A Treatment Planning Method for Better Management of Radiation-Induced Oral Mucositis in Locally Advanced Head and Neck Cancer.

Hao Howard Zhang1, Warren D D'Souza1.   

Abstract

PURPOSE/AIM: To describe a two-phase intensity-modulated radiation therapy (IMRT) treatment planning approach, that is, promising for reduction of oral mucositis risk in locally advanced head-and-neck cancer.
MATERIALS AND METHODS: Ten locally advanced head-and-neck cancer patients who underwent RT were retrospectively collected. Conventional IMRT and volumetric-modulated arc therapy (VMAT) plans were generated for these patients following clinical protocol. Following the first phase of generating conventional IMRT plans, our approach utilized data from Monte Carlo-based kernel superposition dose calculations corresponding to beam apertures (generated from the conventional IMRT plans) and used an exact mathematical programming-based optimization approach applying linear programming (LP) to dose optimization in the second phase.
RESULTS: Compared with conventional IMRT and VMAT treatment plans, our novel method achieved better preservation of oral cavity (16%-29% lower mean dose, P < 0.01), parotid glands (6%-17% lower mean dose, P < 0.04), and spinal cord (3-11 Gy lower maximum dose, P < 0.03) and lower doses to nonorgan-at-risk/nontarget normal tissues, with the same or better target coverage.
CONCLUSIONS: Our LP-based method can be practically implemented in routine clinical use with a goal of limiting radiation-induced oral mucositis for head-and-neck cancer patients.

Entities:  

Keywords:  Head and neck cancer; oral mucositis; radiotherapy; treatment planning

Year:  2018        PMID: 29628628      PMCID: PMC5879831          DOI: 10.4103/jmp.JMP_78_17

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


INTRODUCTION

Worldwide, it is estimated that there are approximately 686,000 new head-and-neck cancers (HNCs) diagnosed every year and 375,000 individuals will die annually as a result of the disease.[1] Locally advanced HNC represents one of the most challenging treatment planning scenarios in radiation therapy (RT). Mucositis is among the most common adverse reactions encountered in RT for HNC. Franzese et al. reported that 54% of patients treated with volumetric-modulated arc therapy developed ≥grade 2 (G2) oral mucositis.[2] Other studies reported that 73%–100% of patients treated with intensity-modulated RT (IMRT) experienced grade 3/4 mucositis[34] and severe oral mucositis occurred in 29%–66% of all patients receiving RT.[56] Oral mucositis is often painful and can significantly affect nutritional intake, mouth care, and quality of life,[7] as well as increase the risk of local and life-threatening systemic infections.[89] Patients with oral mucositis are significantly more likely to have severe pain and loss of >5% of pretreatment body weight.[6] In one study, ~16% of patients receiving RT were hospitalized because of mucositis.[10] Moreover, severe oral toxicities can also compromise the delivery of optimal cancer therapy protocols. Trotti et al. reported that 11% of patients receiving RT for HNC had unplanned breaks in RT because of severe mucositis.[10] These disruptions in dosing as a result of oral complications can directly affect survivorship. Thus, oral mucositis is a major dose-limiting toxicity of RT to the head and neck region.[8911] In addition to serious clinical effects, radiation-induced oral mucositis contributes to economic concerns because of costs associated with pain management, liquid diet supplements, gastrostomy tube placement or total parenteral nutrition, management of secondary infections, and hospitalizations.[12] In a retrospective study, oral mucositis was associated with an increase in costs ranging from $1,700 to $6,000 per patient, depending on the grade of mucositis.[5] In summary, mucositis causes statistically significant increases in direct medical costs, averaging ~$3,000 ± $1,000 per treatment episode.[1314] Despite the high incidence, associated morbidity, and effects on patient quality of life and economic consequences, no standard RT technique is approved for the prevention of mucositis in HNC.[15] The radiation oncology community has concentrated on studies designed to determine whether altering fractionation schedules can improve cancer control, such as hyperfractionation and concomitant boost fractionation for HNC.[16] However, the incidence of grade 3 or 4 acute mucositis increased by 16%–22% in patients treated with accelerated fractionation or hyperfractionation compared with the conventional schedule.[17] It has been shown that oral mucositis is correlated with doses received by the oral cavity (e.g., 15, 30, 40, 45, and 50 Gy).[18192021] Cumulative doses to the oral cavity of 32 Gy were associated with acute mucositis, and a dose >39 Gy was associated with longer duration of mucositis.[22] In an attempt to minimize radiation-induced mucositis, Sanguineti et al. suggested delivering the smallest possible dose to the oral cavity.[23] In this research, we provide a treatment plan, that is, promising for reduction of the risk of oral mucositis in locally advanced HNC compared to clinical treatment plans generated using conventional IMRT and VMAT.

