BACKGROUND: Femoral suspensory fixation for anterior cruciate ligament (ACL) reconstruction has evolved from fixed- to adjustable-loop devices. However, there are still controversies regarding undesired lengthening of adjustable-loop devices. HYPOTHESIS: Adjustable-loop fixation will achieve similar elongation to that of fixed-loop devices, and intraoperative preconditioning will reduce initial elongation for adjustable-loop constructs. STUDY DESIGN: Controlled laboratory study. METHODS: Three adjustable-loop devices (GraftMax, TightRope, and Ultrabutton) and 2 fixed-loop devices (Endobutton and RetroButton) were used in an intraoperative surgical technique workflow according to an in vitro model with porcine bone and bovine tendons (8 specimens per device; N = 40 constructs tested). Each construct underwent 1000 cycles of position- and force-controlled dynamic loading, whereby a total elongation threshold of 3 mm was defined as clinical failure. Constructs were finally pulled to failure at 50 mm/min. RESULTS: There were no statistically significant differences among the devices for total or dynamic elongation. Total elongation (mean ± SD) for adjustable-loop constructs was 4.13 ± 1.46 mm for GraftMax, 2.78 ± 0.85 mm for TightRope, and 2.76 ± 0.45 mm for Ultrabutton; for the fixed-loop devices, total elongation was 2.85 ± 0.74 mm for Endobutton and 2.85 ± 1.03 mm for RetroButton. The GraftMax had a significantly lower initial force (95.5 ± 58.0 N) after retensioning, with the highest initial elongation (0.99 ± 0.60 mm). The Ultrabutton showed the greatest force loss (-105.9 ± 13.5 N) during position control cycling, which was significantly different from the GraftMax (-22.3 ± 28.2 N), with the smallest force loss (P < .001). The TightRope construct had a significantly smaller initial elongation (-0.36 ± 0.22 mm) and the greatest pull-to-failure load (958 ± 40 N) as compared with all of the other devices. CONCLUSION: Adjustable- and fixed-loop configurations achieved statistically comparable fixation strength for total elongation. However, the GraftMax construct exceeded the total elongation threshold of clinical failure. The Ultrabutton produced the greatest loss of force during position control cycling, and the GraftMax button design prevented proper retensioning. The TightRope had a significant greater ultimate strength when compared with all other devices. CLINICAL RELEVANCE: Biomechanical testing according to a surgical technique workflow suggests that adjustable-loop devices can be considered a safe alternative to fixed-loop devices in ACL reconstruction.
BACKGROUND: Femoral suspensory fixation for anterior cruciate ligament (ACL) reconstruction has evolved from fixed- to adjustable-loop devices. However, there are still controversies regarding undesired lengthening of adjustable-loop devices. HYPOTHESIS: Adjustable-loop fixation will achieve similar elongation to that of fixed-loop devices, and intraoperative preconditioning will reduce initial elongation for adjustable-loop constructs. STUDY DESIGN: Controlled laboratory study. METHODS: Three adjustable-loop devices (GraftMax, TightRope, and Ultrabutton) and 2 fixed-loop devices (Endobutton and RetroButton) were used in an intraoperative surgical technique workflow according to an in vitro model with porcine bone and bovine tendons (8 specimens per device; N = 40 constructs tested). Each construct underwent 1000 cycles of position- and force-controlled dynamic loading, whereby a total elongation threshold of 3 mm was defined as clinical failure. Constructs were finally pulled to failure at 50 mm/min. RESULTS: There were no statistically significant differences among the devices for total or dynamic elongation. Total elongation (mean ± SD) for adjustable-loop constructs was 4.13 ± 1.46 mm for GraftMax, 2.78 ± 0.85 mm for TightRope, and 2.76 ± 0.45 mm for Ultrabutton; for the fixed-loop devices, total elongation was 2.85 ± 0.74 mm for Endobutton and 2.85 ± 1.03 mm for RetroButton. The GraftMax had a significantly lower initial force (95.5 ± 58.0 N) after retensioning, with the highest initial elongation (0.99 ± 0.60 mm). The Ultrabutton showed the greatest force loss (-105.9 ± 13.5 N) during position control cycling, which was significantly different from the GraftMax (-22.3 ± 28.2 N), with the smallest force loss (P < .001). The TightRope construct had a significantly smaller initial elongation (-0.36 ± 0.22 mm) and the greatest pull-to-failure load (958 ± 40 N) as compared with all of the other devices. CONCLUSION: Adjustable- and fixed-loop configurations achieved statistically comparable fixation strength for total elongation. However, the GraftMax construct exceeded the total elongation threshold of clinical failure. The Ultrabutton produced the greatest loss of force during position control cycling, and the GraftMax button design prevented proper retensioning. The TightRope had a significant greater ultimate strength when compared with all other devices. CLINICAL RELEVANCE: Biomechanical testing according to a surgical technique workflow suggests that adjustable-loop devices can be considered a safe alternative to fixed-loop devices in ACL reconstruction.
