| Literature DB >> 29291625 |
Annette Baumstark1, Nina Jendrike1, Christina Liebing1, Cornelia Haug1, Guido Freckmann1.
Abstract
An improved test cassette for the integrated Accu-Chek® Mobile system (Roche Diabetes Care GmbH, Mannheim, Germany) has been developed. System accuracy of this improved system was evaluated based on ISO 15197:2013, clause 6.3, for three reagent system lots. According to this standard, at least 95% of the system's measurement results shall be within ±15 mg/dL and ±15% of the results of the comparison method at glucose concentrations <100 mg/dL and ≥100 mg/dL (accuracy criterion A), respectively, and at least 99% of results shall be within consensus error grid zones A and B (accuracy criterion B). In addition, accuracy was evaluated in the hands of users based on ISO 15197:2013, clause 8, with one reagent system lot.Entities:
Keywords: EN ISO 15197:2015; ISO 15197:2013; blood glucose monitoring systems; integrated system; self-monitoring of blood glucose; user performance
Mesh:
Substances:
Year: 2018 PMID: 29291625 PMCID: PMC5851233 DOI: 10.1177/1932296817752958