Investigation of transmission computed tomography (CT) image quality and x-ray dose achievable from an experimental dual-mode benchtop x-ray fluorescence CT and transmission CT system.

Luzhen Deng1,2, Selcuk Yasar2, Md Foiez Ahmed2, Sandun Jayarathna2, Peng Feng1, Biao Wei1, Srinivasan Vedantham3, Andrew Karellas3, Sang Hyun Cho2,4.   

Abstract

OBJECTIVE: To investigate the image quality and x-ray dose associated with a transmission computed tomography (CT) component implemented within the same platform of an experimental benchtop x-ray fluorescence CT (XFCT) system for multimodal preclinical imaging applications.
METHODS: Cone-beam CT scans were performed using an experimental benchtop CT + XFCT system and a cylindrically-shaped 3D-printed polymethyl methacrylate phantom (3 cm in diameter, 7 cm in height) loaded with various concentrations (0.05-1 wt. %) of gold nanoparticles (GNPs). Two commercial CT quality assurance phantoms containing 3D line-pair (LP) targets and contrast targets were also scanned. The x-ray beams of 40 and 62 kVp, both filtered by 0.08 mm Cu and 0.4 mm Al, were used with 17 ms of exposure time per projection at three current settings (2.5, 5, and 10 mA). The ordered-subset simultaneous algebraic reconstruction and total variation-minimization methods were used to reconstruct images. Sparse projection and short scan were considered to reduce the x-ray dose. The contrast-to-noise ratio (CNR) and modulation transfer function (MTF) were calculated.
RESULTS: The lowest detectable concentration of GNPs (CNR > 5) and the highest spatial resolution (per MTF50%) were 0.10 wt. % and 9.5 LP/CM, respectively, based on the images reconstructed from 360 projections of the 40 kVp beam (or x-ray dose of 3.44 cGy). The background noise for the image resulting in the lowest GNP detection limit was 25 Hounsfield units.
CONCLUSION: The transmission CT component within the current experimental benchtop CT + XFCT system produced images deemed acceptable for multimodal (CT + XFCT) imaging purposes, with less than 4 cGy of x-ray dose.

Entities:  

Keywords:  Computed tomography; image resolution; x-ray dose; x-ray fluorescence computed tomography

Year:  2019        PMID: 30909268      PMCID: PMC7027361          DOI: 10.3233/XST-180457

Source DB:  PubMed          Journal:  J Xray Sci Technol        ISSN: 0895-3996            Impact factor:   1.535


  21 in total

1.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  Distance-driven projection and backprojection in three dimensions.

Authors:  Bruno De Man; Samit Basu
Journal:  Phys Med Biol       Date:  2004-06-07       Impact factor: 3.609

3.  Finite detector based projection model for high spatial resolution.

Authors:  Hengyong Yu; Ge Wang
Journal:  J Xray Sci Technol       Date:  2012       Impact factor: 1.535

4.  Characterization of image quality and image-guidance performance of a preclinical microirradiator.

Authors:  R Clarkson; P E Lindsay; S Ansell; G Wilson; S Jelveh; R P Hill; D A Jaffray
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

5.  Analytic comparison between X-ray fluorescence CT and K-edge CT.

Authors:  Peng Feng; Wenxiang Cong; Biao Wei; Ge Wang
Journal:  IEEE Trans Biomed Eng       Date:  2014-03       Impact factor: 4.538

6.  Spatial resolution characterization of a X-ray microCT system.

Authors:  J Rueckel; M Stockmar; F Pfeiffer; J Herzen
Journal:  Appl Radiat Isot       Date:  2014-08-30       Impact factor: 1.513

7.  In vivo characterization of tumor vasculature using iodine and gold nanoparticles and dual energy micro-CT.

Authors:  Darin P Clark; Ketan Ghaghada; Everett J Moding; David G Kirsch; Cristian T Badea
Journal:  Phys Med Biol       Date:  2013-02-19       Impact factor: 3.609

8.  Gold nanoparticles: a new X-ray contrast agent.

Authors:  J F Hainfeld; D N Slatkin; T M Focella; H M Smilowitz
Journal:  Br J Radiol       Date:  2006-03       Impact factor: 3.039

9.  A Monte Carlo Model of a Benchtop X-Ray Fluorescence Computed Tomography System and Its Application to Validate a Deconvolution-Based X-Ray Fluorescence Signal Extraction Method.

Authors:  Md Foiez Ahmed; Selcuk Yasar; Sang Hyun Cho
Journal:  IEEE Trans Med Imaging       Date:  2018-05-15       Impact factor: 10.048

10.  Contrast-noise-ratio (CNR) analysis and optimisation of breast-specific gamma imaging (BSGI) acquisition protocols.

Authors:  Dennis Dickerscheid; Jules Lavalaye; Leo Romijn; Jan Habraken
Journal:  EJNMMI Res       Date:  2013-03-25       Impact factor: 3.138

View more
  1 in total

1.  Joint Optimization of Collimator and Reconstruction Parameters in X-Ray Fluorescence Computed Tomography Using Analytical Point Spread Function and Model Observer.

Authors:  Hsin Wu Tseng; Srinivasan Vedantham; Sang Hyun Cho; Andrew Karellas
Journal:  IEEE Trans Biomed Eng       Date:  2019-12-30       Impact factor: 4.538

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.