Systems for Self-Testing with Built-in Insulin Dose Advisor Based on ISO 15197:2013: System Accuracy and Hematocrit Influence.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Christina Liebing; Alexandra Beer; Frank Flacke; Cornelia Haug; Guido Freckmann
Journal:  Diabetes Technol Ther       Date:  2018-04       Impact factor: 6.118

7.  Accuracy Evaluation of Four Blood Glucose Monitoring Systems in the Hands of Intended Users and Trained Personnel Based on ISO 15197 Requirements.

Authors:  Guido Freckmann; Annette Baumstark; Nina Jendrike; Delia Rittmeyer; Stefan Pleus; Cornelia Haug
Journal:  Diabetes Technol Ther       Date:  2017-02-22       Impact factor: 6.118

8.  Accuracy and precision assessment of a new blood glucose monitoring system.

Authors:  Li-Nong Ji; Li-Xin Guo; Li-Bin Liu
Journal:  Clin Chem Lab Med       Date:  2016-01       Impact factor: 3.694

9.  Assessing System Accuracy of Blood Glucose Monitoring Systems Using Rectangle Target Plots.

Authors:  Peter Müller; Andrew Hattemer; Peter Stephan
Journal:  J Diabetes Sci Technol       Date:  2015-10-09

10.  Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose.

Authors:  Hae-Il Park; Seong-Su Lee; Jang-Won Son; Hee-Sun Kwon; Sung Rae Kim; Hyojin Chae; Myungshin Kim; Yonggoo Kim; Soonjib Yoo
Journal:  J Clin Lab Anal       Date:  2016-04-13       Impact factor: 2.352

View more
  2 in total

1.  Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-12-14

Review 2.  Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.

Authors:  Jessica Hanae Zafra-Tanaka; David Beran; Beatrice Vetter; Rangarajan Sampath; Antonio Bernabe-Ortiz
Journal:  J Diabetes Sci Technol       Date:  2021-03-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.