Literature DB >> 30195894

Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized Clinical Trial.

Sean McCafferty1, Kyle Tetrault2, Ann McColgin3, Warren Chue4, Jason Levine4, Melissa Muller4.   

Abstract

PURPOSE: To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety.
DESIGN: Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis.
METHODS: A GAT and a modified surface GAT prism measured IOP on 173 unique eyes. The study design and analysis complied with FDA 510(k) and ANSIZ80.10-2014 guidelines. All eyes were randomized to IOP measurement by 1 of 5 standard prisms or 5 modified prisms, each from a different manufacturing lot. Pressures were measured by 6 investigators, 2 times with each prism, for a total of 1384 IOP measurements. Analysis included Bland-Altman difference accuracy, intraoperator and interoperator IOP measurement, and manufactured lot repeatability.
RESULTS: Bland-Altman indicated no IOP measurements pairs outside the ±5 mm Hg guidelines. Operator and manufactured lot repeatability F tests and 1-way ANOVAs indicated no statistical difference between the standard and modified prisms (all P > .10). The difference in IOP measurements of the standard and modified prisms correlated well to Dresdner central corneal thickness (CCT) correction (P = .01).
CONCLUSION: A modified surface replacement prism is statistically equivalent to a flat-surfaced prism. The modified surface prism indicated statistically significant correction for CCT requiring further testing outside the ANSI standard limits (0.500 mm < CCT < 0.600 mm) to examine its full potential.
Copyright © 2018 The Author(s). Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30195894      PMCID: PMC6366945          DOI: 10.1016/j.ajo.2018.08.051

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  19 in total

Review 1.  Influence of corneal variables on accuracy of intraocular pressure measurement.

Authors:  Karim F Damji; Rajeev H Muni; Rejean M Munger
Journal:  J Glaucoma       Date:  2003-02       Impact factor: 2.503

2.  Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry.

Authors:  Markus Kohlhaas; Andreas G Boehm; Eberhard Spoerl; Antje Pürsten; Hans J Grein; Lutz E Pillunat
Journal:  Arch Ophthalmol       Date:  2006-04

Review 3.  Intraocular pressure assessment after laser in situ keratomileusis: a review.

Authors:  Andrew S H Tsai; Seng Chee Loon
Journal:  Clin Exp Ophthalmol       Date:  2011-09-02       Impact factor: 4.207

4.  Why Do People (Still) Go Blind from Glaucoma?

Authors:  Remo Susanna; Carlos Gustavo De Moraes; George A Cioffi; Robert Ritch
Journal:  Transl Vis Sci Technol       Date:  2015-03-09       Impact factor: 3.283

5.  Factors that influence intraocular pressure changes after myopic and hyperopic LASIK and photorefractive keratectomy: a large population study.

Authors:  Julie M Schallhorn; Steven C Schallhorn; Yvonne Ou
Journal:  Ophthalmology       Date:  2014-11-14       Impact factor: 12.079

6.  Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer.

Authors:  Aachal Kotecha; Ahmed Elsheikh; Cynthia R Roberts; Haogang Zhu; David F Garway-Heath
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-12       Impact factor: 4.799

7.  Comparison of Intraocular Pressure Measurements Obtained by Rebound, Noncontact, and Goldmann Applanation Tonometry in Children.

Authors:  Chi Shian Feng; Ki Won Jin; Kayoung Yi; Dong Gyu Choi
Journal:  Am J Ophthalmol       Date:  2015-07-23       Impact factor: 5.258

8.  Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo.

Authors:  D L Eisenberg; B G Sherman; C A McKeown; J S Schuman
Journal:  Ophthalmology       Date:  1998-07       Impact factor: 12.079

9.  Goldmann and error correcting tonometry prisms compared to intracameral pressure.

Authors:  Sean McCafferty; Jason Levine; Jim Schwiegerling; Eniko T Enikov
Journal:  BMC Ophthalmol       Date:  2018-01-04       Impact factor: 2.209

10.  Goldmann tonometer error correcting prism: clinical evaluation.

Authors:  Sean McCafferty; Garrett Lim; William Duncan; Eniko T Enikov; Jim Schwiegerling; Jason Levine; Corin Kew
Journal:  Clin Ophthalmol       Date:  2017-05-03
View more
  4 in total

1.  Improved Efficacy of Topical Latanoprost 0.005% Demonstrated by Corneal Biomechanical Correcting Modified Goldmann Prism.

Authors:  Nathan Radcliffe; John Berdahl; Mitchel Ibach; Justin Schweitzer; Jason Levine; Sean McCafferty
Journal:  Clin Ophthalmol       Date:  2020-08-10

2.  Comparison of Three Tonometers in Measuring Intraocular Pressure in Eyes That Underwent Myopic Laser in situ Keratomileusis and Photorefractive Keratectomy.

Authors:  Robert Edward T Ang; Neiman Vincent R Bargas; Gladness Henna A Martinez; George Michael N Sosuan; Maria Isabel Nabor-Umali
Journal:  Clin Ophthalmol       Date:  2022-05-27

3.  Modified Goldmann prism intraocular pressure measurement accuracy and correlation to corneal biomechanical metrics: multicentre randomised clinical trial.

Authors:  Sean Joseph McCafferty; Kyle Tetrault; Ann McColgin; Warren Chue; Jason Levine; Melissa Muller
Journal:  Br J Ophthalmol       Date:  2019-02-22       Impact factor: 4.638

4.  Central Corneal Thickness Among Filipino Patients in an Ambulatory Eye Surgery Center Using Anterior Segment Optical Coherence Tomography.

Authors:  George Michael N Sosuan; Maria Imelda R Yap-Veloso
Journal:  Clin Ophthalmol       Date:  2021-06-23
  4 in total