| Literature DB >> 30109245 |
Jackrit Suthakorn1, Narucha Tanaiutchawoot1, Cholatip Wiratkapan2, Songpol Ongwattanakul1.
Abstract
OBJECTIVE: This paper proposes the development of a breast biopsy navigation system with an assisted needle holder tool for a coaxial needle and a graphical user interface, which utilizes an optical tracking device to localize the needle position relative to the ultrasound image with the aim to improve performance especially for a new radiologist or an inexperienced group.Entities:
Keywords: Assisted needle holder tool; Breast biopsy navigation system; Graphical user interface (GUI); Optical tracking device; Ultrasound image
Year: 2018 PMID: 30109245 PMCID: PMC6090089 DOI: 10.1016/j.ejro.2018.07.001
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Fig. 1(a) minimum and (b) maximum of workspace of assisted needle holder design.
Fig. 2Annual snap fit design for revolute joints.
Fig. 3(a) the components and (b) the prototype of the assisted needle holder.
Fig. 4Guidance image for needle insertion with the trajectory path (1), needle orientation (2) and simulated needle (3).
Fig. 5Graphical user interface for breast biopsy navigation system.
Fig. 6The experiment setup.
Fig. 7The result comparison of (a) time and (b) success rate to finish the procedure by the experience and inexperience intervention.
Biopsy procedure time and success rate for experience and inexperience groups using freehand and assisted needle holder method with statistical analysis.
| Time (s) | Wilcoxon's Z statistic 2 tails | Success Rate (%) | One-proportion Z-test | |||
|---|---|---|---|---|---|---|
| Average | SD | |||||
| Experience | Freehand | 6.022 | 4.191 | Not significant | 95.56 | Not significant |
| Assisted Needle Holder | 6.044 | 4.462 | 100 | |||
| Inexperience | Freehand | 12.925 | 14.880 | Significant | 57.5 | Significant |
| Assisted Needle Holder | 6.725 | 5.996 | 97.5 | |||
Biopsy procedure time and success rate for experience and inexperience groups using the assisted needle holder and the system (Assisted Needle Holder and GUI) method with statistical analysis.
| Time (s) | Wilcoxon's Z statistic 2 tails | Success Rate (%) | One-proportion Z-test | |||
|---|---|---|---|---|---|---|
| Average | SD | |||||
| Experience | Assisted Needle Holder | 6.044 | 4.462 | Significant | 100 | Not significant |
| Assisted Needle Holder + GUI | 10.533 | 6.416 | 97.78 | |||
| Inexperience | Assisted Needle Holder | 6.725 | 5.996 | Significant | 97.5 | Not significant |
| Assisted Needle Holder + GUI | 11.425 | 6.917 | 92.5 | |||
Biopsy procedure time and success rate for experience and inexperience groups using freehand method and the breast biopsy navigation system method with statistical analysis.
| Time (s) | Wilcoxon's Z statistic 2 tails | Success Rate (%) | One-proportion Z-test | |||
|---|---|---|---|---|---|---|
| Average | SD | |||||
| Experience | Freehand | 6.022 | 4.191 | Significant | 95.56 | Not significant |
| Assisted Needle Holder + GUI | 10.533 | 6.416 | 97.78 | |||
| Inexperience | Freehand | 12.925 | 14.880 | Significant | 57.5 | Significant |
| Assisted Needle Holder + GUI | 11.425 | 6.917 | 92.5 | |||