Anterior cruciate ligament (ACL) ruptures are one of the most common injuries. There are
more than 150,000 ACL reconstructions (ACLRs) performed every year.[19,23] Suspensory femoral cortical fixation for ACLR has evolved from a fixed-loop
device (FLD) to an adjustable-loop device (ALD). Advantages of FLDs include strong
fixation that limits initial elongation while the graft is being incorporated. However,
FLDs require precise mathematical calculations to ensure proper bone tunnel measurements
for graft insertion. Some advantages of ALDs include ease of insertion without
intraoperative calculations,[8] a single loop size for all patients,[9] the possibility of a longer amount of tendon within the femoral tunnel that may
optimize graft incorporation, and the capability of retensioning the graft after initial fixation.[8,22] Despite these advantages, a clinical ACLR with a femoral FLD and tibial screw
fixation remains the benchmark for biomechanical performance evaluation for
configurations utilizing adjustable-loop fixation.There is controversy surrounding the increased elongations observed with ALDs as compared
with FLDs.[8,15,18,30] Some researchers have questioned the testing methods as an explanation for the
large elongation values observed with ALDs.[14] Conversely, ALDs have been shown to possess similar elongation values to FLDs[27] and have been implanted in surgery with successful clinical outcomes.[9] Specifically, the influence of retensioning ALDs during testing is often
discussed. Noonan et al[29] demonstrated that retensioning ALDs minimized the total cyclic displacement
values, while Johnson et al[18] found that retensioning did not have a statistically significant difference on
elongation for ALDs. However, both these studies were single-device testing as opposed
to full-construct testing (with bone and soft tissue) at zero load.There have been newly released ALDs, but reported testing of these devices was not
available in the literature at the time of writing.The purpose of this study was to comparatively test previously and newly released ALDs in
a full-construct surgical technique–based manner for stability evaluation in relation to
the benchmark FLD configuration. The FLD benchmark metrics were defined as force
maintenance throughout position-controlled cycling; initial, dynamic, and total
elongation throughout force-controlled cycling; and ultimate load and stiffness during
pull to failure. The hypotheses were that ACLR with ALD fixation would behave comparably
in terms of biomechanics with FLD and that intraoperative preconditioning with graft
precycling and retensioning would significantly reduce initial elongation for ALD
constructs.
Methods
Five femoral cortical suspension devices were selected to be biomechanically tested
in an in vitro model with porcine bone and bovine tendons (8 specimens per device; N
= 40 constructs tested). The 3 ALDs included were the GraftMax (GM; ConMed
Linvatec), TightRope (TR; Arthrex), and Ultrabutton (UB; Smith & Nephew). The 2
FLDs, which served as the positive control, were the Endobutton (EB; Smith &
Nephew) and RetroButton (RB; Arthrex), and both consisted of a 20-mm continuous loop
(Figure 1).
Figure 1.
The 5 suspensory femoral cortical devices tested from left to right:
GraftMax, TightRope, Ultrabutton, Endobutton (fixed-loop device), and
RetroButton (fixed-loop device). The adjustable-loop devices can modify loop
length by pulling on the shortening sutures, and the fixed-loop devices have
a loop length of 20 mm.
The 5 suspensory femoral cortical devices tested from left to right:
GraftMax, TightRope, Ultrabutton, Endobutton (fixed-loop device), and
RetroButton (fixed-loop device). The adjustable-loop devices can modify loop
length by pulling on the shortening sutures, and the fixed-loop devices have
a loop length of 20 mm.
Specimen Preparation
Full-construct testing was set up with the porcine bone and bovine tendons. Fresh
bovine tendons (2 years of age) were obtained from the local slaughterhouse.
Extensor digitorum tendons were harvested in our laboratory from the hind legs,
which have been shown to possess similar viscoelastic properties to human
hamstring tendons.[17] The tendons were stored at –20°C and thawed several hours before testing.
