Yuanhan Chang1, Abhijit Anil Tambe2, Yoshinobu Maeda3, Masahiro Wada3, Tomoya Gonda3.
Abstract
A literature review of finite element analysis (FEA) studies of dental implants with their model validation process was performed to establish the criteria for evaluating validation methods with respect to their similarity to biological behavior. An electronic literature search of PubMed was conducted up to January 2017 using the Medical Subject Headings "dental implants" and "finite element analysis." After accessing the full texts, the context of each article was searched using the words "valid" and "validation" and articles in which these words appeared were read to determine whether they met the inclusion criteria for the review. Of 601 articles published from 1997 to 2016, 48 that met the eligibility criteria were selected. The articles were categorized according to their validation method as follows: in vivo experiments in humans (n = 1) and other animals (n = 3), model experiments (n = 32), others' clinical data and past literature (n = 9), and other software (n = 2). Validation techniques with a high level of sufficiency and efficiency are still rare in FEA studies of dental implants. High-level validation, especially using in vivo experiments tied to an accurate finite element method, needs to become an established part of FEA studies. The recognition of a validation process should be considered when judging the practicality of an FEA study.Entities:
Keywords: Dental implant; Finite element analysis; Validation; Verification
Year: 2018 PMID: 29516219 PMCID: PMC5842167 DOI: 10.1186/s40729-018-0119-5
Source DB: PubMed Journal: Int J Implant Dent ISSN: 2198-4034
Fig. 1Flowchart of literature review. An electronic literature search of PubMed was conducted up to January 2017 using the Medical Subject Headings “dental implants” and “finite element analysis.” After accessing the full texts, the context of each article was searched using the words “valid” and “validation” and articles in which these words appeared were read to determine whether they met the inclusion criteria for the review
All studies in the literature that considered with an actual validation of FEA
| Ranking | Authors | Year | FE model | FEM geometry reference | Material properties of tissues around implants | Validation | Material of validation model | Comparison items |
|---|---|---|---|---|---|---|---|---|
| A | Heckmann et al. [ | 2006 | Implants embedded in a bone block | CAD: bone block with a cortical layer and cancellous layer was constructed by CAD | Homogeneous isotropic linear elasticity: cortical and trabecular bone | (1) In vitro testing: strain gauge on implant support bridge in resin bone model | (1) Epoxy resin | Surface strain of resin/resin |
| B | Hou et al. [ | 2009 | Implants embedded in bone (rat’s mouth) | CT: CT data of the implant in a rat model | Not mentioned | In vivo experiment: implants placed in rat, and histologic findings compared after loading | In vivo: rat | Histologic findings |
| B | Natali et al. [ | 1997 | Implants embedded in a bone block | CAD: bone section constructed by CAD | Homogeneous isotropic linear elasticity: cortical and trabecular bone | In vivo experiment: implant insertion in animal (dog) for loading and creation of sections of bone and implants | In vivo: dog | Visualization of change in bone and stress analysis by FEA |
| B | Cha et al. [ | 2015 | Implant engaged in bone disc (model was used to calculate torque) | CAD: based on histology of the bone–implant interface | Homogeneous isotropic linear elasticity | In vivo experiment: implant insertion in animal (mice) with different insertion torques | In vivo: mice | Histomorphometric analyses |
| C | Nagasao et al. [ | 2009 | Implants embedded in normal mandibles and reconstructed mandible (with fibulae or ribs) and under mastication movement (dynamic condition) | CT: dry mandibles, ribs, and fibulae | Homogeneous isotropic linear elasticity: cortical and cancellous bone of every part of mandible, fibula, and rib | Mechanical testing: implant embedded in 3 full mandibles and surface strain under loading measured by strain gauge | Dry mandible as mandible | Surface strain of bone under same conditions in FEA/experiment |
