Literature DB >> 23464279

Development of XFCT imaging strategy for monitoring the spatial distribution of platinum-based chemodrugs: instrumentation and phantom validation.

Yu Kuang, Guillem Pratx, Magdalena Bazalova, Jianguo Qian, Bowen Meng, Lei Xing.   

Abstract

PURPOSE: Developing an imaging method to directly monitor the spatial distribution of platinum-based (Pt) drugs at the tumor region is of critical importance for early assessment of treatment efficacy and personalized treatment. In this study, the authors investigated the feasibility of imaging platinum (Pt)-based drug distribution using x-ray fluorescence (XRF, a.k.a. characteristic x ray) CT (XFCT).
METHODS: A 5-mm-diameter pencil beam produced by a polychromatic x-ray source equipped with a tungsten anode was used to stimulate emission of XRF photons from Pt drug embedded within a water phantom. The phantom was translated and rotated relative to the stationary pencil beam in a first-generation CT geometry. The x-ray energy spectrum was collected for 18 s at each position using a cadmium telluride detector. The spectra were then used for the K-shell XRF peak isolation and sinogram generation for Pt. The distribution and concentration of Pt were reconstructed with an iterative maximum likelihood expectation maximization algorithm. The capability of XFCT to multiplexed imaging of Pt, gadolinium (Gd), and iodine (I) within a water phantom was also investigated.
RESULTS: Measured XRF spectrum showed a sharp peak characteristic of Pt with a narrow full-width at half-maximum (FWHM) (FWHMKα1 = 1.138 keV, FWHMKα2 = 1.052 keV). The distribution of Pt drug in the water phantom was clearly identifiable on the reconstructed XRF images. Our results showed a linear relationship between the XRF intensity of Pt and its concentrations (R(2) = 0.995), suggesting that XFCT is capable of quantitative imaging. A transmission CT image was also obtained to show the potential of the approach for providing attenuation correction and morphological information. Finally, the distribution of Pt, Gd, and I in the water phantom was clearly identifiable in the reconstructed images from XFCT multiplexed imaging.
CONCLUSIONS: XFCT is a promising modality for monitoring the spatial distribution of Pt drugs. The technique may be useful in tailoring tumor treatment regimen in the future.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23464279      PMCID: PMC3585826          DOI: 10.1118/1.4789917

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  Analytic and iterative reconstruction algorithms in SPECT.

Authors:  Philippe P Bruyant
Journal:  J Nucl Med       Date:  2002-10       Impact factor: 10.057

2.  Tomographic molecular imaging of x-ray-excitable nanoparticles.

Authors:  Guillem Pratx; Colin M Carpenter; Conroy Sun; Ravi P Rao; Lei Xing
Journal:  Opt Lett       Date:  2010-10-15       Impact factor: 3.776

3.  Optimization of a polarized source for in vivo x-ray fluorescence analysis of platinum and other heavy metals.

Authors:  D G Lewis
Journal:  Phys Med Biol       Date:  1994-01       Impact factor: 3.609

4.  EM reconstruction algorithms for emission and transmission tomography.

Authors:  K Lange; R Carson
Journal:  J Comput Assist Tomogr       Date:  1984-04       Impact factor: 1.826

5.  A quantitative x-ray detection system for gold nanoparticle tumour biomarkers.

Authors:  K Ricketts; A Castoldi; C Guazzoni; C Ozkan; C Christodoulou; A P Gibson; G J Royle
Journal:  Phys Med Biol       Date:  2012-08-08       Impact factor: 3.609

Review 6.  Ovarian cancer, Part II: Treatment.

Authors:  R F Ozols
Journal:  Curr Probl Cancer       Date:  1992 Mar-Apr       Impact factor: 3.187

7.  Resistance to platinum-based chemotherapy in lung cancer cell lines.

Authors:  Jianli Chen; Nashwa Emara; Charalambos Solomides; Hemant Parekh; Henry Simpkins
Journal:  Cancer Chemother Pharmacol       Date:  2010-02-21       Impact factor: 3.333

8.  Plane polarized x-ray fluorescence system for the in vivo measurement of platinum in head and neck tumours.

Authors:  P A Ali; A F Al-Hussany; C A Bennett; D A Hancock; A M El-Sharkawi
Journal:  Phys Med Biol       Date:  1998-08       Impact factor: 3.609

Review 9.  Mechanisms of resistance of human tumours to anticancer drugs of the platinum family: a review.

Authors:  P J Ferguson
Journal:  J Otolaryngol       Date:  1995-08

10.  X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays.

Authors:  Seong-Kyun Cheong; Bernard L Jones; Arsalan K Siddiqi; Fang Liu; Nivedh Manohar; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2010-01-13       Impact factor: 3.609

View more
  11 in total

1.  Proton-induced x-ray fluorescence CT imaging.

Authors:  Magdalena Bazalova-Carter; Moiz Ahmad; Taeko Matsuura; Seishin Takao; Yuto Matsuo; Rebecca Fahrig; Hiroki Shirato; Kikuo Umegaki; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

2.  L-shell x-ray fluorescence computed tomography (XFCT) imaging of Cisplatin.

Authors:  Magdalena Bazalova; Moiz Ahmad; Guillem Pratx; Lei Xing
Journal:  Phys Med Biol       Date:  2013-12-13       Impact factor: 3.609

3.  Characterization of a Pixelated Cadmium Telluride Detector System Using a Polychromatic X-Ray Source and Gold Nanoparticle-Loaded Phantoms for Benchtop X-Ray Fluorescence Imaging.

Authors:  Sandun Jayarathna; Md Foiez Ahmed; Liam O'Ryan; Hem Moktan; Yonggang Cui; Sang Hyun Cho
Journal:  IEEE Access       Date:  2021-03-29       Impact factor: 3.367

4.  Feasibility study of Compton cameras for x-ray fluorescence computed tomography with humans.

Authors:  Don Vernekohl; Moiz Ahmad; Garry Chinn; Lei Xing
Journal:  Phys Med Biol       Date:  2016-11-15       Impact factor: 3.609

5.  Technical Note: A benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system with a high-power x-ray source and transmission CT imaging capability.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2018-09-14       Impact factor: 4.071

6.  Distinctive Energy Profile of Water-Soluble, Thiolate-Protected Gold Nanoparticles as Potential Molecular Marker for Vulnerable Plaque Detection with XFCT Imaging.

Authors:  Raiyan T Zaman; Don Vernekhol; Lei Xing
Journal:  J Radiol Radiat Ther       Date:  2020-10-12

7.  Experimental demonstration of direct L-shell x-ray fluorescence imaging of gold nanoparticles using a benchtop x-ray source.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

8.  Improving x-ray fluorescence signal for benchtop polychromatic cone-beam x-ray fluorescence computed tomography by incident x-ray spectrum optimization: a Monte Carlo study.

Authors:  Nivedh Manohar; Bernard L Jones; Sang Hyun Cho
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

9.  Joint Optimization of Collimator and Reconstruction Parameters in X-Ray Fluorescence Computed Tomography Using Analytical Point Spread Function and Model Observer.

Authors:  Hsin Wu Tseng; Srinivasan Vedantham; Sang Hyun Cho; Andrew Karellas
Journal:  IEEE Trans Biomed Eng       Date:  2019-12-30       Impact factor: 4.538

10.  Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography.

Authors:  Nivedh Manohar; Francisco J Reynoso; Parmeswaran Diagaradjane; Sunil Krishnan; Sang Hyun Cho
Journal:  Sci Rep       Date:  2016-02-25       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.