MATERIALS AND METHODS

Data

In this retrospective, institutional review board-approved study, we collected data from 10 patients with locally advanced HNC who had undergone RT in our clinic. Treatment targets included the primary tumor and cervical nodes. A simultaneous integrated boost was employed as the mode of delivery, and the prescribed doses to the primary tumor and high-and low-risk (LR) nodes were 70, 59.4, and 54 Gy, respectively, in 35 fractions. Organs-at-risk (OARs) included the parotid glands, oral mucosa surrounding the oral cavity, brain stem, and spinal cord. We used previously published dose–volume constraints in the optimization process [Table 1].[24]
Table 1

Summary of dose/dose-volume constraints

Summary of dose/dose-volume constraints

Treatment plans

Two types of IMRT plans were generated for each patient. The first was generated using the Pinnacle3 planning system (Philips Healthcare; Fitchburg, WI, USA). The second was generated using our linear programming (LP)-based approach.[25] Although IMRT plans were also generated using the Eclipse planning system (Varian Medical Systems; Palo Alto, CA, USA), both planning systems resulted in comparable IMRT plans. We describe the planning specifics and results for only the Pinnacle3 planning system in this report.

Conventional intensity-modulated radiation therapy plans

In this work, nine beams were used in IMRT treatment planning. Selected angles were input to the commercial IMRT planning systems. The Pinnacle3 planning system used a gradient-based optimization approach for conventional IMRT. These plans will be referred to as Pinn IMRT plans.

Linear programming-based intensity-modulated radiation therapy plans

IMRT plans were also generated using an LP-based optimization approach (referred to as LP IMRT).[25] The details of the LP-based approach can be seen from Zhang et al.[25] Here, we summarize the approach: The LP-based approach allows for explicit incorporation of dose and dose–volume constraints. An advantage of this approach is that if the problem is feasible and bounded (which is the case for dose optimization problems), formulating and solving it using a linear program guarantees the optimal solution. Our approach employed the principle of segment weight optimization. In the first phase, the Pinn IMRT plan was used to generate an initial set of apertures. These physical apertures were used to initiate a second-phase optimization process that utilized aperture-specific dose calculations to provide input data to an LP-based dose optimization solver. The details of the dose optimization problem are as follows: where i is an index corresponding to beam apertures and j is an index corresponding to voxels over which constraints are specified. The LP dose optimization objective involves minimizing the weighted sum of doses delivered to the voxels that violate the thresholds for dose–volume histogram (DVH) constraints. The treatment volume is divided into region O, including OARs; region T, including the planning target volumes; and region N, which corresponds to normal tissue voxels. In this work, normal tissue is defined as nontarget and non-OAR tissue. Input data include beam aperture-specific dose matrices B corresponding to each delivery aperture. DVH constraints are handled through the introduction of penalty variables that account for target, OAR, and normal tissue underdose and overdose as relevant. b,b, b, b and b specify the desired dose thresholds for the target, OAR, normal tissue, and any designated region, respectively, at which constraints are desired. x, j ∈ T, is a vector of underdoses/overdoses (doses below the prescription level or above a certain level [hot spot] for target voxels) delivered to the target region. The variable x, j ∈ O and j ∈ N is a vector of overdoses (doses over a certain critical level) delivered to the OARs and normal tissue. S is a subset of O, which corresponds to a special region, such as the spinal cord voxels, in which hard upper-bound dose constraints instead of DVH constraints are specified. w is the weight (to be optimized) for aperture i of beam angle set θ. α, α,α, and α are the penalty weights assigned for underdosing, overdosing the target, and overdosing the OARs and normal tissues, respectively. Each type of constraint can be enforced on multiple targets and OARs. Multiple constraints can be enforced on each target and OAR as well. The dose optimization problem was formulated in the General Algebraic Modeling System (GAMS Development Corporation; Washington, DC, USA);[26] a high-level modeling system for mathematical programming and optimization. The linear program was then solved with an integrated high-performance commercial solver, Parallel CPLEX (IBM; Armonk, NY, USA)[27] on a cluster of 48 Intel 2.4 GHz CPUs available at our clinic.