Porcine bone (6 months of age) was chosen owing to its previous use in ACL studies[2,30,35] and similarity to human bone.[1,16,28] The bones were stored at –20°C and thawed overnight at room temperature
before testing.The porcine tibias and femurs were initially prepared by removing all the soft
tissue from the bone. The tibias were embedded with RenCast mixture (Huntsman
Advanced Materials) in a custom rectangular steel fixture about 2 cm distal to
the tibial tunnel on the medial side. The lateral plateau of the tibia was sawed
off to create a constant 40-mm bone tunnel.The porcine femurs were prepared by measuring 35 mm from the lateral condyle of
the femur and sawing off the medial condyle at that point. A 35-mm cylinder
drill was used to saw the lateral condyle of the femur to create a bony block.
The cylindrical bony block was docked in a custom-made steel fixture.
Graft Preparation
The bovine tendon grafts for the GM, UB, and 2 FLDs were measured to 325 mm and
trimmed along the fiber orientation to a diameter of 9 mm when the graft was
quadrupled. A graft size of 9 mm was chosen because it represents a standard
graft diameter for human ACLR and smaller grafts (<8 mm) were shown to have
increased failure after 2 years.[24] The last 20 mm of the ends of the tendon were whipstitched with No. 2
suture to create a doubled graft,[7] which was then folded over the loop to create a quadruple-stranded
graft.Biocomposite interference screws were chosen for tibial fixation of the GM and
FLDs according to the corresponding surgical technique guides.[6,12,36,38] To add consistency to the study, a biocomposite interference screw was
also chosen for the UB since the surgical technique guide stated that tibial
fixation was determined by the surgeon’s preference.[36] The GraftLink all-inside technique for the TR, which consisted of an ALD
and button for tibial fixation, was chosen because it was the method promoted by
the manufacturer[5,27] (Figure 2).
Figure 2.
(A) Test setup with femoral bony block in the steel fixture, embedded
porcine tibia, and bovine tendon graft. The tibia and femur are 30 mm
apart. The graft is aligned to create worst-case scenario testing. (B)
Graft setup for all the devices. The arrow indicates the direction of
retensioning for the adjustable-loop devices.
(A) Test setup with femoral bony block in the steel fixture, embedded
porcine tibia, and bovine tendon graft. The tibia and femur are 30 mm
apart. The graft is aligned to create worst-case scenario testing. (B)
Graft setup for all the devices. The arrow indicates the direction of
retensioning for the adjustable-loop devices.The bovine tendon grafts for the TR with the GraftLink technique were measured to
280 mm, and the diameter was sized to 9 mm when quadrupled. The graft was
prepared by threading each end through an adjustable TR loop. Last, the 2 free
ends each went through the adjustable loop again, which created a
quadruple-stranded loop. This loop was then U-stitched with a No. 0 suture with
4 to 5 stitches and ends overlapping for about 5 mm to create a closed
continuous loop.[5,35] The femoral end of the graft included a full TR construct with the button
incorporated around the loop, while the tibial end of the graft included only
the TR loop with sutures. A button was subsequently attached during graft
insertion. Next, the 4 tendon strands on each end of the construct ends were
link stitched with a No. 2 suture incorporating each tendon strand. There were 2
link stitches on the tibial end and 1 link on the femoral side, as altered from
the surgical technique guide describing 2 link stitches on both the tibial and
femoral sides. The final graft preparation for the TR included burying the knots
from the sutures and having the sutures on the inside of the graft.[35]All grafts were pretensioned for conditioning reasons in a graft preparation
board and a 9-mm compression tube with 80 N for 5 minutes prior to insertion.[21,26] The grafts were kept moist during testing with a physiologic saline
solution.
Tunnel Preparation
The guide wire in the tibia represented the native ACL footprint and was used to
drill the bone tunnels in the tibias. The 40-mm full bone tunnel for the GM, UB,
and FLDs were created with a 9-mm cannulated drill. To create a bone socket, the
porcine tibias for the TR were prepared by using a 9-mm cannulated drill for 30
mm and then a 3.5-mm cannulated drill for the remaining 10 mm toward the cortex,
which differed from the surgical technique guide describing the usage of a flip
cutter drill.The femurs for the ALDs were initially prepared with a 9-mm cannulated drill for
a distance of 20 mm. Then, for the last 15 mm, a bone bridge was drilled until
the femoral cortex was reached with a 5-mm, 3.5-mm, and 4.5-mm cannulated drill
for the GM, TR, and UB, respectively. The femurs for the FLDs were prepared with
a 9-mm cannulated drill for 23 mm into the bone. For the remaining last 12 mm, a
4.5-mm hole for the EB and a 3.5-mm hole for the RB were drilled.
