Sean McCafferty1,2,3,4, Jason Levine5,6, Jim Schwiegerling5,7, Eniko T Enikov8.
Abstract
BACKGROUND: Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro.Entities:
Keywords: Applanation; Bias; CCT; CRF; Cadaver eye; Central corneal thickness; Corneal hydration; Corneal resistance factor; Error; Glaucoma; Goldmann; Head position; IOP; In vitro; In vivo; Intracameral; Intraocular pressure; Manometric; Perkins; Supine; Tonometer; Upright
Mesh:
Year: 2018 PMID: 29301514 PMCID: PMC5753488 DOI: 10.1186/s12886-017-0668-z
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Fig. 1CATS tonometer prism with modified applanating surface versus GAT
Fig. 2Ocular globe IOP apparatus for measuring IOP in the supine position, showing a Perkins type tonometer
Fig. 3Ocular globe IOP apparatus for measuring upright IOP showing Goldmann type tonometer
Fig. 4In Vivo Perkins IOP measurement scatterplot using the CATS and GAT prisms over all Intracameral IOPs In live human eyes undergoing cataract surgery
Fig. 5CATS and GAT IOP measurement difference from true intracameral IOP correlated to central corneal thickness (CCT)
Fig. 6CATS and GAT IOP measurement difference from true intracameral IOP correlated to corneal resistance factor (CRF)
Fig. 7Average CATS and GAT IOP measurement error from true intracameral pressure in thin corneas (CCT < 530 μm)
Fig. 8Average CATS and GAT IOP measurement error from true intracameral pressure in thick corneas (CCT > 570 μm)
Fig. 9CATS and GAT IOP measurements compared to an average true intracameral pressure of 11.3 mmHg
Fig. 10CATS and GAT IOP measurements compared to an average true intracameral pressure of 22.0 mmHg
Fig. 11CATS and GAT prism cadaver IOP measurement comparison to intracameral pressure in human cadaver eyes
IOP measurement variance and accuracy with the CATS and GAT prisms in cadaver eyes
| Overall accuracy/repeatability | Parameter | 5 mmHg | 10 mmHg | 20 mmHg | 30 mmHg | 40 mmHg | 50 mmHg | 60 mmHg |
|---|---|---|---|---|---|---|---|---|
| GAT reference prism | Mean IOP meas. | 1.1 | 7.6 | 17.7 | 27.8 | 38.0 | 47.0 | 54.1 |
| Std. Dev. | 1.2 | 3.6 | 1.5 | 2.2 | 1.1 | 3.6 | 2.6 | |
| Coeff. of Variation | 1.1 | 0.5 | 0.1 | 0.1 | 0.0 | 0.1 | 0.0 | |
| Accuracy (error to transducer IOP) | −4.0 | −2.5 | −2.3 | −2.3 | −2.6 | −2.5 | −5.5 | |
| CATS prism | Mean IOP meas. | 3.0 | 7.1 | 19.2 | 29.0 | 41.1 | 48.4 | 55.0 |
| Std. Dev. | 1.7 | 4.0 | 2.2 | 2.6 | 1.2 | 1.2 | 1.6 | |
| Coeff. of Variation | 0.6 | 0.6 | 0.1 | 0.1 | 0.0 | 0.0 | 0.0 | |
| Accuracy (error to transducer IOP) | −2.1 | −3.0 | −0.5 | −0.9 | 0.4 | −1.1 | −4.3 |