Volumetric-modulated arc therapy plans

Single-arc and double-arc VMAT plans were generated using the Pinnacle3 planning system. The Pinnacle3 planning system adopts a multistep optimization algorithm developed by Bzdusek et al. which includes initial segments generation, arc sequencing, machine parameter optimization, and segment weight optimization steps.[28] For each arc, 180 apertures (control points) were used, resulting in 180 apertures for the single-arc (referred to as 1ARC) plans and 360 apertures for the double-arc (referred to as 2ARC) plans. This setting per arc corresponds to the maximum number of apertures available for VMAT from Pinnacle 3. A summary of the treatment plans generated and their differences are shown in Table 2.
Table 2

Summary of treatment plans generated for comparison

Summary of treatment plans generated for comparison

Comparison of plans

We retrospectively tested our approach on data from 10 patients with locally advanced HNC who had undergone conventional RT in our clinic. To compare different treatment plans, we let each approach improve the OAR and normal tissue sparing as much as possible by adjusting weights associated with the individual dose and dose–volume constraints. The adjustments were guided using the best plan obtained from a competing technique as an improvement goal.[29] Each plan was allowed up to 2 h of planning time for improvement. All plans were normalized so that 95% of the target received at least the prescription dose. Final dose distributions for all plans were compared using a previously validated Monte Carlo-based convolution-superposition dose calculation approach.[30] The dosimetric quality of the plans was compared using DVHs and isodose lines. Dose distributions were compared through the conformity index (CI) and dose gradient index (DGI), defined as: CI = (prescription isodose volume)/(target volume)(8) DGI = (50% of prescription isodose volume)/(100% of prescription isodose volume)(9)

Comparison of delivery efficiency

The delivery efficiency of each technique was compared using the number of apertures, monitor units (MUs), and treatment delivery time for each plan [results shown in Table 3]. A maximum dose rate of 600 MU/min was assumed when calculating delivery time for both IMRT and VMAT. Treatment delivery time was approximated by assuming that the beam off time during transition between successive apertures for a given beam angle in IMRT averaged 0.5 s. The maximum gantry rotation speed was 4.8°/s. For IMRT, the delivery time (in seconds) was calculated as:
Table 3

Comparison of total monitor units, number of apertures, calculated delivery time, conformal index, and dose gradient index

Comparison of total monitor units, number of apertures, calculated delivery time, conformal index, and dose gradient index where A is the number of apertures and θl, is the distance (in degrees) between the first and last beam. The beam mode-up time for IMRT was ignored for estimating treatment delivery time. Of interest is the fact that individual beam mode-up times are eliminated on the new TrueBeam™ (Varian Medical Systems; Palo Alto, CA, USA) instrumentation. For VMAT delivery, equation (11) was used to obtain the VMAT delivery time: s is the actual rotation speed at control point l (recall the spacing between the control points is 2°): where S is the actual rotation speed at control point l.

RESULTS

Qualitative dosimetric results are shown as DVH plots [Figure 1] and isodose curves [Figure 2]. Quantitative results were collected at OAR dose constraint setting levels as shown in Figure 3.
Figure 1

Mean dose–volume histogram comparisons in 10 head-and-neck cases for: (a) Primary planning treatment volume, brain stem, and right parotid; (b) High-risk planning treatment volume and oral cavity; (c) Low-risk planning treatment volume and left parotid; and (d) Spinal cord and normal tissue for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans

Figure 2

Isodose curves for: (a) Single-arc volumetric-modulated arc therapy plans; (b) Double-arc volumetric-modulated arc therapy plans; (c) Conventional intensity-modulated radiation therapy plans generated by Pinnacle 3; and (d) Linear programming intensity-modulated radiation therapy plans

Figure 3

Comparison of doses (in Gy) at constraint levels for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans

Mean dose–volume histogram comparisons in 10 head-and-neck cases for: (a) Primary planning treatment volume, brain stem, and right parotid; (b) High-risk planning treatment volume and oral cavity; (c) Low-risk planning treatment volume and left parotid; and (d) Spinal cord and normal tissue for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans Isodose curves for: (a) Single-arc volumetric-modulated arc therapy plans; (b) Double-arc volumetric-modulated arc therapy plans; (c) Conventional intensity-modulated radiation therapy plans generated by Pinnacle 3; and (d) Linear programming intensity-modulated radiation therapy plans Comparison of doses (in Gy) at constraint levels for intensity-modulated radiation therapy and volumetric-modulated arc therapy plans