Device Insertion Techniques
The steel fixture with the femoral bony block was clamped into the tensile
testing machine (Instron ElectroPuls 10000). The tibia was inserted into the
tensile testing machine at an angle such that the graft was in line with the
femoral insertion, creating “worst-case scenario” testing owing to aligned load
axes, and was clamped tightly (Figure 2). A consistent joint space of 29 mm was measured between
the femur and tibia for ACLR primary fixation, which corresponded to a knee in
simulated 30° of flexion and served as a reference for elongation analysis. The
graft was inserted into the femur and tibia with a passing suture.
Countertension was applied on the graft as it was inserted into the femur. For
constructs with the GM or UB, a space of 2 to 3 mm was left in the femoral
socket, allowing retensioning.For the GM, UB, and the FLDs, a 50-N weight was suspended on the graft while a 9
× 28–mm biocomposite interference screw was inserted until it reached the
superior portion of the tibial bone. For the TR tests, the femoral-sided graft
was fully inserted and tensioned until graft tunnel docking, because
retensioning was performed on the tibial side. Next, a button (Adjustable Button
System; Arthrex) was applied to the tibial-sided TR suture loop and manually
tensioned to 50 N by pulling on the shortening strands. During pulling, the
applied force was monitored on the tensile testing machine.
Intraoperative Preconditioning
The testing protocol (Figure
3) included an “ACL length over flexion angle” relationship to
simulate knee flexion activity. In vivo kinematic data have shown that the ACL
experiences a consistent length decrease from 30 mm to 27 mm when the knee is
moved between full extension and 90° of flexion.[20] In reference to the primary fixation position with a graft length of 29
mm, which corresponds to 30° of knee flexion, a knee in full extension and 90°
of angled flexion is simulated by a graft length increase of 1 mm (30-mm joint
space) and a decrease of 2 mm (27-mm joint space), respectively. Therefore, the
testing protocol (Figure
3) for the ALDs started with 10 precycles in position control between
+1 mm and –2 mm at 0.5 Hz to simulate knee movement between full extension and
90° of flexion at the knee.
Figure 3.
Test protocol for the devices. Measurements included initial elongation
(Δad), dynamic elongation (Δde), total
elongation (Δae), force decrease (Δbc), and
stiffness during pull to failure (Δgh). The fixed-loop device
did not undergo preconditioning and retensioning; the TightRope was the
only device knotted on the tibial side.
Test protocol for the devices. Measurements included initial elongation
(Δad), dynamic elongation (Δde), total
elongation (Δae), force decrease (Δbc), and
stiffness during pull to failure (Δgh). The fixed-loop device
did not undergo preconditioning and retensioning; the TightRope was the
only device knotted on the tibial side.An advantage of ALDs is the retensioning, which implies the ability to increase
the graft force after primary fixation.[22] All surgical technique guides recommended retensioning; however, there
was no defined force for the device to be retensioned. It was shown that the FLD
graft force was about 130 N after the screw was inserted and the space between
the tibia and femur was increased 1 mm, representing full extension. Therefore,
the force for retensioning of the ALDs after precycling was chosen to be 200 N,
which was reproducible. If reaching 200 N was not achievable with the device
being tested, the maximal tensioning possible was applied.Retensioning was performed in full extension (30-mm joint space) for the TR and
UB, whereas the GM was tensioned in simulated 30° of knee flexion (29-mm joint
space) to enable the button-locking mechanism. For the TR, the graft was
retensioned on the tibial side to 200 N and then knotted with a surgeon’s knot
and 3 half-hitch knots with an arthroscopic knot pusher. The GM and UB were
retensioned at the femoral side to 200 N or the maximal force possible. All
femoral ALDs were kept knotless, which represents clinical practice by reducing
possible postoperative knot irritation or knot tying through soft tissue layers.
The test protocol for the FLDs did not include precycling and retensioning, as
it was not technically possible to compensate tension loss owing to screw
insertion.
Position Control Loading
After the preconditioning protocol, the position control loading block started
and consisted of 1000 cycles at 0.75 Hz between +1 mm (30-mm joint space) and –2
mm (27-mm joint space). The included slack on the graft enabled a complete
unloading (0 N) to occur, which may have constituted an unfavorable loading
situation for the ALDs. Position-controlled cycling simulated the in vivo
kinematics of the ACL during weightbearing knee flexion, as also established by
Monaco et al.[26] The initial peak and final force loss were measured within the position
control block.
Force Control Loading
After position control loading, there were 1000 cycles of load control between 10
N and 250 N at 0.75 Hz. The 250-N load was chosen because peak ACL forces while
walking and early rehabilitation were estimated to be 303 N[33] and as 250 N is a commonly used load level in ACL testing.[8,15,29,30] The initial elongation was measured as the valley elongation from the
start of testing until the first cycle of the force control block was completed.
