Effect of loading frequency on cyclic fatigue lifetime of a standard-diameter implant with an internal abutment connection.

Yuanyuan Duan1, Jason A Griggs2.   

Abstract

OBJECTIVE: To investigate the effect of loading frequency on the fatigue lifetime of one standard-diameter titanium dental implant system.
METHODS: Thirty-six titanium dental implant specimens (Bone Level RC, Straumann) were assembled following manufacturer's instructions and torqued into cylindrical holder blocks following the apparatus specified by the ISO 14801 test standard. Stainless steel loading hemispheres were bonded on the abutments with a moment arm of 11mm. The holder blocks had layers of differing stiffness to simulate human jaw bone. Constant-stress fatigue lifetime testing was conducted at two frequencies (2Hz and 15Hz) with a stress ratio of 0.1 until fracture in deionized water at 37°C on servo-hydraulic load frames (MTS). The fractured specimens were retrieved and examined using fractographic technique to determine the failure mode. The lifetime data were fit to a general log-linear regression model.
RESULTS: The coefficient for the load amplitude term of the regression model indicated that increasing load amplitude had a statistically significant negative effect on the fatigue lifetime. The coefficients for the cyclic frequency term and the load-frequency interaction term were not significantly different from zero, which indicated that increasing loading frequency did not have an influence on the number of cycles to failure. Fractographic analysis showed that all specimens exhibited an identical combined fracture of abutment and abutment screw adjacent to the bone level. SIGNIFICANCE: Higher loading frequency at least up to 15Hz may be used for future studies of some implant systems to improve the efficiency of fatigue testing.
Copyright © 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Accelerated lifetime testing; CP titanium; Cyclic fatigue; Dental implants; Fractography; General log-linear model; Inverse power law; Strain rate sensitivity; Weibull distribution

Mesh:

Substances:

Year:  2018        PMID: 30220505      PMCID: PMC7249732          DOI: 10.1016/j.dental.2018.09.001

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  17 in total

1.  Retrieval analyses of a blade implant after 231 months of clinical function.

Authors:  L I Linkow; K Donath; J E Lemons
Journal:  Implant Dent       Date:  1992       Impact factor: 2.454

2.  Light and scanning electron microscopic report of four fractured implants.

Authors:  A Piattelli; M Piattelli; A Scarano; L Montesani
Journal:  Int J Oral Maxillofac Implants       Date:  1998 Jul-Aug       Impact factor: 2.804

3.  The effect of oral-like environment on dental implants' fatigue performance.

Authors:  Keren Shemtov-Yona; Daniel Rittel; Liran Levin; Eli E Machtei
Journal:  Clin Oral Implants Res       Date:  2012-12-20       Impact factor: 5.977

Review 4.  Improvements in implant dentistry over the last decade: comparison of survival and complication rates in older and newer publications.

Authors:  Bjarni E Pjetursson; Asgeir G Asgeirsson; Marcel Zwahlen; Irena Sailer
Journal:  Int J Oral Maxillofac Implants       Date:  2014       Impact factor: 2.804

Review 5.  Dental Implants.

Authors:  Jason A Griggs
Journal:  Dent Clin North Am       Date:  2017-10

6.  Young's modulus of trabecular and cortical bone material: ultrasonic and microtensile measurements.

Authors:  J Y Rho; R B Ashman; C H Turner
Journal:  J Biomech       Date:  1993-02       Impact factor: 2.712

Review 7.  Fracture of dental implants: literature review and report of a case.

Authors:  Nirit Tagger Green; Eli E Machtei; Jacob Horwitz; Micha Peled
Journal:  Implant Dent       Date:  2002       Impact factor: 2.454

8.  Evaluation of test protocol variables for dental implant fatigue research.

Authors:  Cornell K Lee; Matthias Karl; J Robert Kelly
Journal:  Dent Mater       Date:  2009-07-30       Impact factor: 5.304

9.  Influence of loading frequency on implant failure under cyclic fatigue conditions.

Authors:  Matthias Karl; J Robert Kelly
Journal:  Dent Mater       Date:  2009-07-29       Impact factor: 5.304

10.  Complication and failure rates with implant-supported fixed dental prostheses and single crowns: a 10-year retrospective study.

Authors:  Julia-Gabriela Wittneben; Daniel Buser; Giovanni E Salvi; Walter Bürgin; Stefan Hicklin; Urs Brägger
Journal:  Clin Implant Dent Relat Res       Date:  2013-04-02       Impact factor: 3.932

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.