Xiaobo Wei1, Mengjiao Liu1, Yun Ding1, Qilin Li1, Changhai Cheng1, Xian Zong1, Wenming Yin1, Jie Chen1, Wendong Gu2.
Abstract
Breast-conserving surgery (BCS) plus postoperative radiotherapy has become the standard treatment for early-stage breast cancer. The aim of this study was to compare the setup accuracy of optical surface imaging by the Sentinel system with cone-beam computerized tomography (CBCT) imaging currently used in our clinic for patients received BCS. Two optical surface scans were acquired before and immediately after couch movement correction. The correlation between the setup errors as determined by the initial optical surface scan and CBCT was analyzed. The deviation of the second optical surface scan from the reference planning CT was considered an estimate for the residual errors for the new method for patient setup correction. The consequences in terms for necessary planning target volume (PTV) margins for treatment sessions without setup correction applied. We analyzed 145 scans in 27 patients treated for early stage breast cancer. The setup errors of skin marker based patient alignment by optical surface scan and CBCT were correlated, and the residual setup errors as determined by the optical surface scan after couch movement correction were reduced. Optical surface imaging provides a convenient method for improving the setup accuracy for breast cancer patient without unnecessary imaging dose.Entities:
Mesh:
Year: 2018 PMID: 29740104 PMCID: PMC5940773 DOI: 10.1038/s41598-018-25644-w
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Correlation studies of CE and SE-before.
| CE | SE-before | Correlation Coefficient | P-value | |
|---|---|---|---|---|
| X(left-right) | 8 mm ± 2.79 mm | −0.62 mm ± 3.13 mm | 0.796 | <0.001 |
| Y(cranial-caudal) | −2.55 mm ± 2.97 mm | −2.67 mm ± 2.93 mm | 0.723 | <0.001 |
| Z(anteroposterior) | −3.58 mm ± 3.17 mm | −2.21 mm ± 2.97 mm | 0.765 | <0.001 |
| Rx(x-rotation) | −0.1° ± 0.8° | 0.2° ± 1.0° | 0.63 | <0.001 |
| Ry(y-rotation) | 0.7° ± 0.8° | 0.6° ± 0.9° | 0.741 | <0.001 |
| Rz(z-rotation) | 0.3° ± 1.0° | 0.2° ± 0.9° | 0.541 | <0.001 |
CE: cone beam CT errors; SE-before: sentinel errors before the treatment couch correction.
Figure 1The scatter spot of CBCT and Sentinel error before moving couch. The correlation of CE and SE-before in the three translation directions (a–c). The correlation of CE and SE-before in the three rotations about the X, Y, Z axes (d–f).
Setup error of CBCT and sentinel before moving the treatment table.
| The group mean (M) | Systemic error (Σ) | Random error (δ) | PTV margin | ||
|---|---|---|---|---|---|
| X-shift | CBCT (mm) | −0.64 | 2.49 | 1.45 | 6.00 |
| Sentinel (mm) | −0.67 | 2.62 | 1.98 | 6.63 | |
| Y-shift | CBCT (mm) | −2.47 | 2.80 | 1.48 | 6.64 |
| Sentinel (mm) | −2.64 | 2.52 | 1.82 | 6.31 | |
| Z-shift | CBCT (mm) | −3.59 | 3.04 | 1.51 | 7.14 |
| Sentinel (mm) | −2.19 | 2.70 | 1.68 | 6.58 | |
| X-rotation | CBCT (°) | − 0.1 | 0.6 | 0.5 | — |
| Sentinel (°) | 0.1 | 0.8 | 0.7 | — | |
| Y-rotation | CBCT (°) | 0.7 | 0.7 | 0.5 | — |
| Sentinel (°) | 0.6 | 0.6 | 0.6 | — | |
| Z-rotation | CBCT (°) | 0.3 | 0.8 | 0.7 | — |
| Sentinel (°) | 0.2 | 0.6 | 0.8 | — | |
CBCT: cone-beam computerized tomography.
Residual setup error of Sentinel.
| The group mean (M) | Systemic errors (∑) | Random errors (σ) | PTV margin | |
|---|---|---|---|---|
| X-shift (mm) | 0.1 | 1.7 | 1.56 | 4.49 |
| Y-shift (mm) | 1.36 | 1.85 | 1.64 | 4.85 |
| Z-shift (mm) | −1.28 | 1.32 | 1.47 | 3.67 |
| X-rotation (°) | 0.3 | 0.6 | 0.7 | — |
| Y-rotation (°) | −0.1 | 0.9 | 0.7 | — |
| Z-rotation (°) | 0.1 | 0.7 | 0.8 | — |
PTV: planning target volume.
Characteristics of the study population (n = 27).
| Clinical characteristics | Value |
|---|---|
|
| |
| Intraductal carcinoma (n) | 6 |
| Invasive carcinoma (n) | 21 |
|
| |
| pTisN0M0 (n) | 6 |
| pT1N0M0 (n) | 17 |
| pT2N0M0 (n) | 4 |
| Mammary gland volume (cm3) | 479.4 ± 180.9 |
| Age (years old) | 49.5 ± 7.2 |
| The average BMI (kg/m2) | 24.8 ± 3.1 |
BMI: body mass index.
Figure 2The images and data of setup corrections by Sentinel (a) and CBCT (a,b): The green area represents the body surface profile reconstructed by CT scan, the blue area represents the body surface profile reconstructed by Sentinel scan, and the dark blue represents areas that match with errors. The data on the left is the result of matching error.
Figure 3Flow chart of research steps.