Plan dosimetric quality

Figure 1 shows the mean DVHs of the 10 patient records considered in this research. Four approaches were compared, with solid lines representing the two IMRT approaches and dashed lines representing the two VMAT approaches. Primary planning treatment volume, high-risk PTV, and LR PTV DVHs in Figure 1 show that all plans achieved essentially the same target coverage. LP IMRT achieved the best dose uniformity, and 1ARC VMAT resulted in the worst. Pinn IMRT and 2ARC VMAT had similar dose uniformity. The maximum difference between PTV DVHs was <10%. In comparing OAR sparing across different treatment plans, Figure 1a shows that all four plans achieved similar brainstem sparing. Pinn IMRT and both VMAT plans resulted in the same sparing of the right parotid. LP IMRT spared 15% more of the right parotid than the other three approaches. From oral cavity DVHs shown in Figure 1b, the LP IMRT plan was superior for preserving oral cavity; 20% better than the Pinn IMRT and 2ARC VMAT plans, which were 10% better than the 1ARC VMAT plan. Left parotid sparing was essentially the same (difference <10%) for all four plans, as shown in Figure 1c. For spinal sparing, the 2ARC VMAT plan delivered 10% less maximum dose to spinal cord than the 1ARC VMAT plan [Figure 1d]. LP IMRT delivered 6% and 12% less maximum doses to spinal cord than the Pinn IMRT and 2ARC VMAT plans, respectively. Looking at more detailed quantitative results (absolute dose in Gy received by each OAR at constraint setting levels, Figure 3), the LP IMRT plan resulted in 2–5 Gy (6%–17%) lower mean doses for both parotids than the other three plans. For oral cavity mean doses, the LP IMRT plan was 5–11 Gy (16%–29%) lower than the other plans. The maximum spinal cord dose of the LP IMRT plan was 3–11 Gy lower than the other plans. Figure 2 shows the isodose curves on an axial slice of one of the cases as an example of dose distribution comparison. Only the 1ARC VMAT plans showed hot spots in the primary PTV. The difference in target coverage in isodose curves between the IMRT plans and 2ARC VMAT plan was negligible. LP IMRT plans displayed better OAR sparing for parotids, oral cavity, and spinal cord than the other plans. From the last two columns of Table 3, we observe that the average CI and DGI achieved by the Pinn IMRT and 2ARC VMAT plans were similar. The LP IMRT plan resulted in the lowest average CI and DGI, whereas the 1ARC VMAT plan resulted in the highest. Paired t-tests were conducted to determine whether differences achieved at OAR dose constraint setting levels between different plans were statistically significant [Table 4]. We can see that except for brain stem sparing, the differences between LP IMRT and the other three plans were significant. The tests confirmed the superior preservation of oral cavity and parotid glands achieved by LP IMRT.
Table 4

P values of paired t-test comparing different treatment plans

P values of paired t-test comparing different treatment plans The mean and standard deviation of total MUs, number of apertures, and treatment delivery time for all plans are summarized in Table 3. The 2ARC plans resulted in 20% more MUs than 1ARC plans. LP IMRT plans resulted in 10% more MUs than Pinn IMRT plans. IMRT plans at least doubled the MUs used by VMAT plans. The number of apertures in the IMRT plans was generally between that of the 1ARC VMAT and 2ARC VMAT plans, which were ~180 and ~360, respectively. LP IMRT plans reduced the number of apertures by 25% compared with Pinn IMRT plans. LP IMRT plans resulted in zero weights for a significant number of the Pinn IMRT apertures while improving plan quality. This finding ultimately resulted in 8% shorter calculated LP IMRT delivery times than Pinn IMRT delivery times. 2ARC VMAT delivery times were ~2.5 min shorter than the estimated LP IMRT delivery times.

