| Literature DB >> 29547536 |
Saverio Affatato1, Alessandro Ruggiero2, Sami Abdel Jaber3, Massimiliano Merola4, Pierangiola Bracco5.
Abstract
Given the long-term problem of polyethylene wear, medical interest in the new improved cross-linked polyethylene (XLPE), with or without the adding of vitamin E, has risen. The main aim of this study is to gain further insights into the mutual effects of radiation cross-linking and addition of vitamin E on the wear performance of ultra-high-molecular-weight polyethylene (UHMWPE). We tested four different batches of polyethylene (namely, a standard one, a vitamin E-stabilized, and two cross-linked) in a hip joint simulator for five million cycles where bovine calf serum was used as lubricant. The acetabular cups were then analyzed using a confocal profilometer to characterize the surface topography. Moreover; the cups were analyzed by using Fourier Transformed Infrared Spectroscopy and Differential Scanning Calorimetry in order to assess the chemical characteristics of the pristine materials. Comparing the different cups' configuration, mass loss was found to be higher for standard polyethylene than for the other combinations. Mass loss negatively correlated to the cross-link density of the polyethylenes. None of the tested formulations showed evidence of oxidative degradation. We found no correlation between roughness parameters and wear. Furthermore, we found significantly differences in the wear behavior of all the acetabular cups. XLPEs exhibited lower weight loss, which has potential for reduced wear and decreased osteolysis. However, surface topography revealed smoother surfaces of the standard and vitamin E stabilized polyethylene than on the cross-linked samples. This observation suggests incipient crack generations on the rough and scratched surfaces of the cross-linked polyethylene liners.Entities:
Keywords: FTIR analysis; cross-linked PE; hip simulator; standard PE; vitamin-E stabilized PE
Year: 2018 PMID: 29547536 PMCID: PMC5873012 DOI: 10.3390/ma11030433
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Schematic representation of the confocal apparatus for the topography acquisitions.
Figure 2Wear behavior for the different configurations of polyethylene tested.
Cumulative mass loss (mean ± standard deviation) for the four sets of polyethylene (PE) acetabular cups tested. Values and statistical analysis performed using a Kruskall–Wallis nonparametric test.
| Cycles (Mc) | Mean ± Standard Deviation (mg) | K–W Test | Post Hoc Test ( | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| STD | VE | XL-50 | XL-75 | ( | STD vs. VE | STD vs. XL-50 | STD vs. XL-75 | VE vs. XL-50 | VE vs. XL-75 | XL-50 vs. XL-75 | |
| 0.5 | 11.4 ± 4.8 | 10.5 ± 1.2 | 2.7 ± 2.1 | 1.7 ± 0.7 | 0.037 | 0.821 | 0.079 | 0.036 | 0.047 | 0.020 | 0.734 |
| 1.0 | 22.9 ± 3.4 | 23.3 ± 3.8 | 9.6 ± 2.8 | 7.5 ± 1.6 | 0.035 | 0.910 | 0.089 | 0.024 | 0.070 | 0.017 | 0.571 |
| 1.5 | 27.6 ± 5.2 | 28.1 ± 3.9 | 13.0 ± 3.4 | 10.9 ± 2.3 | 0.038 | 0.910 | 0.070 | 0.031 | 0.054 | 0.024 | 0.734 |
| 2.0 | 40.9 ± 7.3 | 39.0 ± 5.5 | 17.4 ± 2.7 | 13.9 ± 2.4 | 0.030 | 0.910 | 0.089 | 0.013 | 0.113 | 0.017 | 0.428 |
| 2.5 | 54.1 ± 5.0 | 60.13 ± 8.8 | 22.5 ± 4.3 | 17.2 ± 2.4 | 0.026 | 0.571 | 0.174 | 0.031 | 0.054 | 0.007 | 0.428 |
| 3.0 | 67.0 ± 10.5 | 65.2 ± 9.1 | 25.7 ± 4.6 | 21.0 ± 1.9 | 0.035 | 0.910 | 0.070 | 0.017 | 0.089 | 0.024 | 0.571 |
| 3.5 | 80.8 ± 12.2 | 73.7 ± 14.2 | 31.2 ± 4.5 | 24.3 ± 2.6 | 0.030 | 0.910 | 0.089 | 0.013 | 0.113 | 0.017 | 0.428 |
| 4.0 | 89.4 ± 11.3 | 82.4 ± 13.3 | 34.6 ± 4.9 | 26.6 ± 2.8 | 0.030 | 0.910 | 0.089 | 0.013 | 0.113 | 0.017 | 0.428 |
| 4.5 | 94.8 ± 10.0 | 88.1 ± 13.6 | 37.3 ± 5.1 | 29.3 ± 2.7 | 0.030 | 0.910 | 0.089 | 0.013 | 0.113 | 0.017 | 0.428 |
| 5.0 | 102.6 ± 15.1 | 94.9 ± 16.7 | 39.3 ± 6.7 | 31.8 ± 2.6 | 0.032 | 0.910 | 0.079 | 0.015 | 0.100 | 0.020 | 0.496 |
Statistical significance: p < 0.05.
Figure 3Contour images of the topographies acquired on the inner surfaces of four non-loaded specimens: (a) VE; (b) STD; (c) XL-75; (d) XL-50.
Figure 4First set of contour images of the topographies acquired on the worn inner surfaces of four loaded specimens: (a) VE; (b) STD; (c) XL-75; (d) XL-50.
Figure 5Roughness histograms of the worn surfaces on the mentioned cups. The parameters are Ra, Rq, Rz, and Rt.
Physical–chemical characteristics of the control samples, resulting from FTIR, DSC, and cross-link density measurements.
| Polyethylene | Crystallinity (%) | Trans-Vinylene (mmol/L) | Cross-Link Density (mol/dm3) |
|---|---|---|---|
| STD | 50.3 ± 0.9 | - | - |
| VE | 51.2 ± 1.2 | - | - |
| XL-50 | 40.8 ± 1.0 | 5.2 | 0.132 ± 0.009 |
| XL-75 | 35.5 ± 0.9 | 5.8 | 0.139 ± 0.010 |
Figure 6FTIR spectra of the investigated materials.
Figure 7Mechanism of cross-linking of ultra-high-molecular-weight polyethylene (UHMWPE) through consumption of vinyl double bonds.