The dynamic elongation represents a relative valley elongation during
force-controlled cyclic loading. The total elongation is the sum of initial and
dynamic elongation.
Pull to Failure
For all constructs, a pull to failure at a rate of 50 mm/min was performed with
ultimate failure load and stiffness calculated and mechanism of failure noted.
The ultimate stiffness was determined with the linear portion of the
load-elongation curve within the load range of 200 N and 450 N. Cyclic loading
and load-to-failure data were recorded with Wavematrix software (Instron) with a
sampling rate of 500 Hz.
Statistical Analysis
The programming software MATLAB (v R2015b; MathWorks) was used for data analysis.
Statistical analysis was performed with Sigma Plot Statistics for Windows (v
13.0; Systat Software). The primary statistical analysis included a 1-way
analysis of variance (ANOVA) for each dependent variable. For ANOVAs that were
deemed significant, a Tukey post hoc test was performed to further analyze which
groups were different. Statistical significance was defined as
P ≤ .05, and the desired power level was set at 0.8.The Shapiro-Wilk test was used to confirm that each data set followed a normal
distribution. A nonparametric test, the Kruskal-Wallis, was used for data sets
that failed this test. For Kruskal-Wallis tests that found significance, a Tukey
post hoc test was conducted to further analyze the differences.The mean power value of all 1-way ANOVAs was 0.845, which was higher than the
desired power level of 0.8, leading us to conclude that our sample size was
sufficient.
Results
The results (mean ± SD) for each device are presented in Table 1. The P values for
all the tests among devices are reported in Table 2.
Bone breakage (50) Graft slippage (25) Combination (25)
Bone breakage (50) Graft slippage (25) Suture rupture (25)
Results are presented as mean ± SD unless noted otherwise. TR,
TightRope.
TABLE 2
P Values for Each Tukey Post Hoc Analysis
Adjustable-Loop Devices
Fixed-Loop Devices
GraftMax
TightRope
Ultrabutton
Endobutton
RetroButton
Initial force, N
GraftMax
—
<.001 (<.001)
<.001 (.008)
.440 (.933)
.431 (.945)
TightRope
<.001 (<.001)
—
.942 (.939)
<.001 (.008)
<.001 (.007)
Ultrabutton
<.001 (.008)
.942 (.939)
—
.002 (.081)
.002 (.073)
Endobutton
.440 (.933)
<.001 (.008)
.002 (.081)
—
≥.999 (≥.999)
RetroButton
.431 (.945)
<.001 (.007)
.002 (.073)
≥.999 (≥.999)
—
Force loss, N
GraftMax
—
<.001 (.002)
<.001 (<.001)
.100 (.689)
.413 (.881)
TightRope
<.001 (.002)
—
.284 (.905)
<.001 (.123)
<.001 (.049)
Ultrabutton
<.001 (<.001)
.284 (.905)
—
<.001 (.010)
<.001 (.003)
Endobutton
.111 (.689)
<.001 (.123)
<.001 (.010)
—
.942 (.996)
RetroButton
.413 (.881)
<.001 (.049)
<.001 (.003)
.942 (.996)
—
Initial elongation, mm
GraftMax
—
<.001
.339
.158
.635
TightRope
<.001
—
<.001
.002
<.001
Ultrabutton
.339
<.001
—
.992
.987
Endobutton
.158
.002
.992
—
.879
RetroButton
.635
<.001
.987
.879
—
Dynamic elongation, mm
GraftMax
—
≥.999 (.992)
.182 (.163)
.378 (.495)
.169 (.195)
TightRope
≥.999 (.992)
—
.150 (.055)
.324 (.241)
.139 (.069)
Ultrabutton
.182 (.163)
.150 (.055)
—
.992 (.968)
≥.999 (≥.999)
Endobutton
.378 (.495)
.324 (.241)
.992 (.968)
—
.989 (.981)
RetroButton
.169 (.195)
.139 (.069)
≥.999 (≥.999)
.989 (.981)
—
Total elongation, mm
GraftMax
—
.058 (.095)
.054 (.227)
.081 (.183)
.082 (.261)
TightRope
.058 (.095)
—
≥.999 (.995)
≥.999 (.998)
≥.999 (.990)
Ultrabutton
.054 (.227)
≥.999 (.995)
—
≥.999 (≥.999)
≥.999 (≥.999)
Endobutton
.081 (.183)
≥.999 (.998)
≥.999 (≥.999)
—
≥.999 (≥.999)
RetroButton
.082 (.261)
≥.999 (.990)
≥.999 (≥.999)
≥.999 (≥.999)
—
Ultimate failure load, N
GraftMax
—
.030
.999
.937
.782
TightRope
.030
—
.016
.004
.002
Ultrabutton
.999
.016
—
.984
.896
Endobutton
.937
.004
.984
—
.996
RetroButton
.782
.002
.896
.996
—
Stiffness, N/mm
GraftMax
—
.177
.234
.971
.673
TightRope
.177
—
<.001
.476
.008
Ultrabutton
.234
<.001
—
.068
.932
Endobutton
.971
.476
.068
—
.309
RetroButton
.673
.008
.932
.309
—
values in parentheses correspond to Tukey
post hoc analysis after nonparametric Kruskal-Wallis test.
