Literature DB >> 15090378

Mechanical properties of soft tissue femoral fixation devices for anterior cruciate ligament reconstruction.

Christopher S Ahmad1, Thomas R Gardner, Megan Groh, Johnny Arnouk, William N Levine.   

Abstract

PURPOSE: To evaluate femoral soft tissue fixation for anterior cruciate ligament reconstruction. HYPOTHESIS: Femoral fixation devices have different ultimate strengths and slippage under cyclic loading. STUDY
DESIGN: Controlled laboratory study.
METHODS: Thirty-three porcine femora were used to study interference screw (9), Endobutton (8), Rigidfix cross-pin (8), and Bio-Transfix cross-pin (8) fixation methods. Fixation slippage was evaluated under cyclical load from 50 N to 250 N using a materials testing machine. Ultimate load was determined with a single load to failure.
RESULTS: Total graft slippage was greater (P <.001) for the Rigidfix (6.02 +/- 2.12 mm) and the interference screw (5.44 +/- 3.25 mm) compared to the Endobutton (1.75 +/- 0.97 mm) and the Bio-Transfix (1.14 +/- 0.53 mm). All techniques showed the greatest slippage during the first 100 cycles (Rigidfix 84%, Endobutton 70%, interference screw 56%, and Bio-Transfix 55%). The failure load for the interference screw technique (539 +/- 114 N) was lower (P =.0008) than for the other 3 techniques (737 +/- 140 N for Rigidfix, 746 +/- 119 N for Bio-Transfix, and 864 +/- 164 N for Endobutton).
CONCLUSIONS: The interference screw and the Rigidfix fixation demonstrated inferior fixation biomechanics compared to the Bio-Transfix and the Endobutton techniques.

Mesh:

Year:  2004        PMID: 15090378     DOI: 10.1177/0363546503261714

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  61 in total

1.  Anterior cruciate ligament reconstruction using biodegradable transfemoral fixation at 5-year follow-up: clinical and magnetic resonance imaging evaluation.

Authors:  Sven Nebelung; Gregor Deitmer; Rolf Gebing; Frank Reichwein; Wolfgang Nebelung
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-03-06       Impact factor: 4.342

2.  Interference screws should be shorter than the hamstring tendon graft in the bone tunnel for best fixation.

Authors:  Michael Stalder; Mazda Farshad; Jess G Snedeker; Dominik C Meyer
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-03-21       Impact factor: 4.342

3.  The fixation strength of a novel ACL soft-tissue graft fixation device compared with conventional interference screws: a biomechanical study in vitro.

Authors:  Camilla Halewood; Michael T Hirschmann; Simon Newman; Jaffar Hleihil; Gershon Chaimski; Andrew A Amis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-09-14       Impact factor: 4.342

4.  Pretensioning of quadruple flexor tendon grafts in two types of femoral fixation: quasi-randomised controlled pilot study.

Authors:  Cezar Teruyuki Kawano; Patrícia Maria de Moraes Barros Fucs; Nilson Roberto Severino
Journal:  Int Orthop       Date:  2010-05-22       Impact factor: 3.075

5.  What is the best femoral fixation of hamstring autografts in anterior cruciate ligament reconstruction?: a meta-analysis.

Authors:  Alexis Colvin; Charu Sharma; Michael Parides; Jonathan Glashow
Journal:  Clin Orthop Relat Res       Date:  2011-04       Impact factor: 4.176

6.  The biomechanical strength of a hardware-free femoral press-fit method for ACL bone-tendon-bone graft fixation.

Authors:  M P Arnold; L D Burger; D Wirz; B Goepfert; M T Hirschmann
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-01-07       Impact factor: 4.342

7.  Breakage of bioabsorbable femoral transfix screw after arthroscopic anterior cruciate ligament repair: a report of two cases.

Authors:  A Raviraj; Ashish Anand; K Santosh
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-02-04

8.  Anterior cruciate ligament reconstruction using the Bio-TransFix femoral fixation device and anteromedial portal technique.

Authors:  Michael E Hantes; Zoe Dailiana; Vasilios C Zachos; Sokratis E Varitimidis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2005-10-14       Impact factor: 4.342

9.  Transtibial double bundle posterior cruciate ligament reconstruction using TransFix tibial fixation.

Authors:  Yong Seuk Lee; Jin Hwan Ahn; Young Bok Jung; Joon Ho Wang; Jae Chul Yoo; Ho Joong Jung; Bun Jung Kang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-02-16       Impact factor: 4.342

10.  Comparison of bioabsorbable interference screws and posts for distal fixation in anterior cruciate ligament reconstruction.

Authors:  Krzysztof Gaweda; Jacek Walawski; Robert Wegłowski; Wojciech Krzyzanowski
Journal:  Int Orthop       Date:  2007-12-07       Impact factor: 3.075