Literature DB >> 36267532

Stress analysis and factor of safety in three dental implant systems by finite element analysis.

E Menacho-Mendoza1, R Cedamanos-Cuenca2, A Díaz-Suyo1.   

Abstract

Objective: The purpose of this study was to compare the stress distribution and the factor of safety of three dental implant systems using the finite element method. Materials and methods: Three commercial dental implant systems were designed using Solid Works 2020 software: Model A with an internal octagonal connection and matching platform, Model B with an internal hexagon connection and switching platform, and Model C with an internal 15° conical-cylindrical connection and switching platform. A 200 N load was applied to each design in both axial and 30° oblique directions using the finite element method.
Results: In the three dental implant systems, the maximum von Mises stress was concentrated at the cervical level of the bone-implant interface in all models. Model C showed lower maximum stress values in both axial and 30° oblique loads. The highest maximum stress value was observed with the application of the oblique load in all the study models, and the factor of safety was less than one in Model A when subjected to a 200 N oblique load.
Conclusion: The switching platform models generated lower maximum stress values and a factor of safety higher than one which is considered an acceptable value.Clinical relevance: A dental implant system with an internal hexagon or conical connection and a switching platform generates lower maximum von Mises stress values both on the implant components and on the peri-implant tissues.
© 2022 The Authors. Production and hosting by Elsevier B.V. on behalf of King Saud University.

Entities:  

Keywords:  Dental implants; Factor of safety; Finite element analysis; Stress distribution; von Mises stress

Year:  2022        PMID: 36267532      PMCID: PMC9577351          DOI: 10.1016/j.sdentj.2022.08.006

Source DB:  PubMed          Journal:  Saudi Dent J        ISSN: 1013-9052


  18 in total

1.  Stress analysis in platform-switching implants: a 3-dimensional finite element study.

Authors:  Eduardo Piza Pellizzer; Fellippo Ramos Verri; Rosse Mary Falcón-Antenucci; Joel Ferreira Santiago Júnior; Paulo Sérgio Perri de Carvalho; Sandra Lúcia Dantas de Moraes; Pedro Yoshito Noritomi
Journal:  J Oral Implantol       Date:  2010-10-08       Impact factor: 1.779

Review 2.  Performance of conical abutment (Morse Taper) connection implants: a systematic review.

Authors:  Christian M Schmitt; Getulio Nogueira-Filho; Howard C Tenenbaum; Jim Yuan Lai; Carlos Brito; Hendrik Döring; Jörg Nonhoff
Journal:  J Biomed Mater Res A       Date:  2013-05-09       Impact factor: 4.396

3.  Scientific use of the finite element method in Orthodontics.

Authors:  Luegya Knop; Luiz Gonzaga Gandini; Ricardo Lima Shintcovsk; Marcia Regina Elisa Aparecida Schiavon Gandini
Journal:  Dental Press J Orthod       Date:  2015 Mar-Apr

4.  Three-Dimensional Finite Element Analysis of the Stress Distribution at the Internal Implant-Abutment Connection.

Authors:  Sung-Yong Cho; Yoon-Hyuk Huh; Chan-Jin Park; Lee-Ra Cho
Journal:  Int J Periodontics Restorative Dent       Date:  2016 May-Jun       Impact factor: 1.840

5.  Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: an in vitro study.

Authors:  Abílio Ricciardi Coppedê; Edmilson Bersani; Maria da Gloria Chiarello de Mattos; Renata Cristina Silveira Rodrigues; Ivete Aparecida de Mattias Sartori; Ricardo Faria Ribeiro
Journal:  Int J Prosthodont       Date:  2009 May-Jun       Impact factor: 1.681

Review 6.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08

7.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

8.  Effects of Implant-Abutment Connection Type and Inter-Implant Distance on Inter-Implant Bone Stress and Microgap: Three-Dimensional Finite Element Analysis.

Authors:  Takashi Matsuoka; Tamaki Nakano; Satoshi Yamaguchi; Shinji Ono; Shota Watanabe; Takumi Sato; Hirofumi Yatani
Journal:  Materials (Basel)       Date:  2021-05-06       Impact factor: 3.623

9.  Finite Element Method and Von Mises Investigation on Bone Response to Dynamic Stress with a Novel Conical Dental Implant Connection.

Authors:  Luca Fiorillo; Marco Cicciù; Cesare D'Amico; Rodolfo Mauceri; Giacomo Oteri; Gabriele Cervino
Journal:  Biomed Res Int       Date:  2020-10-07       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.