Effect of automatic tube current modulation on radiation dose and image quality for low tube voltage multidetector row CT angiography: phantom study.

Sebastian T Schindera1, Rendon C Nelson, Terry Yoshizumi, Greta Toncheva, Giao Nguyen, David M DeLong, Zsolt Szucs-Farkas.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate the effect of automatic tube current modulation on radiation dose and image quality for low tube voltage computed tomography (CT) angiography.
MATERIALS AND METHODS: An anthropomorphic phantom was scanned with a 64-section CT scanner using following tube voltages: 140 kVp (Protocol A), 120 kVp (Protocol B), 100 kVp (Protocol C), and 80 kVp (Protocol D). To achieve similar noise, combined z-axis and xy-axes automatic tube current modulation was applied. Effective dose (ED) for the four tube voltages was assessed. Three plastic vials filled with different concentrations of iodinated solution were placed on the phantom's abdomen to obtain attenuation measurements. The signal-to-noise ratio (SNR) was calculated and a figure of merit (FOM) for each iodinated solution was computed as SNR(2)/ED.
RESULTS: The ED was kept similar for the four different tube voltages: (A) 5.4 mSv +/- 0.3, (B) 4.1 mSv +/- 0.6, (C) 3.9 mSv +/- 0.5, and (D) 4.2 mSv +/- 0.3 (P > .05). As the tube voltage decreased from 140 to 80 kVp, image noise was maintained (range, 13.8-14.9 HU) (P > .05). SNR increased as the tube voltage decreased, with an overall gain of 119% for the 80-kVp compared to the 140-kVp protocol (P < .05). The FOM results indicated that with a reduction of the tube voltage from 140 to 120, 100, and 80 kVp, at constant SNR, ED was reduced by a factor of 2.1, 3.3, and 5.1, respectively, (P < .001).
CONCLUSIONS: As tube voltage decreases, automatic tube current modulation for CT angiography yields either a significant increase in image quality at constant radiation dose or a significant decrease in radiation dose at a constant image quality.

Mesh:

Year:  2009        PMID: 19409820     DOI: 10.1016/j.acra.2009.02.021

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  17 in total

1.  Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults.

Authors:  Gaurav S Desai; Raul N Uppot; Elaine W Yu; Avinash R Kambadakone; Dushyant V Sahani
Journal:  Eur Radiol       Date:  2012-04-19       Impact factor: 5.315

Review 2.  [Strategies for reducing the CT radiation dose].

Authors:  S T Schindera; C Nauer; R Treier; P Trueb; G von Allmen; P Vock; Z Szucs-Farkas
Journal:  Radiologe       Date:  2010-12       Impact factor: 0.635

Review 3.  State of the art: technologies for computed tomography dose reduction.

Authors:  Martin L D Gunn; Jennifer R Kohr
Journal:  Emerg Radiol       Date:  2009-11-20

4.  Optimizing radiation exposure in screening of body packing: image quality and diagnostic acceptability of an 80 kVp protocol with automated tube current modulation.

Authors:  Joel Aissa; Johannes Boos; Christian Rubbert; Julian Caspers; Christoph Schleich; Christoph Thomas; Patric Kröpil; Gerald Antoch; Falk Miese
Journal:  Forensic Sci Med Pathol       Date:  2017-03-07       Impact factor: 2.007

Review 5.  Raise the bar and lower the dose: current and future strategies for radiation dose reduction in head and neck imaging.

Authors:  M Ibrahim; H Parmar; E Christodoulou; S Mukherji
Journal:  AJNR Am J Neuroradiol       Date:  2013-02-28       Impact factor: 3.825

Review 6.  The Diagnosis and Treatment of Hemoptysis.

Authors:  Harald Ittrich; Maximilian Bockhorn; Hans Klose; Marcel Simon
Journal:  Dtsch Arztebl Int       Date:  2017-06-05       Impact factor: 5.594

7.  Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison.

Authors:  Aleksander W Krazinski; Felix G Meinel; U Joseph Schoepf; Justin R Silverman; Christian Canstein; Carlo N De Cecco; Lucas L Geyer
Journal:  Eur Radiol       Date:  2014-07-24       Impact factor: 5.315

8.  Hybrid iterative reconstruction technique for liver CT scans for image noise reduction and image quality improvement: evaluation of the optimal iterative reconstruction strengths.

Authors:  Ji Soo Song; Jeong Min Lee; Ji Young Sohn; Jeong-Hee Yoon; Joon Koo Han; Byung Ihn Choi
Journal:  Radiol Med       Date:  2014-08-29       Impact factor: 3.469

9.  80-kVp CT using Iterative Reconstruction in Image Space algorithm for the detection of hypervascular hepatocellular carcinoma: phantom and initial clinical experience.

Authors:  Saebeom Hur; Jeong Min Lee; Soo Jin Kim; Ji Hoon Park; Joon Koo Han; Byung Ihn Choi
Journal:  Korean J Radiol       Date:  2012-03-07       Impact factor: 3.500

10.  Cardiac CT angiography--radiation dose-how effective are we in reducing radiation dose from cardiac CT angiography?

Authors:  Quynh A Truong; Emily Siegel; Christopher P Cannon
Journal:  Rev Cardiovasc Med       Date:  2009       Impact factor: 2.930

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.