Biomechanical evaluation of implant placement in the reconstructed mandible.

Tomohisa Nagasao1, Junpei Miyamoto, Hiromasa Kawana.   

Abstract

PURPOSE: Following resection of malignant tumors from the mandible, immediate reconstruction of the discontinuous jaw is commonly performed. The present study aims to define the risks associated with dynamic loading of the reconstructed mandible.
MATERIALS AND METHODS: Computer-aided design simulations of eight mandibles were produced and termed normal models. The normal models were then modified by removing part of the right body and restoring the defects with bone from the rib or fibula. These modified models were termed reconstructed models. Thereafter, an implant was embedded in the first molar region of the left side for all models. Using finite element analysis, the stresses occurring at the implant-bone interface with simulated mastication were calculated. The Wilcoxon signed rank test was applied to compare the magnitudes of stresses in both models.
RESULTS: In terms of the stress exerted at the implant-bone interface during mastication, the normal models and the reconstructed models showed no significant differences (P > .05).
CONCLUSION: Placement of an implant on the nonreconstructed side following partial resection and mandibular reconstruction presents no significant risk.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20162103

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  5 in total

1.  3D Finite Element Study on: Bar Splinted Implants Supporting Partial Denture in the Reconstructed Mandible.

Authors:  Mohamed El-Anwar; Rami Ghali; Mona Aboelnagga
Journal:  Open Access Maced J Med Sci       Date:  2016-02-09

2.  Finite Element Simulation and Additive Manufacturing of Stiffness-Matched NiTi Fixation Hardware for Mandibular Reconstruction Surgery.

Authors:  Ahmadreza Jahadakbar; Narges Shayesteh Moghaddam; Amirhesam Amerinatanzi; David Dean; Haluk E Karaca; Mohammad Elahinia
Journal:  Bioengineering (Basel)       Date:  2016-12-19

3.  Fixation Release and the Bone Bandaid: A New Bone Fixation Device Paradigm.

Authors:  Narges Shayesteh Moghaddam; Ahmadreza Jahadakbar; Amirhesam Amerinatanzi; Roman Skoracki; Michael Miller; David Dean; Mohammad Elahinia
Journal:  Bioengineering (Basel)       Date:  2017-01-22

Review 4.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08

5.  Metallic Fixation of Mandibular Segmental Defects: Graft Immobilization and Orofacial Functional Maintenance.

Authors:  Narges Shayesteh Moghaddam; Ahmadreza Jahadakbar; Amirhesam Amerinatanzi; Mohammad Elahinia; Michael Miller; David Dean
Journal:  Plast Reconstr Surg Glob Open       Date:  2016-09-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.