| C | Nagasao et al. [ | 2010 | Implants embedded in normal mandibles and reconstructed mandible (with fibulae or ribs) under mastication movement (dynamic condition) | CT: dry mandibles, ribs, and fibulae | Homogeneous isotropic linear elasticity: cortical and cancellous bone of every part of mandible, fibula, and rib | Mechanical testing: implant embedded in 2 full mandibles and surface strain under loading measured by strain gauge | Dry mandible as mandible | Surface strain of bone under same conditions in FEA/experiment |
| C | Eser et al. [ | 2009 | Four implants embedded in the maxilla with bar superstructure | CAD: model of nonanatomic maxilla, individualized arch form according to implant alignment | Homogeneous isotropic linear elasticity: cortical bone, cancellous bone, Ti, Alloy, bar-superstructure | Ex vivo strain gauge measurement of cadaver’s maxilla (with implants) | Cadaver | Surface strain of bone (maxilla) |
| C | Nagasao et al. [ | 2006 | Implants embedded in normal maxilla and cleft maxilla | CT and CAD: normal maxilla: CT from a dry skull; palatal cleft, alveolar cleft, and complete cleft were designed by computer | Homogeneous isotropic linear elasticity: cortical and trabecular bone | Mechanical testing: strain measurement by strain gauge and implant embedded in actual skull model | Dry skull | Surface strain of bone |
| D | Bardyn et al. [ | 2010 | Implants embedded in bone (polyurethane foam and sheep bone) | CT: polyurethane foam block and sheep bone | Nonhomogeneous: calculated from CT data | Mechanical testing in both polyurethane foam and sheep bone: measurement of removal torque of the implant | Polyurethane foam and sheep bone | Removal torque of implants |
| D | Olsen et al. [ | 2005 | Implants embedded in porcine mandibles from CT data and application of loading on the implant of FEM | CT: porcine mandibles | Nonhomogeneous: calculated from CT data | Mechanical testing: comparison of displacement with actual measurements under the same testing load | Block of porcine mandible | Implant displacement under loading |
| D | Huang et al. [ | 2002 | Implant embedded in bone block | CAD: bone block model constructed by CAD | Homogeneous isotropic linear elasticity: cortical and trabecular bone | In vitro model testing experiments: implant in bone cubic and measurement of resonance frequencies | Bone section from lumbar vertebrae of hogs | Value of resonance frequency |
| D | Hasan et al. [ | 2012 | Implant (implant and abutment together) embedded in bovine bone | CT: scan of the models used for the experiment (implant embedded in bovine rib) | Homogeneous isotropic linear elasticity: bovine cortical bone, bovine cancellous bone | Mechanical tests: implant displacement and rotation under loading were measured using a biomechanical measurement system (laser pinhole and camera) | Bovine rib section as mandible bone | Displacement of the abutment |
| D | Chatzigianni et al. [ | 2011 | Mini-implant embedded in bone | CT: scan of the specimen used for the experiment (implant embedded in bovine rib) | Homogeneous isotropic linear elasticity: bovine cortical bone, bovine cancellous bone | Mechanical tests: implant displacement and rotation under loading were measured using a 3D mobility measurement system (laser beams and camera) | Bovine rib section as mandible bone | Displacement of the abutment |
| E1 | Tiossi et al. [ | 2013 | Implants and tooth (acrylic) embedded in resin block model, crowns (splint and non-splint) | CAD: epoxy model block | Nil (in this FEM, there was a resin block only and no living tissue simulation) | Digital image correlation (DIC): images of deforming body captured and strain calculated. Mechanical testing with implants embedded in resin block | Resin block as mandible bone | Calculated surface strain by DIC and FEA |
| E2 | Ozçelik et al. [ | 2007 | Three-unit bridge fixed prosthesis (with rigid connector and non-rigid) supported by an implant and a natural tooth, with an adjacent tooth and surrounding bone | CAD: a bone section (2D) was constructed by CAD with a cortical layer and spongious bone and PDL | Homogeneous isotropic linear elasticity: enamel, dentin, pulp, cortical bone, cancellous bone | Photoelastic stress analysis methods (PSAM): implants placed in photoelastic resin, then force loaded and photograph taken | Photoelastic resin as bone | Stress distribution in bone/resin |
