Accuracy and precision assessment of a new blood glucose monitoring system.

Li-Nong Ji, Li-Xin Guo, Li-Bin Liu.   

Abstract

BACKGROUND: Blood glucose self-monitoring by individuals with diabetes is essential in controlling blood glucose levels. The International Organization for Standardization (ISO) introduced new standards for blood glucose monitoring systems (BGMS) in 2013 (ISO 15197: 2013). The CONTOUR PLUS® (CONTOUR PLUS) BGMS has been demonstrated to meet the 2013 ISO standards; however, no Chinese data on CONTOUR PLUS accuracy and precision have been published.
METHODS: This study evaluated the accuracy and precision of CONTOUR PLUS BGMS in quantitative glucose testing of capillary and venous whole blood samples obtained from 363 patients at three different hospitals.
RESULTS: Results of fingertip and venous blood glucose measurements by the CONTOUR PLUS system were compared with laboratory reference values to determine accuracy. Accuracy was 98.1% (96.06%-99.22%) for fingertip blood tests and 98.1% (96.02%-99.21%) for venous blood tests. Precision was evaluated across a wide range of blood glucose values (5.1-17.2 mmol/L), testing three blood samples repeatedly 15 times with the CONTOUR PLUS blood glucose meter using test strips from three lots. All within-lot results met ISO criteria (i.e., SD<0.42 mmol/L for blood glucose concentration <5.55 mmol/L; CV<7.5% for blood glucose concentration ≥5.55 mmol/L). Between-lot variations were 1.5% for low blood glucose concentration, 2.4% for normal and 3.4% for high.
CONCLUSIONS: Accuracy of both fingertip and venous blood glucose measurements by the CONTOUR PLUS system was >95%, confirming that the system meets ISO 15197: 2013 requirements.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26053011     DOI: 10.1515/cclm-2015-0130

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-12-14

2.  Accuracy of capillary blood glucose for 50-g glucose challenge test for gestational diabetes screening.

Authors:  Dittakarn Boriboonhirunsarn; Amporn Robkhonburi; Maria Asad-Dehghan
Journal:  Diabetol Int       Date:  2022-02-07

3.  Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.

Authors:  Stefan Pleus; Nina Jendrike; Annette Baumstark; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.