Results for Each Device TestedResults are presented as mean ± SD unless noted otherwise. TR,
TightRope.P Values for Each Tukey Post Hoc Analysisvalues in parentheses correspond to Tukey
post hoc analysis after nonparametric Kruskal-Wallis test.A Kruskal-Wallis test was run to analyze statistical differences in the initial
force, as the data did not follow a normal distribution. There was a significant
difference between the TR and GM (P < .001), the TR and both
FLDs (EB, P = .008; RB, P = .007), and the UB
and GM (P = .008) (Figure 4).
Figure 4.
Initial and final force for each device during position-controlled
loading block.
Initial and final force for each device during position-controlled
loading block.The force loss, which was calculated as the difference between the initial and
final force data, did not follow a normal distribution; thus, a Kruskal-Wallis
test was performed. The change in force for the GM was significantly less than
for the UB (P < .001) or the TR (P = .002).
The UB had a significantly greater decrease in force than both FLDs. The TR had
a significantly greater decrease in force compared with the RB
(P = .049) but not the EB.The TR was the only device for which a negative initial elongation was observed.
There was a significant difference in initial elongation between the TR and all
other devices. No other statistical significance was found among other
devices.The 1-way ANOVA for dynamic elongation failed the normality test, so a
Kruskal-Wallis test was run. No statistically significant differences were found
among any of the devices. The power observed for dynamic elongation in the ANOVA
was 0.521, which was below the desired power level of 0.8. Dynamic elongation
for each device is illustrated in Figure 5.
Figure 5.
Box-and-whisker plot depicting dynamic elongation for each device. The
dynamic elongation represents a relative valley elongation during
force-controlled cyclic loading. Values are presented as median (line),
interquartile range (box), and 95% CI (vertical lines). Black points and
circles indicate mean values and outliers, respectively.
Box-and-whisker plot depicting dynamic elongation for each device. The
dynamic elongation represents a relative valley elongation during
force-controlled cyclic loading. Values are presented as median (line),
interquartile range (box), and 95% CI (vertical lines). Black points and
circles indicate mean values and outliers, respectively.For total elongation, the data sets also did not all follow a normal
distribution, and a Kruskal-Wallis test was run. There were no significant
differences among the groups (P = .094). The GM had the largest
total elongation, with 6 of 8 specimens exceeding the 3-mm threshold of clinical
failure, followed by the FLDs, TR, and UB (Figure 6). The post hoc power analysis
reported that the power for total elongation was 0.546, which was lower than the
desired level of 0.8.
Figure 6.
Box-and-whisker plot depicting initial and total elongation for each
device. The initial elongation was measured as the valley elongation
from the start of testing until the first cycle of the force control
block was completed. The total elongation is the sum of initial and
dynamic elongation. Values are presented as median (line), interquartile
range (box), and 95% CI (vertical lines). Black points and circles
indicate mean values and outliers, respectively.
Box-and-whisker plot depicting initial and total elongation for each
device. The initial elongation was measured as the valley elongation
from the start of testing until the first cycle of the force control
block was completed. The total elongation is the sum of initial and
dynamic elongation. Values are presented as median (line), interquartile
range (box), and 95% CI (vertical lines). Black points and circles
indicate mean values and outliers, respectively.
Ultimate Failure Load
The largest pull-to-failure force was observed for the TR, which was
statistically significantly different than all other devices (Figure 7). No other
statistically significant differences were observed. Regarding ultimate
stiffness, the TR had a significantly decreased stiffness when compared with the
UB (P < .001) and the RB (P = .008). The
methods of failure differed among the devices. The main failure mode was suture
slippage for the GM, suture rupture for the TR, and button breakage combined
with femoral bone breakage for the UB. For the FLDs, the main methods of failure
were either femoral bone breakage or graft slippage.
Figure 7.
Box-and-whisker plot depicting ultimate failure load for each device.
Values are presented as median (line), interquartile range (box), and
95% CI (vertical lines). Black points indicate mean values.
Box-and-whisker plot depicting ultimate failure load for each device.