| E3 | Chou et al. [ | 2014 | A section of mandible and implant | CT data and 2D FE model used in previous study | Homogeneous isotropic linear elasticity | Mechanical testing: implant embedded in resin bone and strain measured by strain gauge | 3D printer to build acrylic-based polymer | Surface strain of bone |
| E3 | Mobilio et al. [ | 2013 | Implant embedded in a bone block | CAD: bone block built by CAD with a cortical (1.5 mm) and trabecular (28.5 mm) layer | Homogenous anisotropic linear elasticity cortical bone: orthotopic linearly elastic material; trabecular bone: transversely isotropic linearly elastic material | Mechanical testing: implant embedded in resin block and strain measured by strain gauge | Resin block as mandible bone | Load and strain relationship |
| E3 | Chang et al. [ | 2012 | Short implants with crowns embedded in left posterior segment of maxilla | CT: CT scan of a dry human male skull | Homogeneous isotropic linear elasticity: cortical bone, cancellous bone (high and low density) | Mechanical testing: strain measured by strain gauge and implant embedded in resin block under loading | ABS plastic bone as maxillary bone | Surface strain of bone/resin |
| E3 | Tu et al. [ | 2010 | Implant embedded in resin block | CAD: a resin block with a cortical layer and cancellous layer was constructed by CAD | Nil (in this FEM, there was a resin block only and no living tissue simulation) | Mechanical testing: strain measured by strain gauge and implant embedded in resin mandible section | Resin bone as mandible bone | Surface strain of bone/resin |
| E3 | Lin et al. [ | 2010 | Implant embedded in the left maxilla with crown | CT of intact healthy male patient | Homogeneous isotropic linear elasticity: cortical bone, cancellous bone | Mechanical testing: strain measured by strain gauge and implant embedded in resin mandible section | ABS plastic bone as maxillary bone | Surface strain of bone/resin |
| E3 | Qian et al. [ | 2009 | Implant embedded in bone block | CAD: a bone cubic with cortical layer and cancellous layer was constructed by CAD | Homogeneous isotropic linear elasticity: cortical bone, cancellous bone | (1) In vitro experiment: mechanical testing with resin bone and digital image correlation to calculate displacement of implant and strain on bone | Resin block as mandible bone | (1) Displacement of implant and strain on bone |
| E3 | Karl et al. [ | 2009 | Implant embedded in base made by 3 materials | CAD: acrylic, G10 epoxy resin, aluminum | Homogeneous isotropic linear elasticity: acrylic, G10 epoxy resin, aluminum | Mechanical testing: strain gauge. FEA-calculated strain was compared with strain gauge results | Acrylic resin, glass-filled epoxy, aluminum | Surface strain of acrylic resin, glass-filled epoxy, aluminum |
| E3 | Hsu et al. [ | 2009 | Implant embedded in resin block (with resin’s parameter for consistence with experiment) | Nil | Homogeneous isotropic linear elasticity: Resin (epoxy and Tempron) | Mechanical testing: implant embedded in resin bone section and surface strain under loading was measured by strain gauge | Resin block as mandible bone | Surface strain of resin/resin |
| E3 | Nagasawa et al. [ | 2008 | Implant embedded in a bone block (only compact bone) | CAD: a bone block (compact bone) was constructed by CAD | Homogeneous isotropic linear elasticity: compact bone | Mechanical loading test for implant, sectioned longitudinally | Nil | Implant deformation; no scientific values |
| E3 | Huang et al. [ | 2005 | Splinted or non-splinted 2-unit crowns supported by 2 or 3 implants embedded in bone | CT: CT of posterior portion of a cadaver mandible | Homogeneous anisotropic linear elasticity cortical bone: orthotopic linearly elastic material; trabecular bone: transversely isotropic linearly elastic material | Mechanical test: strain measured by strain gauge on model | Acrylic resin as mandible bone | Surface strain of resin/bone |
| E3 | Iplikçioğlu et al. [ | 2003 | Implant embedded in bone block | CAD: a resin block model was constructed by CAD | Nil (in this FEM, there was a resin model only and no living tissue simulation) | Mechanical test: measurement of stress on the implant, abutment, and resin | Resin block as bone | Stress distribution in resin and implants |
