Quantitative X-ray fluorescence imaging of gold nanoparticles using joint L1 and total variation regularized reconstruction.

Junwei Shi1, Blaine Granger1, Keying Xu1, Yidong Yang1,2,3.   

Abstract

BACKGROUND: This work proposed a joint L1 and total variation (TV) regularized reconstruction method for X-ray fluorescence tomography (XFT), and investigated the performance of this method in quantitative imaging of gold nanoparticles (GNPs).
METHODS: We developed a dual-modality XFT/CT imaging system which consisted of a benchtop X-ray source, a translation/rotation stage, a silicon drift detector for X-ray fluorescence (XRF) detection, and a flat panel detector for transmission X-ray detection. A pencil-beam collimator was 3D printed with steel and employed in sample excitation. The sensitivity of the XFT imaging system was determined by imaging water phantoms with multiple inserts containing GNP solutions of various concentrations (0.02-0.16 wt.%). A joint L1 and total variation (TV) regularized algorithm was developed for XFT reconstruction, where the L1 regularization was used to reduce image artifacts and the TV regularization was used to preserve the shape of targets. Nonlinear conjugate gradient (NCG) descent algorithm with backtracking line search was adopted to solve the reconstruction problem. We compared the L1 + TV regularization method with filtered back projection (FBP), maximum likelihood expectation maximization (ML-EM), L1 regularization, and TV regularization methods. Contrast-to-noise ratio (CNR), Dice similarity coefficient (DSC) and localization error (LE) metrics were used to compare the performance of different methods. The CT and XFT imaging doses were also measured using EBT2 radiochromic films.
RESULTS: The 3D printed pencil-beam collimator shaped an excitation beam with a 2 mm full width at half maximum at the imaging isocenter. Based on the phantom imaging experiments, the joint L1 and TV regularization method performed better than FBP, ML-EM, L1 regularization and TV regularization methods, with higher localization accuracy (offset <0.6 mm), CNR and DSC values. Compared with CT, XFT with L1 + TV regularized reconstruction demonstrated higher sensitivity in GNP imaging, and could detect GNP at a concentration of 0.02 wt.% or lower. Moreover, there existed a significant linear correlation (R2>0.99) between the reconstructed and true GNP concentration. The estimated XFT imaging dose is about 41.22 cGy under current setting.
CONCLUSIONS: The joint L1 + TV regularized reconstruction algorithm performed better in noise suppression and shape preservation. Using the L1 + TV regularized reconstruction, the XFT system is able to localize GNP targets with submillimeter accuracy and quantify GNP distribution at a concentration of 0.02 wt.% or lower. 2020 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  L1 regularization; X-ray fluorescence (XRF); computed tomography; gold nanoparticles; total variation regularization

Year:  2020        PMID: 31956541      PMCID: PMC6960426          DOI: 10.21037/qims.2019.10.15

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  27 in total

1.  Anti-CD4-targeted gold nanoparticles induce specific contrast enhancement of peripheral lymph nodes in X-ray computed tomography of live mice.

Authors:  Wolfgang Eck; Anthony I Nicholson; Hanswalter Zentgraf; Wolfhard Semmler; Sönke Bartling
Journal:  Nano Lett       Date:  2010-07-14       Impact factor: 11.189

2.  Prior image constrained compressed sensing (PICCS): a method to accurately reconstruct dynamic CT images from highly undersampled projection data sets.

Authors:  Guang-Hong Chen; Jie Tang; Shuai Leng
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

3.  Optical molecular imaging-guided radiation therapy part 1: Integrated x-ray and bioluminescence tomography.

Authors:  Junwei Shi; Thirupandiyur S Udayakumar; Zhiqun Wang; Nesrin Dogan; Alan Pollack; Yidong Yang
Journal:  Med Phys       Date:  2017-07-20       Impact factor: 4.071

4.  L-shell x-ray fluorescence computed tomography (XFCT) imaging of Cisplatin.

Authors:  Magdalena Bazalova; Moiz Ahmad; Guillem Pratx; Lei Xing
Journal:  Phys Med Biol       Date:  2013-12-13       Impact factor: 3.609

5.  The use of gold nanoparticles to enhance radiotherapy in mice.

Authors:  James F Hainfeld; Daniel N Slatkin; Henry M Smilowitz
Journal:  Phys Med Biol       Date:  2004-09-21       Impact factor: 3.609

6.  Experimental demonstration of direct L-shell x-ray fluorescence imaging of gold nanoparticles using a benchtop x-ray source.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

7.  X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays.

Authors:  Seong-Kyun Cheong; Bernard L Jones; Arsalan K Siddiqi; Fang Liu; Nivedh Manohar; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2010-01-13       Impact factor: 3.609

8.  Gold nanoparticle imaging and radiotherapy of brain tumors in mice.

Authors:  James F Hainfeld; Henry M Smilowitz; Michael J O'Connor; Farrokh Avraham Dilmanian; Daniel N Slatkin
Journal:  Nanomedicine (Lond)       Date:  2012-12-24       Impact factor: 5.307

Review 9.  Nanocarriers as an emerging platform for cancer therapy.

Authors:  Dan Peer; Jeffrey M Karp; Seungpyo Hong; Omid C Farokhzad; Rimona Margalit; Robert Langer
Journal:  Nat Nanotechnol       Date:  2007-12       Impact factor: 39.213

10.  Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography.

Authors:  Nivedh Manohar; Francisco J Reynoso; Parmeswaran Diagaradjane; Sunil Krishnan; Sang Hyun Cho
Journal:  Sci Rep       Date:  2016-02-25       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.