System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.

Annette Baumstark1, Nina Jendrike1, Christina Liebing1, Cornelia Haug1, Guido Freckmann1.   

Abstract

An improved test cassette for the integrated Accu-Chek® Mobile system (Roche Diabetes Care GmbH, Mannheim, Germany) has been developed. System accuracy of this improved system was evaluated based on ISO 15197:2013, clause 6.3, for three reagent system lots. According to this standard, at least 95% of the system's measurement results shall be within ±15 mg/dL and ±15% of the results of the comparison method at glucose concentrations <100 mg/dL and ≥100 mg/dL (accuracy criterion A), respectively, and at least 99% of results shall be within consensus error grid zones A and B (accuracy criterion B). In addition, accuracy was evaluated in the hands of users based on ISO 15197:2013, clause 8, with one reagent system lot.

Entities:  

Keywords:  EN ISO 15197:2015; ISO 15197:2013; blood glucose monitoring systems; integrated system; self-monitoring of blood glucose; user performance

Mesh:

Substances:

Year:  2018        PMID: 29291625      PMCID: PMC5851233          DOI: 10.1177/1932296817752958

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  12 in total

1.  Influence of fruit juice on fingertips and patient behavior on self-monitoring of blood glucose.

Authors:  Masayuki Arakawa; Chie Ebato
Journal:  Diabetes Res Clin Pract       Date:  2012-02-18       Impact factor: 5.602

2.  Accuracy Evaluation of an Integrated Blood Glucose Monitoring System With Improved Test Cassettes Following ISO 15197:2013.

Authors:  Annette Baumstark; Manfred Kraft; Andrew Hattemer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2015-10-01

3.  Standardized evaluation of nine instruments for self-monitoring of blood glucose.

Authors:  Gunn B B Kristensen; Grete Monsen; Svein Skeie; Sverre Sandberg
Journal:  Diabetes Technol Ther       Date:  2008-12       Impact factor: 6.118

Review 4.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

5.  Performance of 10 systems for self-monitoring of blood glucose by trained healthcare professionals and in the hands of the users.

Authors:  Una Ø Sølvik; Marianne Risa; Camilla E Jacobsen; Grete Monsen; Sverre Sandberg
Journal:  Clin Chem       Date:  2015-03-03       Impact factor: 8.327

6.  Performance of a new blood glucose monitoring system in the hands of intended users.

Authors:  Timothy S Bailey; Jane F Wallace; Joan Lee Parkes; Scott Pardo; Ji Yu; Holly C Schachner; David A Simmons; Amy Chu
Journal:  Diabetes Technol Ther       Date:  2012-06-29       Impact factor: 6.118

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Effect of adverse storage conditions on performance of glucometer test strips.

Authors:  Richard Bamberg; Kathleen Schulman; Melissa MacKenzie; Joanna Moore; Sarah Olchesky
Journal:  Clin Lab Sci       Date:  2005

9.  Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician.

Authors:  Svein Skeie; Geir Thue; Kari Nerhus; Sverre Sandberg
Journal:  Clin Chem       Date:  2002-07       Impact factor: 8.327

10.  Glucose monitoring after fruit peeling: pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: wash your hands with tap water before you check blood glucose level.

Authors:  Takahisa Hirose; Tomoya Mita; Yoshio Fujitani; Ryuzo Kawamori; Hirotaka Watada
Journal:  Diabetes Care       Date:  2011-01-31       Impact factor: 19.112

View more
  3 in total

1.  Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.

Authors:  Nina Jendrike; Annette Baumstark; Stefan Pleus; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-12-14

2.  Clinical Study of a High Accuracy Green Design Blood Glucose Monitor Using an Innovative Optical Transmission Absorbance System.

Authors:  Takeyuki Moriuchi; Yuto Otaki; Hiroya Satou; Fumihiko Chai; Yuma Hayashida; Ryokei Aikawa; Takayuki Sugiyama; Koji Sode
Journal:  J Diabetes Sci Technol       Date:  2021-12-10

3.  Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.

Authors:  Stefan Pleus; Nina Jendrike; Annette Baumstark; Jochen Mende; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.