| E3 | Chang et al. [ | 2016 | Ball attachment overdenture (mandible, implant and attachment, mucosa, denture) | CT: from a single human mandible (edentulous 65-year-old woman) | Homogeneous isotropic linear elasticity | Mechanical test: strain measured by strain gauge on surface of bone model | Rapid prototype ABS plastic bone model, and a 3-mm layer of silicone to simulate mucosa | Surface strain of resin/bone |
| E3 | Rezende et al. [ | 2015 | Bone section with embedded implant and prosthesis (metal coping and porcelain), screws | CT: in vitro model (resin bone) | Homogeneous isotropic linear elasticity | Mechanical test: strain measured by strain gauge on surface of bone model | Polyurethane resin | Surface strain of resin/bone |
| E3 | Chang et al. [ | 2012 | Implants embedded in maxilla section with imperfect and perfect osseointegration under force loading | CT: data of maxillary first molar area | Inhomogeneous anisotropic linear elasticity cortical bone: anisotropic Trabecular bone: transversely isotropic linearly elastic material | Mechanical testing: strain measured by strain gauge and implant embedded in resin block | Resin block as maxillary bone | Surface strain of bone/resin block |
| E3 | Chang et al. [ | 2012 | Implants and crowns in a section of the maxilla | CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD | Homogeneous anisotropic linear elasticity compact bone, cancellous bone | Mechanical testing: strain measured by strain gauge and implant embedded in resin block | ABS resin block as mandible bone | Surface strain of bone/resin block |
| E4 | Zhiyong et al. [ | 2004 | (1) Single tooth in bone block | CAD: a bone block model was constructed by CAD | Homogeneous isotropic linear elasticity: cortical and trabecular bone, dentin, PDL | Mechanical testing: comparison of displacement with actual measurements under the same tested loading conditions | Not mentioned | Implant displacement under loading |
| E4 | Chang et al. [ | 2012 | Implant embedded in a bone block | CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD | Homogeneous isotropic linear elasticity: compact bone, cancellous bone | Mechanical testing: pullout testing of mini-implant inserted in synthetic bone material | Synthetic bone material as mandible bone | Pullout strength of mini-implant |
| E5 | Inglam et al. [ | 2013 | Implant embedded in a bone block | CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD | Homogeneous anisotropic linear elasticity cortical bone: orthotopic isotropic | Mechanical testing: strain measured by strain gauge and implant embedded in resin block | Resin block as mandible bone | Surface strain of bone/resin block |
| E5 | Necchi et al. [ | 2003 | Implant (fixture, abutment, and connecting screw) | Nil | Nil | Mechanical failure tests: preloading and functional loading conditions | Not mentioned | Maximum breaking force |
| E5 | Genna et al. [ | 2003 | Implant embedded in bone block | CAD: a resin block model was constructed by CAD | Nil (in this FEM, there was a resin model only and no living tissue simulation) | Cyclic mechanical fatigue testing: implant placed in epoxy resin and section of specimen examined under microscope | Epoxy resin block as bone | Comparison of locations of stress focus |
| E5 | Perriard et al. [ | 2002 | Different types of implant bodies and abutments embedded in resin | CAD: epoxy resin as bone | Nil (in this FEM, there was a resin model only and no living tissue simulation) | Mechanical fatigue testing of implant model: until half of samples still survived under loading | Resin block | Comparison of locations of stress concentrations |
| F1 | Bruno Salles Sotto-Maior et al. [ | 2016 | A bone model of mandibular right posterior region | CT: from a patient’s mandible | Homogeneous isotropic linear elasticity: cortical and trabecular bone | Clinical findings of bone loss at 1-year follow-up | In vivo: radiographic films of patients | Mechanoregulatory tissue model was employed to monitor the morphological changes in bone subjected to biomechanical loading |