Values are presented as median (line), interquartile range (box), and
95% CI (vertical lines). Black points indicate mean values.
Discussion
This full–ACL construct study included a test methodology with intraoperative graft
preconditioning containing precycling and retensioning according to a surgical
technique workflow under in vitro loading parameters that replicate the in vivo ACL
environment. All clinically relevant treatment options for FLDs and ALDs were
utilized to objectively compare the devices and allow for graft optimization.
Retensioning is a major benefit for ALDs, which is not possible for FLDs after
primary fixation. During precycling, which simulates intraoperative knee flexion,
primary elongations are due to the settling effects of the ACLR. Apart from possible
adverse effects of excessive graft tensioning, such as abnormal articulation and
cartilage or graft degeneration,[3,34] the benefit of retensioning ALDs was shown to eliminate these elongations and
further optimize graft tension to mitigate ultimate knee laxity.[29]Retensioning allows for ALDs to establish a higher initial force level as compared
with an FLD within the position control block. During simulated knee flexion in the
position control block, a slack was introduced within the graft, similar to ACL
behavior during midflexion angles.[4,31,37] This repetitive graft loading-unloading situation (0 N) creates an
unfavorable loading condition for an ALD to experience loop lengthening after ACLR,
which could be a reason for a greater force loss at reaching the end of position
control. An FLD is not affected by a complete unloading situation, owing to its
continuous loop design. Within this study, the all-inside TR constructs demonstrated
the highest initial force and, at a force loss of 43%, the highest absolute final
force. Because our test protocol included position-controlled cycling simulating
early rehabilitation, it can be assumed that the high tension that was built up
during intraoperative preconditioning would also be preserved in the patient’s knee
over the first days subsequent to ACLR.In our study, the UB and the GM revealed the highest and lowest force loss during
position-controlled loading, with a 54% and 23% decrease as referenced to initial
force, respectively. The GM could not be retensioned to achieve the desired 200-N
level in full extension and had an absolute lower initial force of 95.5 N (see the
online Video Supplement). The GM button with its locking mechanism design limits the
retensioning. During loop shortening, the pulling sutures lift up the locking suture
loop to allow for adjustable loop shortening. After the pulling sutures are
released, the locking suture loop above the button is given slack and moves downward
with the pulling sutures into the button-locking pocket when tension is applied on
the graft side. Although retensioning was performed several times, no further
increase of the initial force was possible, owing to this locking-unlocking
mechanism. Only the GM utilizes a button-locking mechanism, while the TR and UB
utilize a suture-locking mechanism. The lower absolute initial force level as well
as the difference in the locking mechanism might explain the decreased force loss
during position control cycling.However, there are differences among the ALDs, as evidenced by the amount of initial
elongation. For TR constructs, a negative initial elongation was observed. This
finding suggests that retensioning was more effective for this device, resulting in
the highest initial force level. Combined with a decreased force loss, less
elongation was needed to reach the valley load of 10 N after the first peak of 250 N
in force-controlled cyclic loading. This resulted in a significant difference in
initial elongation between an all-inside graft with TR and all other devices
utilizing tibial screw fixation (P = .002 for EB;
P < .001 for GM, UB, and RB).The majority of ACLR elongation (dynamic) occurred during the force-controlled cyclic
loading, which could be attributed to graft viscoelastic stretching of the soft
tissue material and to fixation device elongation. Because no statistically
significant differences were found among any of the devices, it could be assumed
that the stretching of the soft tissue material plays the major role for ACLR
elongation. However, the statistical analysis regarding the dynamic elongation was
underpowered. The highest amount of dynamic elongation was assessed for the TR,
which might be a result of a longer elongation distance owing to an increased graft
construct length with extracortical fixation points and all-inside graft
preparation.Another difference that was noted but not statistically different among devices was
total elongation. As clinical failure of ACLR was reported as a side-to-side
difference of >3 mm as measured with the KT-1000 during anterior tibial
translation, we used this value as the failure threshold for total elongation.[13] The GM was the only device with a total elongation >3 mm (6 of 8 specimens
exceeded the threshold), which may be explained by the aforementioned retensioning
limitation resulting in higher initial elongation.The pull-to-failure loads for all ALD constructs were greater than the forces
experienced during walking and early rehabilitation of the ACL[33]; therefore, all the ALDs can be interpreted as a suitable fixation option
when benchmarked to FLDs. The TR had the largest pull-to-failure load
(P ≤ .030 compared with all other constructs), followed by the
GM and UB. A possible explanation of this finding is the graft preparation for an
all-inside technique with both-sided cortical button fixation rather than an
interference screw for tibial graft fixation. Tibial fixation was previously
considered to be the weakest link of a construct, with lower pull-to-failure loads.
