Technical Note: A benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system with a high-power x-ray source and transmission CT imaging capability.

Nivedh Manohar1, Francisco J Reynoso1, Sang Hyun Cho1,2.   

Abstract

PURPOSE: This report describes upgrades and performance characterization of an experimental benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system capable of determining the spatial distribution and concentration of metal probes such as gold nanoparticles (GNPs). Specifically, a high-power (~3 kW) industrial x-ray source and transmission CT capability were deployed in the same platform under the cone-beam geometry.
METHODS: All components of the system are described in detail, including the x-ray source, imaging stage, cadmium-telluride detector for XFCT, and flat-panel detector for transmission CT imaging. The general data acquisition scheme for XFCT and transmission CT is also explicated. The detection limit of the system was determined using calibration samples containing water and GNPs at various concentrations. Samples were then embedded in a small-animal-sized phantom and imaged with XFCT and CT. The reconstructed XFCT and CT images were compared and analyzed using the contrast-to-noise ratio for each GNP-containing region of interest. Also, measurements of the incident beam spectra used for XFCT and CT imaging were made and the corresponding x-ray dose rates were estimated, along with the imaging dose.
RESULTS: The present configuration produced a GNP detection limit of 0.03 wt. % with the delivery of an effective dose of 1.87 cGy per projection. XFCT scan of an animal-sized phantom containing low concentrations (down to 0.03 wt. %) of GNP-loaded inserts can be performed within an hour.
CONCLUSIONS: The high performance of the system combined with the ability to perform transmission CT in tandem with XFCT suggests that the currently developed benchtop cone-beam XFCT/CT system, in conjunction with GNPs, can be used for routine multimodal preclinical imaging tasks with less stringent dose constraints such as ex vivo imaging. With further effort to minimize XFCT imaging dose as discussed in this report, it may also be used for in vivo imaging.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  gold nanoparticles; multimodal imaging; preclinical imaging; x-ray fluorescence computed tomography

Mesh:

Year:  2018        PMID: 30125950      PMCID: PMC6181758          DOI: 10.1002/mp.13138

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Small-animal X-ray dose from micro-CT.

Authors:  John M Boone; Orlando Velazquez; Simon R Cherry
Journal:  Mol Imaging       Date:  2004-07       Impact factor: 4.488

2.  Improved signal-to-noise ratio for non-perpendicular detection angles in x-ray fluorescence computed tomography (XFCT).

Authors:  Martin Sjölin; Mats Danielsson
Journal:  Phys Med Biol       Date:  2014-10-13       Impact factor: 3.609

3.  Three-dimensional x-ray fluorescence mapping of a gold nanoparticle-loaded phantom.

Authors:  Liqiang Ren; Di Wu; Yuhua Li; Ge Wang; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

4.  Iterative methods for the three-dimensional reconstruction of an object from projections.

Authors:  P Gilbert
Journal:  J Theor Biol       Date:  1972-07       Impact factor: 2.691

5.  Sheet beam x-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticles.

Authors:  Chelsea A S Dunning; Magdalena Bazalova-Carter
Journal:  Med Phys       Date:  2018-04-16       Impact factor: 4.071

6.  A quantitative x-ray detection system for gold nanoparticle tumour biomarkers.

Authors:  K Ricketts; A Castoldi; C Guazzoni; C Ozkan; C Christodoulou; A P Gibson; G J Royle
Journal:  Phys Med Biol       Date:  2012-08-08       Impact factor: 3.609

7.  X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays.

Authors:  Seong-Kyun Cheong; Bernard L Jones; Arsalan K Siddiqi; Fang Liu; Nivedh Manohar; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2010-01-13       Impact factor: 3.609

8.  Development of XFCT imaging strategy for monitoring the spatial distribution of platinum-based chemodrugs: instrumentation and phantom validation.

Authors:  Yu Kuang; Guillem Pratx; Magdalena Bazalova; Jianguo Qian; Bowen Meng; Lei Xing
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

9.  Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography.

Authors:  Nivedh Manohar; Francisco J Reynoso; Parmeswaran Diagaradjane; Sunil Krishnan; Sang Hyun Cho
Journal:  Sci Rep       Date:  2016-02-25       Impact factor: 4.379

10.  Evaluation of X-ray doses and their corresponding biological effects on experimental animals in cone-beam micro-CT scans (R-mCT2).

Authors:  Nobuyuki Miyahara; Toshiaki Kokubo; Yukihiro Hara; Ayuta Yamada; Takafumi Koike; Yoshinori Arai
Journal:  Radiol Phys Technol       Date:  2015-10-06
View more
  6 in total

1.  Characterization of a Pixelated Cadmium Telluride Detector System Using a Polychromatic X-Ray Source and Gold Nanoparticle-Loaded Phantoms for Benchtop X-Ray Fluorescence Imaging.

Authors:  Sandun Jayarathna; Md Foiez Ahmed; Liam O'Ryan; Hem Moktan; Yonggang Cui; Sang Hyun Cho
Journal:  IEEE Access       Date:  2021-03-29       Impact factor: 3.367

2.  Investigation of transmission computed tomography (CT) image quality and x-ray dose achievable from an experimental dual-mode benchtop x-ray fluorescence CT and transmission CT system.

Authors:  Luzhen Deng; Selcuk Yasar; Md Foiez Ahmed; Sandun Jayarathna; Peng Feng; Biao Wei; Srinivasan Vedantham; Andrew Karellas; Sang Hyun Cho
Journal:  J Xray Sci Technol       Date:  2019       Impact factor: 1.535

3.  A detector's eye view (DEV)-based OSEM algorithm for benchtop x-ray fluorescence computed tomography (XFCT) image reconstruction.

Authors:  Luzhen Deng; Md F Ahmed; Sandun Jayarathna; Peng Feng; Biao Wei; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2019-04-08       Impact factor: 3.609

4.  Sensitivity enhancement of an experimental benchtop x-ray fluorescence imaging system by deploying a single crystal cadmium telluride detector system optimized for high flux x-ray operations.

Authors:  Hem Moktan; Sandun Jayarathna; Sang Hyun Cho
Journal:  Nucl Instrum Methods Phys Res A       Date:  2021-12-13       Impact factor: 1.455

5.  Quantitative X-ray fluorescence imaging of gold nanoparticles using joint L1 and total variation regularized reconstruction.

Authors:  Junwei Shi; Blaine Granger; Keying Xu; Yidong Yang
Journal:  Quant Imaging Med Surg       Date:  2020-01

6.  Bias-voltage dependent operational characteristics of a fully spectroscopic pixelated cadmium telluride detector system within an experimental benchtop x-ray fluorescence imaging setup.

Authors:  Hem Moktan; Raj Kumar Panta; Sang Hyun Cho
Journal:  Biomed Phys Eng Express       Date:  2021-12-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.