DISCUSSION

Our results showed that double-arc VMAT plans improved plan quality over single-arc VMAT plans. IMRT plans and double-arc VMAT plans generated by Pinnacle 3 were comparable. However, our LP-based IMRT consistently provided better overall plan quality than both the IMRT and double-arc VMAT plans. The CI and DGI of LP IMRT plans were consistently better than those with other plans. DVHs for normal tissues were better for LP IMRT plans than for VMAT plans. For locally advanced HNC cases, OARs either sit within the concavity of the targets or are very close to the targets. Mallick and Waldron[31] showed that the rate of oral mucositis is quite high in patients being treated with fields involving the oral cavity. Since VMAT treatment plans deliver radiation through all possible beams around the patient, it is difficult to avoid those fields involving the oral cavity. As our results showed, VMAT plans were not as competitive as IMRT delivery techniques for preserving the oral cavity. It has been shown that most patients who receive >50 Gy to the oral mucosa will develop severe ulcerative oral mucositis.[18] Clinical severity is directly proportional to the radiation dose administered. Shogan et al. showed a correlation between oral cavity dose and mucositis when using IMRT.[19] A statistically significant correlation between acute mucositis grade and percentage of volume of oral cavity receiving 15, 30, 40, and 45 Gy was identified. In other studies, a correlation between grade 3+ mucositis and oral cavity doses of 9.5 and 10.1 Gy per week was found.[2021] Cumulative doses of 32 Gy to the oral cavity were associated with acute mucositis, and a dose >39 Gy was associated with longer duration of mucositis.[22] Delivery of the smallest dose possible to the oral cavity has been suggested to avoid radiation-induced mucositis.[23] As the results show in this work, using any of these dose levels as an evaluation criteria, our LP IMRT approach is the most promising to prevent radiation-induced mucositis. Clinical comparisons of IMRT and VMAT treatment plans for HNC have reported either similar or better results with VMAT than IMRT.[32333435] Some of these reports did not include oral cavity sparing as the main comparative objective.[3235] The limitation in the respective IMRT plans was that fewer beams and apertures utilized (e.g., 39 apertures over 5–7 beams by Bertelsen et al.[34]) may constrain performance in these plans. Most of the comparisons did not report the number of apertures used. Moreover, to deliver radiation continuously as the gantry rotates around the patient, VMAT requires that successive beam apertures be linked. The link between successive apertures is primarily dictated by the physical speed of the multileaf collimator leaves. As a result, the selection of beam apertures in VMAT is constrained by the need to link apertures to their neighbors. This constraint may adversely influence plan quality, which may be the reason that VMAT could not spare the oral cavity better than IMRT plans in this study. On the other hand, the superior results achieved by our LP IMRT approach were contributed from both accurate dose calculations and exact optimization algorithm. Most commercial planning systems tend to employ fast, but somewhat inaccurate dose calculation algorithms that provide input to the optimization algorithm. The achieved optimal solution differs from the dose distribution obtained when one performs a more accurate dose calculation. Al-Hallaq et al. demonstrated that more accurate dose calculation algorithms, such as convolution/superposition or Monte Carlo approaches, are better suited for IMRT in tailoring doses in HNC cases.[36] The dose calculation algorithm utilized in this research is Monte Carlo kernel superposition.[30] The other factor that may influence results is the optimization approach employed, as suggested by Bortfeld and Webb.[37] Most commercial planning systems employ heuristic or local search-based optimization approaches. Exact optimization approaches such as LP guarantee optimal solutions and provide explicit handling of constraints (soft and hard) over target and OAR voxels. Other exact optimization approaches, such as quadratic programming, should also provide similar results. On another note, IMRT and VMAT explore the degrees of freedom in generating conformal plans differently (aperture space vs. angular space, respectively). Our results show that for reducing the risk of oral mucositis in locally advanced HNC, LP IMRT becomes a more promising approach than VMAT. This result is observed despite the decrease in the number of beam apertures employed in the IMRT plans compared with 2ARC VMAT plans. Our work further suggests that when one considers comparisons between IMRT and VMAT using commercial software, on average, it is possible to obtain comparable plans. However, the LP IMRT approach produces substantially improved results. For locally advanced HNC, many structures in addition to the oral cavity are at risk for complications. We included these as nontarget/non-OAR normal tissues in this research because of concerns about reducing the risk of secondary cancer and complications for those structures. Hall[38] and Suit et al.[39] have pointed out that consideration in any radiation technique is the potential for increasing the number of radiation-induced secondary cancers. They noted that the volume of normal tissue exposed to leakage or low doses is directly related to secondary cancers. Guckenberger et al. have shown that using multiple arcs increases the spread of low doses to normal tissue.[40] Our results for normal tissue doses are consistent with this finding. In this work, normal tissue dose was explicitly considered in the constraints. This led to the finding that LP-based fixed-field IMRT plans resulted in lower normal tissue doses compared with VMAT plans for all 10 test cases considered. Our approach is analogous to segment weight optimization, which has been widely used in clinical practice. The limitation of our approach in the current form is that the starting point of the second phase is based on apertures generated from conventional IMRT plans. The quality of our approach could be further improved if a larger pool of high-quality apertures could be considered. This is the goal of next steps in our research. The entire framework as described here can be easily implemented and adapted to practical clinic use.