This was also proved in our model.[10,25,32,35]Additionally, Chandrashekar et al[11] reported a stiffness of 199 ± 88 N/mm and 308 ± 89 N/mm for females and
males, respectively, leading to a combined mean stiffness of 250 ± 102 N/mm.
Stiffness results of ACL constructs tested in this study were within the overall
range (111-397 N/mm), especially for male patients on the lower side. The lowest
stiffness was observed for all-inside constructs with TR, which can be explained by
the longer elongation distance of this technique. Especially for those less-stiff
graft constructs, a high initial tension is beneficial, as shown by Amis and Jakob.[3]For the UB and GM, it was difficult to compare findings with other studies, as these
devices have not been formally tested at the time of this study. Furthermore,
comparison between the test results of our study and others is challenging owing to
differences of the test protocol and test setup.Limitations of this study include the use of porcine tibias and femurs as well as
bovine tendons as substitutes for human tissue. However, these have been shown to
best resemble human bones in density[1,28] and tendons. Force application was done in line with the tunnel axis, which
differs from clinical observations but is in accordance with a worst-case loading
scenario for ACLR testing.Statistical analysis regarding dynamic and total elongation did not show significant
difference yet was underpowered. This increases the likelihood of not detecting an
existing difference and thereby constitutes a further limitation to this study.Moreover, this study was designed as a surgical technique–based study allowing ideal
circumstances for each device. Therefore, the TR was tested as an all-inside
construct with graft tunnel docking, and the GM was retensioned at 30° instead of 0°
of flexion. This means that devices were used in an application-oriented manner at
the expense of full comparability. We felt that the limitation resulting from the
chosen study design was acceptable with regard to the benefits.This time-zero in vitro biomechanical study does not factor in any postoperative bone
healing that might occur and cause mitigated knee laxity. Clinical studies are
required to further support the GM and UB devices. This biomechanical model may also
facilitate additional research as techniques and devices advance.
Conclusion
The results of the current study suggest that ALDs behave comparably with FLDs with
regard to biomechanical fixation strength and ultimate knee laxity while also having
unique advantages as compared with FLDs, such as an increased bone-tendon interface
and a simplified application. However, some differences between the tested ALDs were
observed. The GM button design prevented proper retensioning, so the initial force
was less than, and total elongation greater than, the TR or the UB. During complete
loading-unloading situations, the UB had the largest force loss. Last, the TR
achieved the smallest initial elongation with the greatest ultimate failure load as
compared with all other devices.
Authors: Raul Mayr; Christian Heinz Heinrichs; Martin Eichinger; Christian Coppola; Werner Schmoelz; René Attal Journal: Am J Sports Med Date: 2015-03-12 Impact factor: 6.202
Authors: Benjamin M Petre; Sean D Smith; Kyle S Jansson; Peter-Paul de Meijer; Thomas R Hackett; Robert F LaPrade; Coen A Wijdicks Journal: Am J Sports Med Date: 2012-12-20 Impact factor: 6.202
Authors: Nathan A Mall; Peter N Chalmers; Mario Moric; Miho J Tanaka; Brian J Cole; Bernard R Bach; George A Paletta Journal: Am J Sports Med Date: 2014-08-01 Impact factor: 6.202
Authors: Michael W Mariscalco; David C Flanigan; Joshua Mitchell; Angela D Pedroza; Morgan H Jones; Jack T Andrish; Richard D Parker; Christopher C Kaeding; Robert A Magnussen Journal: Arthroscopy Date: 2013-10-17 Impact factor: 4.772
Authors: Meghan W Richardson; Nicholas D Tsouris; Chaudry R Hassan; Justen H Elbayar; Yi-Xian Qin; David E Komatsu; Angelo V Rizzi; James M Paci Journal: Arthroscopy Date: 2019-04-12 Impact factor: 4.772
Authors: Peter J C McEwen; Milford McArthur; Sarah G Brereton; William B O'Callaghan; Matthew P R Wilkinson Journal: Asia Pac J Sports Med Arthrosc Rehabil Technol Date: 2020-10-07
Authors: Bastian Uribe-Echevarria; Justin A Magnuson; Annunziato Amendola; Matthew J Bollier; Brian R Wolf; Carolyn M Hettrich Journal: Iowa Orthop J Date: 2020
Authors: Tobias Götschi; George Rosenberg; Xiang Li; Chen Zhang; Elias Bachmann; Jess G Snedeker; Sandro F Fucentese Journal: Orthop J Sports Med Date: 2020-02-25