Sensitivity enhancement of an experimental benchtop x-ray fluorescence imaging system by deploying a single crystal cadmium telluride detector system optimized for high flux x-ray operations.

Hem Moktan1, Sandun Jayarathna1, Sang Hyun Cho1,2.   

Abstract

In this work, an energy-resolving thermoelectrically cooled single crystal cadmium telluride (CdTe) detector system upgraded with the latest firmware was optimized for high x-ray flux operations using high bias voltage and fast peaking time. This detector system was deployed into an experimental benchtop x-ray fluorescence (XRF) imaging/computed tomography (XFCT) system developed for quantitative imaging of metal nanoprobes such as gold nanoparticles (GNPs). Using the firmware-upgraded and existing/old CdTe detector systems, the Compton/XRF spectra from small (8 mm diameter) GNP-containing phantoms were acquired. The phantoms were irradiated with 1.8 mm Sn-filtered 125 kVp cone beam x-rays at 24 mA. The firmware-upgraded detector system produced relatively lower dead time under high x-ray flux, compared with the old detector system, and performed well with the spectral resolution of ~0.7 keV (in full width at half maximum) at 69 keV photon energy. Given the same 2 mm aperture detector collimator and irradiation time of 10 s, this detector system managed to score nearly 50% more gold XRF signals than the existing one at all GNP concentrations tested. This improvement resulted in the GNP detection limit of 0.02 wt. % which was lower than that (0.03 wt. %) achievable with the existing detector system. When combined with the detector collimator containing a larger (3 mm) aperture, the firmware-upgraded detector system produced drastically more gold XRF signal at a given GNP concentration (e.g., 9 times more for 1 wt. % GNP solution and irradiation time of 10 s), leading to further reduction in the GNP detection limit (i.e., 0.01 wt. %). The present investigation showed that the firmware upgraded CdTe detector system optimized for high x-ray flux operations allowed for better photon counting efficiency, thus leading to sensitivity enhancement of an experimental benchtop XRF/XFCT imaging system.

Entities:  

Keywords:  benchtop x-ray fluorescence imaging; gold nanoparticles; single crystal cadmium telluride detector; x-ray fluorescence

Year:  2021        PMID: 35340930      PMCID: PMC8942383          DOI: 10.1016/j.nima.2021.166198

Source DB:  PubMed          Journal:  Nucl Instrum Methods Phys Res A        ISSN: 0168-9002            Impact factor:   1.455


  13 in total

1.  Estimation of strontium in animal bone using x-ray fluorescence analysis.

Authors:  W M ROBERTS
Journal:  Nature       Date:  1959-03-28       Impact factor: 49.962

2.  Experimental validation of L-shell x-ray fluorescence computed tomography imaging: phantom study.

Authors:  Magdalena Bazalova-Carter; Moiz Ahmad; Lei Xing; Rebecca Fahrig
Journal:  J Med Imaging (Bellingham)       Date:  2015-10-08

3.  Improved signal-to-noise ratio for non-perpendicular detection angles in x-ray fluorescence computed tomography (XFCT).

Authors:  Martin Sjölin; Mats Danielsson
Journal:  Phys Med Biol       Date:  2014-10-13       Impact factor: 3.609

Review 4.  Elemental and chemically specific X-ray fluorescence imaging of biological systems.

Authors:  M Jake Pushie; Ingrid J Pickering; Malgorzata Korbas; Mark J Hackett; Graham N George
Journal:  Chem Rev       Date:  2014-08-07       Impact factor: 60.622

5.  Three-dimensional x-ray fluorescence mapping of a gold nanoparticle-loaded phantom.

Authors:  Liqiang Ren; Di Wu; Yuhua Li; Ge Wang; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

6.  Technical Note: A benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system with a high-power x-ray source and transmission CT imaging capability.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2018-09-14       Impact factor: 4.071

7.  X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays.

Authors:  Seong-Kyun Cheong; Bernard L Jones; Arsalan K Siddiqi; Fang Liu; Nivedh Manohar; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2010-01-13       Impact factor: 3.609

8.  Experimental demonstration of benchtop x-ray fluorescence computed tomography (XFCT) of gold nanoparticle-loaded objects using lead- and tin-filtered polychromatic cone-beams.

Authors:  Bernard L Jones; Nivedh Manohar; Francisco Reynoso; Andrew Karellas; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2012-11-08       Impact factor: 3.609

9.  A Monte Carlo Model of a Benchtop X-Ray Fluorescence Computed Tomography System and Its Application to Validate a Deconvolution-Based X-Ray Fluorescence Signal Extraction Method.

Authors:  Md Foiez Ahmed; Selcuk Yasar; Sang Hyun Cho
Journal:  IEEE Trans Med Imaging       Date:  2018-05-15       Impact factor: 10.048

10.  Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography.

Authors:  Nivedh Manohar; Francisco J Reynoso; Parmeswaran Diagaradjane; Sunil Krishnan; Sang Hyun Cho
Journal:  Sci Rep       Date:  2016-02-25       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.