CONCLUSIONS

In addition to clinical evaluations of commercially available planning software, we applied a different optimization approach for fixed-field IMRT: LP-based IMRT. Our results showed that LP-based fixed-field IMRT plans resulted in superior dosimetric quality compared with conventional IMRT and VMAT plans. VMAT plans, however, consistently resulted in the lowest treatment delivery times. LP-based IMRT was the most promising treatment plan, for patients with locally advanced HNC, that preserves oral cavity and parotid glands so that the risk of oral mucositis can be reduced.

Financial support and sponsorship

This work was financially supported in part by a grant NIH/NCI CA130814.

Conflicts of interest

There are no conflicts of interest.
  34 in total

1.  Convolution/superposition using the Monte Carlo method.

Authors:  Shahid A Naqvi; Matthew A Earl; David M Shepard
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

2.  Single-Arc IMRT?

Authors:  Thomas Bortfeld; Steve Webb
Journal:  Phys Med Biol       Date:  2008-12-10       Impact factor: 3.609

3.  RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemoradiotherapy.

Authors:  Patricia Doornaert; Wilko F A R Verbakel; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-04-24       Impact factor: 7.038

4.  Intensity-modulated radiotherapy in patients with head and neck cancer: a European single-centre experience.

Authors:  D Van Gestel; D Van Den Weyngaert; D Schrijvers; J Weyler; J B Vermorken
Journal:  Br J Radiol       Date:  2011-04       Impact factor: 3.039

Review 5.  Oral mucositis.

Authors:  Rajesh V Lalla; Douglas E Peterson
Journal:  Dent Clin North Am       Date:  2005-01

Review 6.  Radiation therapy for head and neck cancers.

Authors:  Indranil Mallick; John N Waldron
Journal:  Semin Oncol Nurs       Date:  2009-08       Impact factor: 2.315

Review 7.  Mucositis incidence, severity and associated outcomes in patients with head and neck cancer receiving radiotherapy with or without chemotherapy: a systematic literature review.

Authors:  Andy Trotti; Lisa A Bellm; Joel B Epstein; Diana Frame; Henry J Fuchs; Clement K Gwede; Eugene Komaroff; Luba Nalysnyk; Marya D Zilberberg
Journal:  Radiother Oncol       Date:  2003-03       Impact factor: 6.280

8.  Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study.

Authors:  Wilko F A R Verbakel; Johan P Cuijpers; Daan Hoffmans; Michael Bieker; Ben J Slotman; Suresh Senan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-01       Impact factor: 7.038

Review 9.  Consequences of mucositis-induced treatment breaks and dose reductions on head and neck cancer treatment outcomes.

Authors:  David I Rosenthal
Journal:  J Support Oncol       Date:  2007-10

10.  Mucositis: current management and investigations.

Authors:  Adam S Garden
Journal:  Semin Radiat Oncol       Date:  2003-07       Impact factor: 5.934

View more
  3 in total

1.  Mouthwash Containing Vitamin E, Triamcinolon, and Hyaluronic Acid Compared to Triamcinolone Mouthwash Alone in Patients With Radiotherapy-Induced Oral Mucositis: Randomized Clinical Trial.

Authors:  Farzaneh Agha-Hosseini; Mona Pourpasha; Massoud Amanlou; Mahdieh-Sadat Moosavi
Journal:  Front Oncol       Date:  2021-03-11       Impact factor: 6.244

Review 2.  Precision medicine for risk prediction of oral complications of cancer therapy-The example of oral mucositis in patients receiving radiation therapy for cancers of the head and neck.

Authors:  Stephen T Sonis
Journal:  Front Oral Health       Date:  2022-08-18

3.  Local tissue electrical parameters predict oral mucositis in HNSCC patients: A diagnostic accuracy double-blind, randomized controlled trial.

Authors:  Gabriela Luize Guimarães Sanches; Agna Soares da Silva Menezes; Laércio Ives Santos; Cristina Paixão Durães; Larissa Lopes Fonseca; Marcelo Perim Baldo; Thais de Oliveira Faria; Luciano Alves de Araújo Andrade; Petr Iakovlevitch Ekel; Sérgio Henrique Sousa Santos; Alfredo Maurício Batista de Paula; Lucyana Conceição Farias; Marcos Flávio Silveira Vasconcelos D'Angelo; André Luiz Sena Guimarães
Journal:  Sci Rep       Date:  2020-06-12       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.