Long-term Results of Primary Hip Arthroplasty with Cup Inclination Angle Bigger than Fifty Degrees.

Jialiang Tian1, Li Sun1, Ruiyin Hu1, Wei Han1, Xiaobin Tian1,2.   

Abstract

BACKGROUND: Most of studies suggest the cup inclination angle is controversial. All of the studies previous have not control the prosthesis type and head diameter, it is difficult to determine relative or synergistic effects of cup angle.
METHODS: We retrospectively reviewed 54 patients (61 hips) with primary total hip arthroplasties which cup inclination angle bigger than 500 after a mean 11.8 years follow-up. All the prosthesis are noncemented cup with a 28 mm metal head(Trilogy Acetabular Shell, Zimmer). The cup inclination was divided into three groups, 50-550 in 26 hips, 55-600 in 21 hips, and bigger than 600 in 14 hips. An immediate postoperative radiograph was compared with a follow-up radiograph. The polyethylene wear rates and abduction of the acetabular cups was measured in all of the patients.
RESULTS: The preoperative mean Harris hip score improved from 47.36 to 94.3 points at 10 years. The survivorship of the cup was 100% at 10 years. The mean rate of liner wear was 0.144 ± 0.031 mm/y (0.105-0.178 mm/y) in cup inclination angle between 500-550, and 0.260 ± 0.043 mm/y (0.215-0.394 mm/y) in angle between 550-600, 0. 403 ± 0.016 mm/y (0.378-0.423 mm/y) in angle bigger than 600. The different cup inclination groups are different with liner and volumetric wear.
CONCLUSION: For the metal-on-polyethylene prostheses, the liner wear obvious correlate with cup inclination after angle bigger than 500. It can be concluded that the ideal abduction angle for metal-on-polyethylene prostheses should be less than 55° in hip total replacement.

Entities:  

Keywords:  Cementless; Cup inclination; Hip arthroplasty; Linear wear; Metal-on-polyethylene

Year:  2017        PMID: 29896015      PMCID: PMC5995005          DOI: 10.1016/j.jcot.2017.03.007

Source DB:  PubMed          Journal:  J Clin Orthop Trauma        ISSN: 0976-5662


  21 in total

1.  Optimizing acetabular component position to minimize impingement and reduce contact stress.

Authors:  D D D'Lima; P C Chen; C W Colwell
Journal:  J Bone Joint Surg Am       Date:  2001       Impact factor: 5.284

Review 2.  Dislocation after total hip arthroplasty: implant design and orientation.

Authors:  Robert L Barrack
Journal:  J Am Acad Orthop Surg       Date:  2003 Mar-Apr       Impact factor: 3.020

3.  Mid-term wear characteristics of an uncemented acetabular component.

Authors:  J H M Goosen; C C P M Verheyen; N J A Tulp
Journal:  J Bone Joint Surg Br       Date:  2005-11

4.  Determination of polyethylene wear in total hip replacements with use of digital radiographs.

Authors:  J M Martell; S Berdia
Journal:  J Bone Joint Surg Am       Date:  1997-11       Impact factor: 5.284

5.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

6.  Dislocations after total hip-replacement arthroplasties.

Authors:  G E Lewinnek; J L Lewis; R Tarr; C L Compere; J R Zimmerman
Journal:  J Bone Joint Surg Am       Date:  1978-03       Impact factor: 5.284

7.  Ectopic ossification following total hip replacement. Incidence and a method of classification.

Authors:  A F Brooker; J W Bowerman; R A Robinson; L H Riley
Journal:  J Bone Joint Surg Am       Date:  1973-12       Impact factor: 5.284

8.  Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis.

Authors:  B M Jolles; P Zangger; P-F Leyvraz
Journal:  J Arthroplasty       Date:  2002-04       Impact factor: 4.757

9.  Polyethylene wear and acetabular component orientation.

Authors:  Shantanu Patil; Arnie Bergula; Peter C Chen; Clifford W Colwell; Darryl D D'Lima
Journal:  J Bone Joint Surg Am       Date:  2003       Impact factor: 5.284

10.  Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up.

Authors:  Pacharapol Udomkiat; Lawrence D Dorr; Zhinian Wan
Journal:  J Bone Joint Surg Am       Date:  2002-07       Impact factor: 5.284

View more
  1 in total

1.  The Impingement-free, Prosthesis-specific, and Anatomy-adjusted Combined Target Zone for Component Positioning in THA Depends on Design and Implantation Parameters of both Components.

Authors:  Karl-Heinz Widmer
Journal:  Clin Orthop Relat Res       Date:  2020-08       Impact factor: 4.755

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.