| F1 | Wang et al. [ | 2013 | A 3D model of maxillary bone | CT image of maxillary bone section missing both central incisors | Homogeneous isotropic linear elasticity | Radiographs qualitatively compared regarding resemblance between computational remodeling results and clinical data | In vivo: radiographic films of patients | Comparison of variations in maxillary bone densities |
| F1 | Choi et al. [ | 2012 | Implants embedded in anterior maxilla | CT: anterior maxillary bone | Homogeneous isotropic linear elasticity compact bone, cancellous bone | Comparison of model implant displacement under the same load with clinical outcomes in literature | Literature | Model implant displacement |
| F1 | Shen et al. [ | 2010 | Implant embedded in mandibular right first molar area | CAD: a bone block with a cortical layer and cancellous layer was constructed by CAD | Homogeneous anisotropic linear elasticity: cortical bone, cancellous bone | Clinical data; comparison of implant displacement value under 20-N loading from clinical data | Clinical results | Implant displacement under loading |
| F1 | Lin et al. [ | 2010 | Implant embedded in mandible (cortical and cancellous bone), crown, teeth | CT: in vivo CT of a segment of mandible | Inhomogeneous anisotropic linear elasticity: cortical bone, cancellous bone (properties varied with density) | Clinical data: comparison of bone density with other clinical follow-up X-ray images | X-ray images of human | X-ray images |
| F2 | MacGinnis et al. [ | 2014 | 3D skull model with masked sutures | CT: from 42-year-old man, 3D skull image excluding the mandible | Homogeneous isotropic linear elasticity | Comparison with past literature | Comparison with conclusions of past literature | |
| F2 | Fanuscu et al. [ | 2004 | Unilateral edentulous posterior maxilla with grafted sinus | CAD: unilateral edentulous posterior maxilla with grafted sinus was constructed by computer | Homogeneous isotropic linear elasticity: cortical and trabecular bone | Validation with previous study by one of the authors in which photoelastic modeling with similar geometry was used | Location of stress | |
| F2 | Mellal et al. [ | 2004 | Cylindrical implant, bone consisting of a cancellous core coated with cortical envelope | CAD: a bone section model was constructed by CAD | Homogeneous isotropic linear elasticity: cortical and trabecular bone | Literature: systematic search of the literature was conducted to relate the numerical predictions to existing in vivo data | ||
| F2 | Zarone et al. [ | 2003 | Mandible with 6 implants and prosthetic superstructure | Laser: a man’s total mandible by laser digitizer | Homogeneous isotropic linear elasticity: cortical and trabecular bone | Data from previous experiments: comparisons of range of medial convergence during opening and protrusive movements | ||
| G | Bulaqi et al. [ | 2015 | Implants embedded in a bone block | CT data: mandible | Homogeneous isotropic linear elasticity | Comparison with theoretically predicted values (calculated with the equations) | values of conical to wretch torque ratio | |
| G | Vayron et al. [ | 2015 | Implants embedded in a bone block | CAD: cortical bone, newly formed bone, and trabecular bone | Homogeneous isotropic mechanical properties | Comparison with results using a 2D finite difference numerical model |
Fig. 2Hierarchy of validations based on their similarity to real biomechanical behaviors. The articles (n = 47) were categorized according to their validation method as follows: in vivo experiments in humans (n = 1) and other animals (n = 3), model experiments (n = 32), others’ clinical data and past literature (n = 9), and other software (n = 2)
Fig. 3Proportion of dental implant FEA articles with a validation. (Left) Among totally 522 FEA articles of dental implants which we were able to access English full text up to January 2017, there are only 47 articles with a validation. (Right) The articles with a validation were categorized according to their validation method as follows levels: A, in vivo (human bodies); B, performed in vivo (heterogeneous animals); C, model experiment performed using part of a cadaver; D, model experiment performed using heterogeneous bone; E, model experiment performed using artificial materials; F, comparison with past literature; and G, performed with other software (n = 2)